Chapter 4 - Equipment Management PDF

Document Details

ImmenseWerewolf

Uploaded by ImmenseWerewolf

Georgian Technical University

Tags

laboratory equipment equipment management quality management laboratory procedures

Summary

This document offers an overview of equipment management in a laboratory setting, covering essential elements such as selection, installation, calibration, maintenance, and troubleshooting. It provides insights into important considerations and considerations for a well-functioning laboratory environment.

Full Transcript

Chapter 4 Equipment 4-1. Equipment Management Overview Equipment management is one of the essential elements of a quality management system. Proper management of the equipment in the laboratory is necessary to ensure accurate, reliable, and timely testing. The benefits of a good equipment ma...

Chapter 4 Equipment 4-1. Equipment Management Overview Equipment management is one of the essential elements of a quality management system. Proper management of the equipment in the laboratory is necessary to ensure accurate, reliable, and timely testing. The benefits of a good equipment management program are many: helps to maintain a high level of laboratory performance; reduces variation in test results, and improves the technologist’s confidence in the accuracy of testing results; lowers repair costs, as fewer repairs will be needed for a well-maintained instrument; lengthens instrument life; reduces interruption of services due to breakdowns and failures; increases safety for workers; produces greater customer satisfaction. A great deal of thought and planning should go into equipment management. As the laboratory puts an equipment management program in place the following elements should be considered: Selection and purchasing—When obtaining new equipment what criteria should be used to select equipment? Should equipment be purchased, or would it be better to lease? Installation—For new equipment, what are the installation requirements, and who will install the new instrument? Calibration and performance evaluation—What is needed to calibrate and validate that the equipment is operating correctly? How will these important procedures be conducted for both old and new instruments? Maintenance—What maintenance schedule is recommended by the manufacturer? Will the laboratory need additional preventive maintenance procedures?  Are current maintenance procedures being conducted properly? Troubleshooting—Is there a clear procedure for troubleshooting for each instrument? Service and repair—What is the cost? Can the laboratory obtain the necessary service and repair in its geographical area? Retiring and disposing of equipment—What must be done to dispose of old equipment when it needs to be replaced? It is the responsibility of the laboratory director to: oversee all the equipment management systems in the laboratory; ensure that all persons who will be using the instruments have been appropriately trained and understand how to both properly operate the instrument and perform all necessary routine maintenance procedures. Equipment management responsibility may be specifically assigned to a technologist in the laboratory. In many laboratories there is a person who has good skills with equipment maintenance and troubleshooting. Giving this person the role of oversight of all equipment is recommended. Oversight of an equipment management program includes: assigning responsibilities for all activities; assuring that all personnel are trained on operation and maintenance; monitoring the equipment management activities: 1. Review all equipment records routinely; 2. update maintenance procedures as necessary; 3. ensure that all procedures are followed. Note: day-to-day maintenance should be the responsibility of the technical operator. Everyone who uses the equipment should be trained in calibration and daily maintenance. 4-2. Selecting and Acquiring Equipment Selecting the best instrument for the laboratory is a very important part of equipment management. Some criteria to consider when selecting laboratory equipment are listed below. Why and how will the equipment be used? The instrument should be matched against the service the laboratory provides. What are the performance characteristics of the instrument? Is it sufficiently accurate and reproducible to suit the needs of the testing to be done? What are the facility requirements, including the requirements for physical space? Will the cost of the equipment be within the laboratory’s budget? Will reagents be readily available? Will reagents be provided free of charge for a limited period of time? If so, for how long? How easy will it be for staff to operate? Will instructions be available in a language that is understood? Is there a retailer for the equipment in the country, with available services? Does the equipment have a warranty? Are there any safety issues to consider? If the decisions about purchasing are made outside the laboratory, for example by a central purchasing body, the laboratory manager should provide information that will support selecting equipment that will best serve the needs of the laboratory. In areas where there are national programs for purchasing standard equipment, the laboratories of the country should have some input to decisions. In addition, in areas where donors are likely to provide some of the equipment that is used, laboratory management should have input into choice of equipment. If this is not possible, management should consider declining equipment, if inappropriate for laboratory needs. Is it better to purchase, rent or lease equipment? When making this decision, it is a good idea to factor in repair costs. The initial cost of an instrument may seem reasonable, but it may be expensive to repair. Also consider savings that could be negotiated if the laboratory needs more than one piece of equipment. The manufacturer should