ALTAIR 5X PID.pdf
Document Details
Uploaded by ValiantSunstone
Tags
Full Transcript
Description 3 Description 3.1 Overview 1 1 8 9...
Description 3 Description 3.1 Overview 1 1 8 9 11 2 7 10 3 6 4 5 13 12 Fig. 2 Device view LEDs 1 8 IRDA communication port 2 red "Alarm", 1 green "Safe" and 1 yellow "Fault" 2 Horn 9 Pump inlet 3 Display 10 Filter US 4 Button 11 RFID tag 5 q Button 12 Charging port 6 Button 13 Charge Status LED 7 Bluetooth Status LED The device monitors gases in ambient air and in the workplace. The ALTAIR 5X PID is available with a maximum of five sensors, which can display readings for six separate gases (one Two-Tox Sensor provides both CO and H2S or CO and NO2 sensing capabilities in a single package). The ALTAIR 5X PID Multigas Detector is only available with a color display. The alarm levels for the individual gases are factory-set and can be changed through the Instrument Setup Menu. These changes can also be made through MSA Link Software. Ensure that the latest version of the MSA Link software has been downloaded from MSA’s website www.msasafety.com. It is recommended that after making changes using MSA Link software, the device should be turned OFF and ON. ALTAIR® 5X PID 14 Description While the device can detect up to 30 % oxygen in ambient air, it is approved for use only up to 21 % oxygen. 3.2 Device Hardware Interfaces Device operation is dialog driven from the display with the aid of the three function buttons (→ Fig. 2 "Device view"). The device has three buttons for user operation. Each button can function as a "soft key", as defined on the display directly above the button. Button Definitions Button Description q The q button is used to turn the device ON or OFF and to confirm user action selections. The button is used to page down through data screens or to decrease the values in setup mode. This button is also used to initiate a Bump Test for the installed sensors, directly from the MEASURING page.If the user is granted access to the MotionAlert setting feature, this button can be used to activate the InstantAlertTM alarm. See chapter 4.5 "Device Setup" for the means to allow/disallow user access. The button is used to reset Peak, STEL, TWA and alarms (where possible) or perform calibra- tion in measuring mode. It is also used as page up or to increase the values in setup mode. When the button and the button are pressed simultaneously while in normal measure mode, the Setup mode can be entered after the password is confirmed. LED Definitions LED Description The red alarm LEDs are visual indications of an alarm condition or any type of error in the RED (Alarm) device. The Safe LED flashes once every 15 seconds to notify the user that the device is ON and operating under the conditions defined below: The green SAFE LED is enabled Combustible reading is 0 % LEL or 0 % Vol Oxygen (O2) reading is 20.8 % GREEN (Safe) US All other sensor readings are 0 ppm No gas alarms are present (low or high) Device is not in Low Battery warning or alarm STEL and TWA readings are 0 ppm This option can be turned OFF through the MSA Link software. The Fault LED activates if any of several fault conditions are detected during device oper- ation. This includes: A device memory error YELLOW (Fault) A sensor determined to be missing or inoperative A pump fault These faults are also indicated by activation of device alarm LEDs, horn, and vibrating alarm. The blue LED is a visual indication of the Bluetooth connection status. BLUE Off = Bluetooth board OFF or Undiscoverable (Bluetooth Status) Fast Flash = Discoverable Mode Slow Flash = Connected ALTAIR® 5X PID 15 Description 3.3 Alarms The device is equipped with multiple alarms for increased user safety: Icon Alarm The device vibrates when any alarm condition is active. This can be turned OFF Vibrating Alarm through the SETUP- ALARM OPTIONS menu (→ chapter 4.5 "Device Setup"). The device is equipped with an audible alarm. Horn The horn can be turned OFF through the SETUP- ALARM OPTIONS menu (→ chapter 4.5 "Device Setup"). The InstantAlert exclusive feature allows the user to manually activate an audible alarm to alert those nearby to potentially dangerous situations. Holding the InstantAlert™ button for approximately 5 seconds while in Normal Measure Mode activates Alarm the InstantAlert alarm. Access to this feature may be restricted by user settings. See chapter 4.5 "Device Setup" for means to allow/disallow user access. If MotionAlert is turned ON (→ chapter 4.5 "Device Setup"), the device activates a "Man Down" alarm if motion is not detected within 30 seconds. The Alarm LEDs MotionAlert™ flash, and the horn activates with an increasing audible frequency. MotionAlert is Alarm always turned OFF when the device is turned OFF. Access to this feature may be restricted by user settings. See chapter 4.5 "Device Setup" for means to allow/disallow user access. Stealth Mode disables the visual, audible and vibrating alarms. MSA recommends that this feature be left in its default "OFF" state. Stealth mode can be turned ON Stealth Mode through the SETUP - INSTRUMENT OPTIONS menu (→ chapter 4.5 "Device Setup"). On the display, all three alarm icons are shown as OFF. The device evaluates the condition of the sensors during Calibration. As the end of a sensor’s life approaches, a warning is provided. While the sensor is still fully functional, the warning gives the user time to plan for a replacement sensor to minimize downtime. The Sensor Life indicator ♥ displays during ongoing operations as a reminder of a sensor’s pending end of life. When a sensor’s end-of-life is reached, sensor calibration will not be successful, and the user is then alerted by a Sensor Life Alarm. A flashing Sensor Life Sensor Life indicator ♥ displays during ongoing operations until the sensor is replaced and/or US Alarm successfully calibrated. On the display, each displayed gas will have its own Sensor Life indicator. If a sensor is in end-of-life warning, its indicator will be an orange ♥. If a sensor has reached end-of-life, it is in alarm and its Sensor Life indicator will be a continuous blinking red ♥. See chapter 4.10 "Calibration" for additional details on Sensor Life determination and indication. The backlight automatically activates when any front panel button is pressed and remains ON for the duration of user-selected timeout. Backlight This duration can be changed using the SETUP - INSTRUMENT SETUP (→ chapter 4.5 "Device Setup") or through MSA Link software. ALTAIR® 5X PID 16 Description Icon Alarm This operating beep activates every 30 seconds by momentarily beeping the horn and flashing the alarm LEDs under the following conditions: Operating beep is enabled Operating Device is on normal Measure Gases page Beep Device is not in battery warning Device is not in gas alarm. The Operating Beep can be disabled using the SETUP - INSTRUMENT OPTIONS (→ chapter 4.5 "Device Setup") or through MSA Link software. 