provide all of the necessary information to operate and maintain equipment. Before purchasing ask if: wiring diagrams, computer software information, a list of parts needed, and an operator’s manual are provided; the manufacturer will install the equipment and train staff (covering travel expenses as necessary) as part of the purchase price; the warranty includes a trial period to verify that the instrument performs as expected; the manufacturer’s maintenance can be included in the contract and if so, whether maintenance is provided on a regular basis. Determine if the laboratory can provide all the necessary physical requirements, such as electricity, water, and space. There must be adequate room to move the equipment into the laboratory; consider door openings and elevator access. Before equipment is installed, verify that all physical requirements (electrical, space, doors, ventilation, and water supply) have been met. Other things to consider are: The vendor’s responsibilities for installation should be confirmed in writing prior to beginning the installation process. A checklist of the expected performance specifications should be developed, so that performance can be quickly verified as soon as the equipment is installed. Whenever possible, it is best to have the manufacturer install laboratory equipment; this will likely improve the conditions of the warranty, and also may ensure that the installation is done properly and quickly. If equipment is installed by the laboratory: check that the package contents contain all of the parts; make a copy of any software that is part of the system; do not allow the equipment to be used before it is completely installed, performance is verified, and testing personnel are trained. 4-3. Getting Equipment Ready for Service After equipment has been installed, the following details need to be addressed before putting the equipment into service: assign responsibility for performing the maintenance and operation programs; develop a system for recording the use of parts and supplies implement a written plan for calibration, performance verification, and proper operation of the equipment; establish a scheduled maintenance program that includes daily, weekly, and monthly maintenance tasks; provide training for all operators; only personnel who have been trained specifically to properly use the equipment should be authorized as operators. Designate those authorized to use the equipment and when it is to be used. Follow the manufacturer’s directions carefully when performing the initial calibration of the instrument. It is a good idea to calibrate the instrument with each test run, when first putting it into service. Determine how often the instrument will need to be recalibrated, based on its stability and on manufacturer’s recommendation. It may be advantageous to use calibrators provided by or purchased from the manufacturer. Prior to testing patient specimens, it is important to evaluate the performance of new equipment to ensure it is working correctly with respect to accuracy and precision. In addition, test methods using kits or laboratory instruments need to be evaluated for the ability to detect disease (sensitivity, specificity, positive and negative predictive value), and to determine normal and reportable ranges. Verification of manufacturers’ performance claims—Manufacturers provide performance evaluations for testing methods using their kits or instruments, and include the information in the package inserts or operators’ manuals. However, laboratories need to verify the manufacturers’ performance claims, and demonstrate they can get the same results using the kits or equipment in them laboratory, with their personnel. Some of the steps that should be followed to verify performance include: testing samples with known values and comparing the results to the expected or certified value; if equipment is temperature controlled, establishing the stability and uniformity of the temperature. Validation of new equipment and associated techniques— If the equipment and/or associated techniques are new, validation processes will be important. This can be carried out by running samples in parallel on both previous and new equipment/methods for a period of time to determine that the expected results can be obtained. These validation procedures should be completely recorded. In order to verify that equipment is working according to the manufacturer’s specifications, it is necessary to monitor instrument parameters by performing periodic function checks. This should be done before using the instrument initially, then with the frequency recommended by the manufacturer. These function checks should also be done following any instrument repairs. Some examples of function checks are daily monitoring of temperatures and checking the accuracy of wavelength calibration. 4-3. Implementing an Equipment Maintenance Program Preventive maintenance includes measures such as systematic and routine cleaning, adjustment, and replacement of equipment parts at scheduled intervals. Manufacturers generally recommend a set of equipment maintenance tasks that should be performed at regular intervals: daily, weekly, monthly, or yearly. Following these recommendations will ensure that the equipment performs at maximum efficiency and will increase the lifespan of the equipment. This will also help to prevent: inaccurate test results due to equipment failure delays in reporting results lower productivity large repair costs. A maintenance plan will include preventive maintenance procedures as well as provision for inventory, troubleshooting, and repair of equipment. When implementing an equipment maintenance program, some of the initial steps will include what follows. Assign responsibility for providing oversight. Develop written policies and procedures for maintaining equipment, including routine maintenance plans for each piece of equipment. The plan should specify the frequency with which all maintenance tasks should be performed. Develop the format for records, create logs and forms, and establish the processes to maintain records. Train staff on the use and maintenance of the equipment, and assure that all staff understand their specific responsibilities. It is recommended that a label be attached to the instrument indicating when the next maintenance or service should be performed. The laboratory should keep an inventory log of all equipment in the laboratory. The log should be updated with information on new equipment, as it is added, and include documentation of when old equipment is retired. For each piece of equipment, the equipment inventory log should have a record of: instrument type, make and model number, and serial number of the instrument, so that any problems can be discussed with the manufacturer; date the equipment was purchased, and whether it was purchased new, used, or reconditioned; manufacturer/vendor contact information; presence or absence of documentation, spare parts, and maintenance contract; warranty’s expiration date; specific inventory number indicating the year of acquisition; this is especially useful for larger laboratories. For example, use the style “YY-number” (04 001, 04-002, etc.) where “YY-number” equals the last two numbers of the year followed by a number attributed in the year. An inventory process must be conducted if the laboratory does not have an existing inventory system for equipment. This could be conveniently organized following a model grid room by room, for example, reception, sampling, serology, and parasitology. During the inventory, the condition of the equipment should be documented as: o functional o partially functional, or o non-functional. Equipment that is not functioning needs to be evaluated as to whether or not it can be repaired. Non-repairable equipment should be retired, and work should be scheduled for equipment needing repair. To ensure that the laboratory does not run out of spare parts, an inventory record of those used most frequently should be kept for each piece of equipment. The record should include: part name and number; average use of the part, and the minimum to keep on hand; cost; date when the part is placed into storage, and when it is used (in and out stock log); quantity of each part remaining in inventory. 3-5. Troubleshooting, Service, Repair, and Retiring Equipment Problems with equipment may present in many ways. The operator may notice subtle changes such as drift in quality control or calibrator values or obvious flaws in equipment function. Sometimes, the equipment fails to operate. It is important to teach operators to troubleshoot equipment problems in order to quickly get the equipment functioning and resume testing as rapidly as possible. When an operator observes instrument drift, it is important to repeat the preventive maintenance procedures as a first step to resolve the problem. If this does not work, proceed with troubleshooting processes. Manufacturers frequently provide a flowchart that can help determine the source of problems. Some of the questions to consider are listed below: Is the problem related to a poor sample? Has the sample been collected and stored properly? Are factors such as turbidity or coagulation affecting instrument performance? Is there a problem with the reagents? Have they been stored properly, and are they still in date? Have new lot numbers been introduced without updating instrument calibration? Is there a problem with the water or electrical supply? Is there a problem with the equipment? Make one change at a time based on symptoms. If equipment is the problem, review the manufacturer’s instructions to verify that all procedures are being followed correctly. If problems cannot be identified and corrected in-house, attempt to find a way to continue testing until the equipment can be repaired. Some ways to achieve this are as follows: Arrange to have access to back-up instruments. It is often too costly for the laboratory to have its own back-up instruments, but sometimes a central stores agency can maintain back-up instruments to be shared throughout the local area or country. Ask the manufacturer to provide a replacement instrument during repairs. Send the samples to a nearby laboratory for testing. Be sure to notify the appropriate providers that that there are problems and that there will probably be delays in completing the testing. Do NOT use faulty equipment! Seek help from the manufacturer or other technical expert. Place a note on the equipment so all staff are aware that it is not in use. Manufacturers may provide service and repair of equipment that is purchased from them. Be sure to set up a procedure for scheduling service that must be periodically performed by the manufacturer. When instruments need repair, remember that some warranties require that repairs be handled only by the manufacturer. Large facilities sometimes have biomedical service technicians in-house who perform equipment maintenance and repair. Routine service should be scheduled so as not to interrupt the flow of work. It is very important to have a policy and procedures for retiring older laboratory equipment. This will usually occur when it is clear that the instrument is not functioning and is not repairable, or when it is outmoded and should be replaced with new equipment. Once a piece of equipment is fully retired and it has been determined that it has no further usefulness, it should be disposed of in an appropriate manner. This last step is often neglected in laboratories, and old equipment accumulates, taking up valuable space and sometimes creating a hazard. When disposing of equipment, salvage any useable parts, particularly if the equipment is being replaced with another similar one. Then consider any potential biohazards, and follow all safety disposal procedures. 