3.4 On-Screen Indicators 5 6 7 8 9 10 11 4 3 2 12 13 1 14 Fig. 3 Color display US 1 "Soft Key" Indicator 8 Vibration alarm OFF Horn OFF or successful Bump Test/ 2 Gas reading 9 Calibration Indicator 3 Gas Concentration Units 10 LED OFF 4 Current Time 11 Battery charge level 5 Motion Alert symbol ON 12 Gas Type 6 Bluetooth ON 13 "Soft Key" Indicator Combustible Gas Type/ VOC Gas Type alternates 7 14 ♥ Sensor Life Indicator (the display switches between each type every few seconds) ALTAIR® 5X PID 17 Description Battery Charge Level Indicator The battery condition icon continuously displays in the upper right-hand corner of the display. A bar represents the charging level of the battery. The nominal run-time of the device (COMB, O2, CO, H2S, and PID sensor) at room temperature is 12 hours. Actual run-time will vary depending on ambient temperature, battery and alarm conditions. Low Battery Warning WARNING! If battery warning alarm activates while using the device, leave the area immediately as the end of battery life is approaching. Failure to follow this warning can result in serious personal injury or death. Fig. 4 Battery Warning The duration of remaining device operation during a Low Battery Warning depends on ambient temperatures, battery condition alarm status. Nominal battery life is 30-60 minutes after the Battery Warning activates. When the device goes into battery warning the: battery life indicator continuously blinks alarm sounds and alarm LEDs flash every 30 seconds Safe LED no longer flashes device continues to operate until it is turned OFF or battery shutdown occurs. Battery Shut Down US WARNING! If battery alarm displays, stop using the device as it no longer has enough power to indicate potential hazards, and persons relying on this product for their safety could sustain serious personal injury or death. The device goes into battery shutdown mode 60 seconds before final shutdown (when the batteries can no longer operate the device): "BATTERY ALARM” flashes on the display Alarm sounds Alarm LEDs flash Fault LED is on No other pages can be viewed; after approximately one minute, the device automatically turns OFF. ALTAIR® 5X PID 18 Description Fig. 5 Battery Shut Down When battery shutdown condition occurs (shown in Fig. 5): (4) Leave the area immediately. (5) Recharge or replace the battery pack. Battery Charging WARNING! Risk of explosion: Do not recharge device in hazardous area. WARNING! Use of any charger, other than the charger supplied with the device, may damage or improperly charge the batteries. For users in Australia/ New Zealand: The charge cradle is a Class A product. In a domestic environment, this product may cause radio interference, in which case, the user may be required to take adequate measures. The charger is capable of charging a completely depleted pack in less than six hours in normal, room- temperature environments. Allow very hot or cold devices to stabilize for one hour at room temperature before attempting to charge. US Minimum and maximum ambient temperature to charge the device is 10 °C (50 °F) and 35 °C (95 °F), respectively. For best results, charge the device at room temperature 23 °C (73 °F). To Charge the Device Firmly insert the charger connector into the charge port on the back of the device. An LED in the battery pack is used to indicate on the charge status. Red = charging, Green = charged, yellow = fault If a problem is detected during charging (LED turns yellow): Disconnect the charger momentarily to reset the charge cycle. The battery pack may be charged separately from the device. During periods of non-use, the charger may remain connected to the device/battery pack. ALTAIR® 5X PID 19 Description The charger must be disconnected for the device to operate. 3.5 Viewing Additional Pages The Main Screen appears at device turn-ON. Additional displays can be viewed by pressing the button to move to the screen as indicated by the "soft key". The sequence of pages are as follows and are described below: Advance on MAIN MEASURE BUMP TEST ? PEAK * MIN * 1 PAGE 1 STEL * * IF ENABLED Advance on 1 NOT VALID FOR ALL SENSORS TWA * 1 US MOTION ALERT STATUS * DATE DISCOVERY CHANGE LAST CAL CAL DUE MODE* VOC DATE * DATE * GAS*? ALTAIR® 5X PID 20 Description Bump Test (BUMP page) This page allows the user to perform an automated Bump Test on the device. To perform the test, the (YES) button is pressed. See chapter 4.9 "Bump Test" for details on performing the Bump Test. If the button is pressed, the Bump Test is not performed, and the display shows the next page in the sequence (PEAK). If the button is pressed, the Bump Test is not performed, and the display reverts back to the normal MEASURE page. Peak Readings (PEAK page) This page shows the highest levels of gas recorded by the device since turn-ON or since peak readings were reset. To reset the peak readings: (1) Access the PEAK page. (2) Press the button. This page can be de-activated through MSA Link software. Minimum Readings (MIN page) This page shows the lowest level of oxygen recorded by the device since turn-ON or since the MIN reading was reset. It is only shown if an oxygen sensor is installed and enabled. To reset the MIN reading: (1) Access the MIN page. (2) Press the button. Short Term Exposure Limits (STEL page) US WARNING! If the STEL alarm activates, leave the contaminated area immediately; the ambient gas concentration has reached the preset STEL alarm level. Failure to follow this warning will cause over-exposure to toxic gases and persons relying on this product for their safety could sustain serious personal injury or death. This page shows the average exposure over a 15-minute period. When the amount of gas detected by device is greater than the STEL limit: Alarm sounds, alarm lights flash. Alarm LEDs flash “STEL ALARM” message flashes. To reset the STEL: ALTAIR® 5X PID 21 Description (3) Access the STEL page. (4) Press the button. The STEL alarm is calculated over a 15-minute exposure. STEL calculation examples: Assume the device has been running for at least 15 minutes: 15 minute exposure of 35 ppm: (15 minutes x 35 ppm) = 35 ppm 15 minutes 10 minute exposure of 35 ppm and 5 minutes exposure of 15 ppm: (10 minutes x 35 ppm) + (5 minutes x 5 ppm) = 25 ppm 15 minutes This page can be de-activated through MSA Link software. Time Weighted Average (TWA page) WARNING! If the TWA alarm activates, leave the contaminated area immediately; the ambient gas concentration has reached the preset TWA alarm level. Failure to follow this warning will cause over-exposure to toxic gases and persons relying on this product for their safety could sustain serious personal injury or death. This page shows the average exposure over 8 hours since the device was turned ON or since the TWA reading was reset. When the amount of gas detected is greater than the eight-hour TWA limit: US Alarm sounds Alarm LEDs flash “TWA ALARM” message flashes. To reset the TWA Readings: (5) Access the TWA page. (6) Press the button. The TWA alarm is calculated over an eight-hour exposure. ALTAIR® 5X PID 22 Description TWA calculation examples: 1 hour exposure of 50 ppm: (1 hour x 50 ppm) + (7 hours x 0 ppm) = 6.25 ppm 8 hours 4 hour exposure of 50 ppm and 4 hour exposure of 100 ppm: (4 hours x 50 ppm) + (4 hours x 100 ppm) = 75 ppm 8 hours 12 hour exposure of 100 ppm: (12 hours x 100 ppm) = 150 ppm 8 hours This page can be de-activated through MSA Link software. Date Display Current date appears on the display in the format: MM-DD-YY. Last cal page Displays the device last successful calibration date in the format: MM-DD-YY. This page can be de- activated through MSA Link software or the SETUP - CAL OPTIONS page. Cal due page Displays the days until the device's next calibration is due (user selectable). This page can be de-acti- vated through MSA Link software or the SETUP - CAL OPTIONS page. Discoverable Mode page US Allows the user to put the device into Bluetooth discoverable mode in order to pair with another device. This page can be deactivated through the SETUP - INSTRUMENT OPTIONS page. Change VOC Gas? Page This page is selectable if the "Menu Enable" feature is ON as described in section 4.5 "Device Setup". This page contains the 10 Favorite PID gases, the All Gases list and the Custom Gas list. An example of this screen is shown below: ALTAIR® 5X PID 23 Description Motion Alert Activation Page When the MotionAlert feature is active, the + symbol appears. The device enters pre-alarm when no motion is detected for 20 seconds. This condition can be cleared by moving the device. MotionAlert is turned OFF each time the device is powered OFF. After 30 seconds of no motion, the full MotionAlert alarm is triggered. This alarm can only be cleared by pressing the button. This page displays if it was selected in Setup Mode. To activate or deactivate the MotionAlert feature, press the button while the MOTIONALERT ACTIVATION page is displayed. 3.6 Sensor Missing Alarm Enabled PID and XCell sensors are continuously monitored for proper function. If, during operation, the PID or an XCell sensor is detected as failed or disconnected, this alarm message appears. "SENSOR MISSING" flashes on the display. The problematic sensor is indicated. The alarm sounds and the Fault and Alarm LEDs flash. The alarm can be silenced by pressing the button; no other pages can be viewed. WARNING! When this alarm occurs, the device is inoperative for measuring gases. The user must exit the hazardous area, the device must be powered down, and the sensor situation must be corrected. 3.7 Monitoring Toxic Gases US The device can monitor the concentration of a variety of toxic gases in ambient air. Which toxic gases are monitored depends on the installed sensors. The device displays the gas concentration in parts per million (ppm), μmol/mol or mg/m3 on the Measuring page. Gas units are selected in the SETUP - INSTRUMENT OPTIONS page. WARNING! If an alarm is triggered while using the device, leave the area immediately.Remaining in the area under such circumstances can cause serious personal injury or death. The device has four gas alarms: HIGH Alarm LOW Alarm STEL Alarm TWA Alarm ALTAIR® 5X PID 24 Description Fig. 6 Alarm Conditions (here High Alarm) The carbon monoxide channel in the device is equipped with an internal filter. The purpose of this filter is to protect the CO sensor from acid gases (H2S, SO2, etc.) and from the hydrocarbons that the device is intended to measure, including the calibration gas, isobutylene. In normal use, an interferent signal for calibration or bump checking the device should not be observed on the CO channel. However, exposure to large amounts of certain hydrocarbons (either long exposure times or high concentrations) can overwhelm the filter and appear as signals on the CO channel. In normal operation, after the hydrocarbon exposure is ended, the filter is designed to outgas absorbed hydrocarbons at a rate that will not cause a signal on the CO channel. However, if the unit is exposed to high temperature (>40°C), this desorption rate increases and spurious signals may be observed on the CO channel due to gassing of previously absorbed hydrocarbons. Typically the CO sensor will recover within 24 hours but extremely high exposures can extend this time. After the recovery period if the CO sensor can no longer be calibrated or shows an elevated reading that cannot be brought to zero with a zero calibration, the CO sensor should be replaced. WARNING! Extremely high levels of VOCs will send the CO sensor into alarm and the sensor may not recover or the recovery period will be significant. Take the impact on sensor performance into account when installing sensors. If the gas concentration reaches or exceeds the alarm set point or the STEL or TWA limits, the: alarm message displays and flashes in combination with the corresponding gas concentration backlight turns on US alarm sounds (if active) alarm LEDs flash (if active) vibrating alarm triggers (if active) 3.8 Monitoring Oxygen Concentration The device monitors the oxygen concentration in ambient air. The alarm set points can be set to acti- vate on two different conditions: Enriched - oxygen concentration > 20.8% or Deficient - oxygen concentration < 19.5%. While the device can detect up to 30% oxygen in the ambient air, it is approved for use only up to 21% oxygen-content. ALTAIR® 5X PID 25 