3-6. Equipment Maintenance Documentation Equipment documents and records are an essential part of the quality system. The policies and procedures for maintenance should be defined in appropriate documents, and keeping good equipment records will allow for thorough evaluation of any problems that arise. (Module 16: Documents and Records) Each major piece of equipment will have its own equipment maintenance document. Smaller, commonly used equipment such as centrifuges and pipettes may be managed with an equipment maintenance document or manual that deals with all such equipment in the laboratory. An equipment maintenance document should include: step-by-step instructions for routine maintenance, including frequency of performance, and how to keep records of performance; instructions for carrying out function checks, frequency of performance, and how to record the results; directions for calibrating the instrument; guide for troubleshooting; any required manufacturer’s service and repair; list of any specific items needed for use and maintenance, such as spare parts. For major equipment, include identification of the specific instrument, and perhaps information on its performance. Each piece of equipment should have a dedicated logbook documenting all characteristics and maintenance elements: preventive maintenance activities and schedule; recording of function checks and calibration; any maintenance performed by the manufacturer; full information on any problem that the instrument develops, the subsequent troubleshooting activity, and follow-up information regarding resolution of the problem. In recording problems, be sure to record:  date problem occurred, and when equipment was removed from service;  reason for breakdown or failure;  corrective action taken; including a note about any service provided by the manufacturer;  date returned to use;  any changes to procedure for maintenance or function checks as a result of the problem. Some of the tools that are helpful for keeping records on equipment management are: charts logs checklists graphs service reports. The log book should be available for review during the entire life of the equipment. Summary All laboratories should have a well-organized equipment management program. The program should address equipment selection, preventive maintenance, and procedures for troubleshooting and repair. It is essential that good documents and records be maintained. These will include a complete and accurate inventory of all laboratory equipment, documents provided by the manufacturer on operation, maintenance, and troubleshooting, and records of all preventive maintenance and repair activities. A good equipment maintenance program results in a high level of performance and greater confidence in the reliability of results. A significant benefit to the laboratory will be fewer interruptions in test performance, lower repair costs, and elimination of premature replacement of equipment. Increased safety for laboratory workers will result from well-maintained equipment. 3-7. Laboratory Centrifuge Centrifuges are instruments that spin samples at high speeds, forcing the heavier particles to the bottom of the container (usually a tube). The part of the centrifuge that holds the tubes and rotates during operation is the rotor. Centrifuges are frequently used to separate the cellular components of blood from serum or plasma and to centrifuge urine to obtain urine sediment. Centrifuges vary in size, capacity, and speed capability. Clinical centrifuge is the name given to models that can be used for urinalysis or serum separation. Clinical centrifuges can be large floor models or small enough to fit on a benchtop. These usually have a speed capacity of 0 to 3000 rpm (revolutions per minute), and hold tubes ranging in size from 5 to 50 mL, depending on the rotor or tube carriers. Correct way to balance tubes in a centrifuge rotor: Top, placing tubes in the rotor of a tabletop clinical centrifuge; Bottom, A-C, diagrams showing correct placement of tubes (small, dark blue circles) in rotor and, D, incorrect placement of tubes. Unquestionably the medical laboratory plays an important role in the detection, diagnosis, and ultimately the treatment of medical disorders. As emphasized earlier in this program, the meticulous handling of patient samples from collection, preparation, and all the way through analysis is essential for accurate test results. Centrifugation is one of the most important steps in the preparation of clinical specimens for analysis. Consequently, it is paramount that those working in the medical laboratory not only have a basic knowledge in the theory of centrifugation but also be proficient in how to safely use and maintain centrifuges in their laboratory. The next two sections will focus on the steps necessary to safely use a centrifuge and how to keep it functioning properly. Fortunately, centrifuge accidents are rare. When they do occur, human error is the most likely cause. Over the last several years, advancements in centrifugation technology have resulted in enhanced performance, ease of use, and built-in safety features. Nevertheless, one must always keep in mind that a well-trained, centrifuge-savvy laboratory staff will be of immense assistance in preventing centrifuge accidents and costly repairs due to misuse.  It is important that the laboratory staff whose work requires a centrifuge be familiar with the manufacturer’s operating instructions for each make and model of centrifuge located in the laboratory.  When using a centrifuge, proper protective equipment should be employed, such as eyewear, gloves, and a laboratory coat.  Make sure the centrifuge is on a flat, sturdy surface prior to operation.  Confirm that the rotor is properly seated on the drive shaft, that the chamber and tube holders/ buckets are clean and dry, and the spindle is clean.  