Description WARNING! If an alarm activates while using the device, leave the area immediately. Remaining in the area under such circumstances can cause serious personal injury or death. When the alarm set point is reached for either of the above conditions: the alarm message displays and flashes in combination with the corresponding gas concentration backlight turns on alarm sounds (if active) alarm LEDs flash (if active) vibrating alarm triggers (if active) The LOW alarm (oxygen deficient) is latching and will not automatically reset when the O2 concentra- tion rises above the LOW set point. To reset the alarm press the button. If the alarm is latching, the button silences the alarm for five seconds. Alarms can be made latching or unlatching via MSA Link software. False oxygen alarms can occur due to changes in barometric pressure (altitude), humidity or extreme changes in ambient temperature. It is recommended that an oxygen calibration be performed at the temperature and pressure of use. Be sure that the device is in known fresh air before performing a calibration. 3.9 Monitoring Combustible Gases The device can be equipped with a catalytic combustible sensor that detects a variety of combustible gases up to 100 % LEL and displays the reading as either % LEL or % CH4. WARNING! If an alarm is triggered while using the device, leave the area immediately.Remaining in the area under such circumstances can cause serious personal injury or death. The catalytic combustible sensor has two alarm setpoints: HIGH Alarm LOW Alarm US If the gas concentration reaches or exceeds the alarm set point, the device: alarm message displays and flashes in combination with the corresponding gas concentration: backlight turns on alarm sounds (if active) alarm LEDs flash (if active) Gas Exposure of 100 % LEL When gas reading exceeds 100 % of the lower explosive limit (LEL), the catalytic combustible sensor enters a Lock Alarm state and displays “XXX” in place of the actual reading. WARNING! A catalytic combustible gas reading of "XXX” indicates the atmosphere could be above 100 %LEL or 5.00 % Vol CH4 and an explosion hazard exists. Move away from contaminated area immediately. ALTAIR® 5X PID 26 Description The user can clear the LockAlarm state only by turning the device OFF, and then ON again in a fresh air environment. When catalytic combustible gas reading digits appear, the device is available again for measuring gases. Check national standard values for 100 % LEL. 3.10 Monitoring VOC Gases The device is equipped with a PID sensor that detects a variety of VOC gases. The device displays the gas concentration in parts per million (ppm), µmol/mol or mg/m3 on the Measuring page. WARNING! If an alarm is triggered while using the device, leave the area immediately. Remaining in the area under such circumstances can cause serious personal injury or death. The device has four gas alarms: HIGH Alarm LOW Alarm STEL Alarm TWA Alarm If the gas concentration reaches or exceeds the alarm set point or the STEL or TWA limits, the: alarm message displays and flashes in combination with the corresponding gas concentration backlight turns on alarm sounds (if active) alarm LEDs flash (if active) vibrating alarm triggers (if active) To reset the alarm press the button. False VOC alarms can occur due to changes in barometric pressure (altitude), humidity or extremechanges in ambient temperature. US It is recommended that a VOC calibration be performed at the temperature, humidity and pressure of use. Be sure that the device is in known fresh air before performing a calibration. For optimal lamp strike, the PID lamp should be started within the normal temperature range. When the device is calibrated in an dry, air conditioned environment and taken to a high temperature and high humidity outdoor environment, a VOC Low or High alarm may be triggered by this sudden change. It is recommended that the PID sensors be cleaned prior to this transition to avoid this situation, or to acclimate the sensor to the outdoor conditions in a known safe area. ALTAIR® 5X PID 27 Description 3.11 Displaying Current Response Factor The current Response Factor (RF) is displayed at device startup along with the PID lamp potential in eV value, sensor range and VOC gas type. During operations, the RF can be displayed through several menus. If the Menu Enable option is On, use the button on the Main Measuring page to scroll through the menu options and select YES on "Change VOC Gas?". Selecting any gas on this page will display the 8 character gas name, the Response Factor, the Maximum Value of the VOC gas and the current High and Low Alarm values. The Maximum Value is calculated by multiplying the sensor range by the RF. For example, the US Max Value for Hexane is 2000 *4.5 = 9000 ppm. The maximum value cannot exceed 9999 ppm. It is the responsibility of the user to change the VOC Low and High Alarms as appropriate for the applied RF. The selection of the alarm limits must be under the direction of a qual- ified safety professional who has carefully evaluated the specific hazards of the jobsite where it will be used and who is completely familiar with the product and its limitations. A complete list of the 8 character gas name and Response Factors for all VOC gases is contained in chapter 10 "PID Response Factor Table". ALTAIR® 5X PID 28 Operation 4 Operation Device operation is dialog driven from the display with the aid of the three function buttons (→ Fig. 2 "Device view"). For more information, see the flow charts in chapter 11 "Flow Charts". 4.1 Environmental Factors A number of environmental factors may affect the gas sensor readings, including changes in pressure, humidity and temperature. Pressure and humidity changes affect the amount of oxygen actually present in the atmosphere. Pressure Changes If pressure changes rapidly (e.g., stepping through airlock), the oxygen sensor reading may temporarily shift and possibly cause the device to go into alarm. While the percentage of oxygen may remain at or near 20.8 Vol %, the total amount of oxygen present in the atmosphere available for respiration may become a hazard if the overall pressure is reduced by a significant degree. Humidity Changes If humidity changes by any significant degree (e.g., going from a dry, air conditioned environment to outdoor, moisture laden air), oxygen readings can be reduced by up to 0.5 %, due to water vapor in the air displacing oxygen. The oxygen sensor has a special filter to reduce the effects of humidity changes on oxygen readings. This effect will not be noticed immediately, but slowly impacts oxygen readings over several hours. Temperature Changes The sensors have built-in temperature compensation. However, if temperature shifts dramatically, the sensor reading could shift. Combined Humidity and Temperature Changes When the device is calibrated in an dry, air conditioned environment and taken to a high temperature and high humidity outdoor environment, a VOC Low or High alarm may be triggered by this sudden change. It is recommended that the PID sensors be cleaned prior to this transition to avoid this situa- tion, or to acclimate the sensor to the outdoor conditions in a known safe area. 4.2 Turning ON and Fresh Air Setup Device operation is dialog driven from the display with the aid of the three function buttons (→ Fig. 2 "Device view"). For more information, see the flow charts in chapter 11 "Flow Charts". US Turn the device ON with the q button. The device performs a self test: During the self test, the device checks alarm LEDs, audible alarm, vibrating alarm and installed sensors. The device displays: Startup logo Software version, device serial number, company name, department and user names IC / FCC ID Identifier Sampling system safety test During the turn-ON sequence, if a sensor was changed since the previous device operation, the current listing of the installed sensors displays and user interaction is required. The user must accept the new configuration by pressing the button. ALTAIR® 5X PID 29 Operation If the current sensor configuration is not accepted, the device alarms and is not usable. FCC Identification page Combustible gas type, and installed sensor indication VOC gas type, lamp value, detectable range and Response Factor Alarm setpoints Low Alarm Alarm setpoints High Alarm Alarm setpoints STEL Alarm (if enabled) Alarm setpoints TWA Alarm (if enabled) Settings for calibration cylinder Current date Last calibration date (if enabled) CAL due date. If the calibration due date is enabled, the message "CAL DUE; X DAYS" appears on the device display. X = the number of days until a calibration is due, user selectable for 1 to 180 days. If the number of days until calibration is due reaches 0, an alert occurs and "CAL DUE, NOW" displays. Press the π button to clear the alert Sensor warm-up period Fresh Air Setup option (if enabled). The Main Measure Page will appear. The presence of a ♥ indicator on the display means a sensor is approaching or has reached its end- of-life. See chapter 3.3 "Alarms" for details on the Sensor Life Alarm situation. Refer to flowchart in chapter 11.1 "Basic Operation". Sampling System Safety Test Upon startup, an alarm (visual, audible and vibrating) is triggered and the customer is prompted to block the pumps/sampling system of the device within 30 seconds. When the device detects a pump flow block, it will display a PASS message. The startup sequence will resume. If the device does not detect a pump flow block, it will display an error message. US The device will shut OFF after the customer acknowledges this message by pressing the button. Check the sampling system if this occurs and contact MSA as needed. Users can check the operation of the sampling system any time during operation by blocking the sampling system to generate a pump alarm. WARNING! Do not use the pump, sample line, or probe unless the pump alarm activates when the flow is blocked. Lack of an alarm is an indication that a sample may not be drawn to the sensors, which could cause inaccurate readings.Failure to follow the above can result in serious personal injury or death.Never let the end of the sampling line touch or go under any liquid surface. If liquid is drawn into the device, read- ings will be inaccurate and device could be damaged. MSA recommends the use of an MSA sample probe containing a special membrane filter, permeable to gas but impermeable to water, to prevent such an occurrence. ALTAIR® 5X PID 30 Operation Fresh Air Setup (FAS) at device Turn-ON The Fresh Air Setup (FAS) is for automatic ZERO adjustment of the device. The FAS has limits. If a hazardous level of gas is present, the device ignores the FAS command and the device alarm activates. The ability to perform an FAS at device turn-ON can be disabled by using MSA Link software. WARNING! Do not perform the Fresh Air Setup unless you are certain you are in fresh, uncontaminated air; other- wise, inaccurate readings can occur which can falsely indicate that a hazardous atmosphere is safe. If you have any doubts as to the quality of the surrounding air, do not use the Fresh Air Setup feature. Do not use the Fresh Air Setup as a substitute for daily calibration checks. The calibration check is required to verify span accuracy. Failure to follow this warning can result in serious personal injury or death. Fig. 7 Fresh Air Setup The device displays a blinking "FRESH AIR SETUP?", prompting the user to perform a Fresh Air Setup: (7) Press the button to bypass the Fresh Air Setup. The Fresh Air Setup is skipped and the device goes to the Measuring page (Main page). (8) Press the button to perform the Fresh Air Setup. The device starts the FAS sequence and the FAS screen displays. US A progress bar shows the user how much of the FAS has been completed. At the end of the FAS, the device displays either "FRESH AIR SETUP PASS" or "FRESH AIR SETUP FAIL". If the FAS fails, perform a zero calibration (→ chapter 4.10 "Calibration"). 4.3 Special Consideration for Oxygen Sensor Under the following situations, the oxygen sensor display reading may be suppressed for up to 30 minutes at device turn-ON as a sensor 'cook down' is performed. This could occur if: the oxygen sensor was just installed the battery pack was allowed to be deep-discharged the battery pack was removed from the device. ALTAIR® 5X PID 31 Operation During this time, the oxygen sensor numeric position on the display indicates "PLEASE WAIT". While this message displays, the device cannot respond to a: Fresh Air Setup Calibration Bump Test procedure. When the numeric oxygen reading appears, the FAS, calibration, or Bump Test procedures may be performed. 