Verify that all tube holders and buckets are in place and matched. (See Figures 1 & 2) Do not interchange tube holders and buckets with other centrifuges. Use only the tube holders and buckets that came with that model of centrifuge.  Before operating the centrifuge inspect all tube holders/buckets and rotor for any cracks or flaws.  When centrifuging possible hazardous material, make sure that biocontainment lids are in use and locked into place.  Sample tubes should be evenly filled, taking care not to overfill. Both swing-out and fixed rotors must be loaded symmetrically, with opposing tubes being of the same type and filled to the same mass. For example, if you have a tube filled with 7 mL of sample and another tube with 4 mL of sample you will need two additional tubes in order to load the rotor symmetrically. This is accomplished by filling another tube with 7 mL of water or a liquid of similar density and then a second tube to match the 4 mL sample.  When loading the rotor, the opposing tubes must be of the same weight in order to “balance” the centrifuge.  Capped tubes should always remain capped during centrifugation. This not only helps prevent possible contamination by infectious agents but also any dehydration of the fluid contained in the tubes resulting from the centrifugation.  As noted previously, for each tube placed into the rotor, a tube of equal weight must be placed into the rotor directly opposite it. By doing so, the operator is ensuring that the center of gravity remains at the center of the rotor. This is critical because balancing the centrifuge not only prevents damage but also better assures a safe operation.  Do not exceed the manufacturer’s recommended maximum speed. The operator should remain with the centrifuge until it reaches the set speed.  A minor vibration in the centrifuge is normal and to be expected, but excessive amounts of vibration can mean danger. If the centrifuge begins to shake or wobble, it more than likely means that it is not balanced properly and should be stopped immediately. 3-8. Microliter pipette Micropipettes come in different sizes. At Labster we use the following pipettes: To adjust the volume, turn the adjustable wheel or the plunger (depending on the pipette) until the desired volume is displayed on the readout. It is very important to stay within the volume range of the respective pipette to avoid contaminating the pipette or producing inaccurate measurements. Proper pipetting technique is important to ensure an accurate volume of liquid is transferred. Microliter pipette instructions After selecting a pipette with an appropriate volume, you are ready to transfer liquids using the following steps. Drawing up liquid Before drawing up liquid, make sure you add a sterile tip to the pipette. Then press the plunger from the rest position to the first stop. Hold the pipette vertically and dip the pipette tip 3-4 mm into the liquid and carefully release the plunger in a controlled manner (1). Dispensing liquid To dispense liquid, hold the pipette tip against the side of the tube at a 30-40 deg angle or dip the tip beneath the surface of the liquid already contained in the tube. Then press the plunger to the second stop to make sure no fluid remains inside the tip. Discard the tip into the trash after each pipetting event by using the ejector. By serial dilution Whereas if it was done by serial dilution, we first look at the final volume and concentrations needed, then determine the dilution factor. So if we need 10 ml of a 1 mM solution, our dilution factor is: Dilution factor = 1 / (final concentration / initial concentration) = 1 / (0.001M / 1M) = 1000X dilution factor equals one divided by the final concentration divided by the initial concentration, which equals one divided by 0.001 molar divided by 1 molar, which equals 1000 We can split up 1000X into multiple factors: 10 X 10 X 10. This means we can set up 3 tubes to perform the serial dilution. When it is a 10X dilution, it means that it is 1 part of the solution, 9 parts of the diluent (usually it is water). Similarly, for a 15X dilution, a thousand times dilution into multiple factors: 3 times a 10-fold dilution. This means we can set up 3 tubes to perform the serial dilution. When it is a 10-fold dilution, it means that it is 1 part of the solution, 9 parts of the diluent (usually it is water). Similarly, for a 15-fold dilution, it will be 1 part of the solution, 14 parts of the diluent. 1 mL of stock is transferred to the second tube that contains 9 mL of diluent (e.g water). Then 1 mL (aliquot) of the second tube is taken and transferred to the third tube. The process of transferring aliquot is repeated until the desired dilution concentration is achieved. Although serial dilution requires more preparation, it will give more accurate dilutions, provided that your technique is precise. terilization refers to the removal of all microorganisms in a material or on an object. Sterilization ensures that even fungal spores or highly resistant bacterial endospores are killed. Sterilizing media The medium and its container must be sterilized by autoclaving before use, to avoid contamination. The autoclave utilizes high-pressure saturated steam at 121o C for 15 to 20 minutes to kill the microorganisms in the medium and its container. To ensure that the autoclave procedure has been conducted correctly, the bottle that contains the medium is labeled with a piece of autoclave tape. The autoclave tape has a white stripe which will turn black if subjected to high-temperature steam. Therefore, we can identify whether a bottle of medium has been autoclaved/sterilized or not by looking at the stripe color of the autoclave tape. Autoclave tape. Before autoclave; white stripe, After autoclave: black stripe Sterilizing equipment Lab equipment can be sterilized by autoclaving, heating it up on a flame or using disinfectants such as ethanol to kill germs.

Use Quizgecko on...
Browser
Browser