4.4 Measurement Mode [Normal Operation] The following options pages can be executed from the Measurement screen: BUMP page This page allows user to perform a Bump Test on installed sensors Peak Page* This page shows the peak readings for all sensors. Min Page This page shows the minimum readings for the oxygen sensor. STEL Page* This page shows the calculated STEL readings of the device. TWA Page* This page shows the calculated TWA readings of the device. Date Page This page shows actual date settings of the device. Last Cal Date This page shows the date of the last calibration. Cal Due* This page shows the set date for the next calibration. Change VOC This page allows the VOC gas type to be changed Gas? Discoverable This page allows the user to put the device into Bluetooth discoverable US Mode mode in order to pair with another device. Motion Alert This page allows the Motion Alert Feature to be activated or deactivated. * The display of these pages can be de-activated through MSA Link software For further information see chapter 12 "Changeable Feature Summary". ALTAIR® 5X PID 32 Operation 4.5 Device Setup The device has provisions to access and modify the following parameters through direct button inter- face: Calibration Options Alarm Options Instrument Options These menus can be accessed only from the measure page by pressing and holding the and buttons simultaneously until prompted for a password. The operation is as follows: (1) Turn the device ON and wait until the measure page appears. (2) Simultaneously press and hold the and buttons for approximately five seconds. a) The default password is "672". PASSWORD 000 (3) Enter the first digit by pressing the or button and confirm with the q button. The cursor jumps to the second digit. (4) Enter the second as well as the third digits. Incorrect password: device returns to the Main Page. Correct password: user can enter the Setup mode. The password can be changed with a PC through the MSA Link software.If the password is forgotten, it can be reset by using MSA Link software. Contact MSA Customer Service for assistance. The following Options are available by pressing the and buttons: Calibration Options - see chapter "Calibration Setup" Alarm Options - see chapter "Alarm Setup" Instrument Options - see chapter "Instrument Options" Calibration Setup US CALIBRATION OPTIONS The Calibration Options menu has provisions to: modify the calibration cylinder settings (CYLINDER SETUP) enable/disable calibration due and to set the number of days (CAL DUE OPTIONS) enable/disable the option to show the last cal date at turn on and (LAST CAL DATE) When enabled, the date of the last device calibration displays during the turn-ON process. enable/disable the option for password protected calibration (CAL PASSWORD) When enabled, the device setup password must be entered prior to calibration. ALTAIR® 5X PID 33 Operation Press: the button go to next page the button to go previous page the q button to enter setup. Setting Calibration Cylinder This option has a dialog similar to the span calibration dialog. The display shows all active sensors. (1) Press the q button to enter setup. The screen for the first calibration cylinder displays. (2) Press the or button to change the value. the q button to confirm the setup. With this confirmation the device automatically moves to the next cylinder setting. (3) Repeat the sequence for changing the required settings for all necessary gas values. After the last setting is performed, the device returns to the Calibration Options menu. The only allowed calibration gas for the 0-2000 ppm PID sensor is 100 ppm isobutylene balanced in air. Higher concentrations can cause false readings of the CO sensor. Setting Cal Due Options (1) Press the q button to enter setup. (2) Press the or button to enable/disable this option. (3) Press the q button to confirm. (4) After confirmation the device prompts the user to enter the number of days for the reminder. (5) Change number of days by pressing the or button. (6) Press the q button to go to the next menu. Setting Last Cal Date (1) Press the q button to enable/disable this option. (2) Press the button to go to the next page. US (3) Press the button to go to the previous page. Setting Calibration Password (1) Press the q button to enable/disable this option. (2) Press the button to go to the next page. (3) Press the button to go to the previous page. Back To Main Menu (1) Press the q button to go to Device Setup Menu The Cal Options screen displays (2) Press the button to go to the next (Alarm options) or the button to exit the Setup menu. ALTAIR® 5X PID 34 Operation Alarm Setup ALARM OPTIONS The Alarm Options Menu allows the user to: enable/disable the vibrating alarm enable/disable the audible alarm (horn) enable/disable the Alarm LEDs enable/disable the MOTIONALERT SELECTION page. If disabled, the user cannot change the device MotionAlert setting. set Sensor Alarms. Press the button go to next page the button to go previous page the q button to enter setup. Setting Vibrating Alarm Press the q button to enable/disable this option. Setting Horn Alarm Press q button to enable/disable this option. Setting LED Alarm Press q button to enable/disable this option. Setting MotionAlert Access Setting this parameter allows the user to access the MOTIONALERT page from the MEASURE page. If access is denied here: the user cannot access the MOTIONALERT page to enable or disable that feature the InstantAlert feature (chapter 3.3 "Alarms") cannot be activated. US (1) To grant or deny user access to the MOTIONALERT page, use the button to change the indi- cated selection. User access is: permitted when the setting indicates ON. denied when the setting indicates OFF. (2) The selection is confirmed by pressing either the or button. Setting Sensor Alarms This page allows modifying the preset alarm values of: LOW Alarm HIGH Alarm STEL Alarm TWA Alarm. ALTAIR® 5X PID 35 Operation Factory set alarm levels are shown in chapter 6.1 "Factory-set Alarm Thresholds and Setpoints". (1) Press the q button to enter Sensor Alarm setup. LOW Alarm Setup screen displays. Fig. 8 Sensor Alarm Setup (2) Press the button to abort the operation or the button to go to next alarm setup or the q button to change the alarm setpoints. Alarm Value for the first Sensor displays. US Fig. 9 Sensor Alarm Setup (3) Set values for Sensor Alarm by pressing the or button. (4) Press the q button to confirm set value. (5) Repeat setting for all other sensors. (6) Press the button to return to the Alarm Options menu. (7) Repeat setting for all other alarm types. ALTAIR® 5X PID 36 Operation Instrument Options SETTINGS The Instrument Options menu allows modification of different device options: Sensor Setup (enable/disable the channel) Language Setup Time Date Setup Datalog Intervals Stealth Mode Operating Beep Backlight Options VOC Gas Setup Bluetooth Press the button go to next page the button to go previous page the q button to enter setup. Setting Sensor Options (1) Press the q button to enter setup. Following screen displays: US Fig. 10 Sensor Options Setup (2) Press the button to select sensor, press the q button to make changes. The sensor information is displayed and the sensor can be enabled or disabled. Other operations such as changing the combustible gas type (Methane, Butane, Propane etc.) and units (ppm to mg/m3) are only possible using the MSA Link software. ALTAIR® 5X PID 37 Operation (3) Change status by pressing the or button. (4) Press the q button to confirm and advance to next screen (next sensor). (5) Perform the sequence for all other sensors. After setting up the last sensor the device goes to the next Setup Page. Language Setup This option is for setting the language of the device. (1) Press the q button to enter setup. (2) Change language by pressing the or button. (3) Confirm with the q button. The device goes to the next Setup Page. Time and Date Setup This option is for setting the device time and date. The device first prompts to set the time and then it prompts for the date. The time can be set up for either regular AM/PM or military time (through MSA Link soft- ware). AM/PM time is the default setting. (1) Press the q button to enter setup. (2) Change hours by pressing the or button. (3) Confirm with the q button. (4) Change minutes by pressing the or button. (5) Confirm with the q button. The device goes to the Set Date Page. (6) Change month, date and year by pressing the or button and confirming with the q button. The device goes to the next Setup Page. (7) Confirm with the q button. The device goes to the next Setup Page. Setting Stealth Mode Stealth mode disables the visual, audible and vibrating alarms. (1) Press the q button to change mode (ON/OFF). US (2) Press the button to go to the next page or the button to return to previous page. Setting Datalog Intervals This option is for setting the intervals at which all the readings will be logged. (1) Press the q button to enter setup. (2) Change interval by pressing the or button. (3) Confirm with the q button. The device goes to the next Setup Page. Setting Operating Beep (1) Press the q button to change mode (ON/OFF). (2) Press the button to go to the next page or the button to return to previous page. Setting Backlight (1) Press the q button to enter setup. Change option by pressing the or button. (2) Press the q button to enter. (3) Change timeout by pressing the or button. (4) Press q button to confirm timeout. ALTAIR® 5X PID 38 Operation PID Setup WARNING! It is very important to have an understanding of PID basics when changing PID settings. Failure to properly identify the VOC gas being measured and/or failures to select the correct Response Factor alarm values (exposure, STEL, TWA) that match the desired Response Factor and/or the correct lamp will result in erroneous readings or erroneous alarm limits that could cause death or serious personal injury. The PID sensor should be configured prior to initial operation. (1) Enter the correct password enter the 'Instrument Setup' menu and press the button until VOC Gas Setup is highlighted, then select OK. There are five configuration pages available: Configuration page Menu Enable On allows the VOC gas to be changed without entering the pass- word. When Menu Enable On is active, the "Change VOC Gas?" option is avail- Menu Enable able from the Main Measuring Page as described in section 3.4 "On-Screen Indicators". The default setting is ON. Maintain VOC Gas ON retains the currently selected VOC gas when the device is powered down and restarted. If this option is set to OFF the device will always Maintain VOC power up with isobutylene selected as the VOC gas type. This option should be Gas set to ON if the same VOC gas is to be monitored on every use. The default setting is ON. This set of pages allows the default Favorites list to be changed with VOC gases suitable for the user's particular environment. The ten default Favorites will be displayed on initial use. The first screen in the Favorites Setup page will ask which Favorite is to be replaced. (1) Select the gas to be replaced by using the or button to highlight then select OK. Favorites Setup The next screen will display the current 10 Favorites, and options for Custom Gas and All Gases. US (2) Highlight the gas to add to the Favorites and select OK. A Confirmation Screen will be displayed showing the gas to be replaced and the gas to be added to Favorites. Selecting YES will return to the Favorites list showing the new gas, selecting NO will return to the Favorites list showing the previous Favorites and selecting ABORT will return to the menu page. ALTAIR® 5X PID 39 Operation Configuration page This menu displays all of the gases available for detection by this PID sensor type. The gases are listed by an 8 character abbreviation. The full gas names are listed in chapter 10 "PID Response Factor Table" of this manual. he first 10 gases listed are the Favorites gases followed by options for the All Gas List and the Custom List. Gas names starting with the letters A-Z are listed alphabetically. Each page contains 14 gas names. (1) Select the gas of interest by using the or button to highlight then select OK. Holding down the or button for more than 2 seconds will scroll a full page at a time. VOC Gas Selec- Selecting OK will display a confirmation page that contains the following informa- tion tion: 8 character short name Response Factor (RF) for the selected gas Maximum Value of that gas (Full scale sensor value x RF). The Maximum Value is calculated by multiplying the sensor range by the RF. For example, the Max Value for Hexane is 2000 *4.5 = 9000ppm. The maximum value cannot exceed 9999ppm due to display resolution limitations. High Alarm - this reflects the current High Alarm value. Change if necessary for the chosen Response Factor Low Alarm - this reflects the current High Alarm value. Change if necessary for the chosen Response Factor The Custom Gas Setup allows a unique 8 character gas name and associated Response Factor to be entered for up to 10 Custom gases. (1) Select which Custom Gas (1 - 10) to enter or replace. a) Confirm with OK. (2) On the next screen, enter the 8 character gas name using the or Custom Gas buttons to select letters and numbers. Setup a) Select OK when the appropriate alpha-numeric characters are reached. (3) After the 8th character is entered, enter the Response Factor (0.1- 40.0). US Once the RF is complete, a final confirmation page will display. (4) Select OK to apply the Custom Gas as the current gas or select NO and abort to the menu page. WARNING! Failure to properly identify the VOC gas being measured and/or failures to select the correct alarm values (exposure, STEL, TWA) that match the desired Response Factor and/or the correct lamp will result in erroneous readings that could cause death or serious personal injury. ALTAIR® 5X PID 40 Operation Enabling Bluetooth The device is configured with a Bluetooth capable communications feature. (1) Press the q button to enable or disable the Bluetooth communications device (ON/OFF). (2) Press the button to to return to the Main Menu or the button to return to previous page. Back To Main Menu There are three options at this point: the button Sensor Options menu the button Previous Setup page in the Instrument Options menu the q button Instrument Options menu 4.6 Bluetooth Operation The Bluetooth communication device must be enabled for any Bluetooth functions to operate. See chapter 4.5 "Device Setup". Compatible Bluetooth host with appropriate software is required for proper operation. Bluetooth Security The Bluetooth connection is encrypted and secured with a unique six digit pin that must be double confirmed on both device and Bluetooth host at the time of pairing. Discovery Mode This device mode is used to enable a Bluetooth host to pair with the device for the first time or if a different Bluetooth host was connected with the device previously. Note that the device will automatically enter discovery mode for five minutes at device turn on if Bluetooth has been enabled. Discovery mode will also be entered for 5 minutes following a disconnection. To manually enter discovery mode: (1) Page down through the menu pages in Measurement Mode using the button until the Discovery Mode page is displayed. (2) Press the q button to enter discovery mode. The blue led will blink rapidly indicating that the device is in Discovery Mode. Connecting the device to a Bluetooth host for the first time US (1) Ensure that the device is on and in Discovery Mode (2) On the Bluetooth host, locate the Bluetooth device list. Select "A5X-xxxxxxxx" from the list. Both the device and Bluetooth host will display a unique six digit security code to ensure that the correct devices are being paired. (3) After confirming that the six digit codes match, confirm the pairing request on the device by pressing the button. (4) Confirm on the Bluetooth host as well. Connecting the Device to a Bluetooth Host If this was the last device connected to the Bluetooth host, the Bluetooth host can connect to the device whether or not the device is in discovery mode as long as Bluetooth is enabled. The six digit code confirmation will not be displayed. The device will only recall the last Bluetooth host it was paired with. If connecting to another Bluetooth host, the device must be placed into discovery mode to be detected. ALTAIR® 5X PID 41 Operation Pairing the device to a Bluetooth host This device has an integrated RFID chip to facilitate a faster Bluetooth pairing process with a Bluetooth host that supports a RFID or NFC reader with appropriate software. Simply align the RFID or NFC reader of the Bluetooth host directly over the MSA logo on the front of the device.The device and Blue- tooth host should become paired and connected. Disconnecting the Device from a Bluetooth Host The device does not have a disconnect feature as this would be initiated by the Bluetooth host. Use the Bluetooth host functions to purposefully disconnect the device from the Bluetooth host. Device Configuration over Bluetooth Connection The device has the ability to receive updates to device settings over the Bluetooth connection. The user must successfully pair the device and Bluetooth host confirming that the six digit security code matches both on the device and the Bluetooth host. After a configuration change has been initiated, the user must confirm the request on the device by pressing the button. Evacuation Alert over Bluetooth Connection The device has the ability to receive an evacuate message over the Bluetooth connection. The user must successfully pair the device and Bluetooth host confirming that the six digit security code matches both on the device and the Bluetooth host. Once connected, an evacuate message sent to the device will send the device into alarm while displaying EVAC on the display. Press the button to silence the Evacuation alert and confirm the alert was received. Press the button a second time to reset the Evacuation alert once in a safe area. 4.7 MSA Link Operation Connecting device to PC (1) Switch ON the device and align the Datalink Communication port on the device to the IR inter- face of the PC. (2) Start the MSA Link software on the PC and start the connection by clicking the connect icon. 4.8 Function Tests on the Device Alarm Test (1) Turn ON the device. The user should verify that: alarm LEDs flash horn sounds briefly US vibrating alarm triggers briefly. 4.9 Bump Test WARNING! Perform a Bump Test before each day’s use to verify proper device operation. Failure to perform this test can result in serious personal injury or death. Bump test frequency is often stipulated by national or corporate regulations; however, bump testing before each day’s use is generally the accepted best safety practice and is therefore MSA’s recommendation. This test quickly confirms that the gas sensors are functioning. Perform a full calibration periodically to ensure accuracy and immediately if the device fails the Bump Test. The Bump Test can be performed using the procedure below or automatically using the GALAXY GX2 Test Stand. ALTAIR® 5X PID 42 Operation CSA requires (per 22.2 NO. 152) that combustible sensor sensitivity be tested before each day's use on a known concentration of methane equivalent to 25 to 50 % of full scale concentration. ACCURACY MUST BE WITHIN 0 to +20 % OF ACTUAL. Correct accuracy by performing the calibration procedure described in chapter 4.10 "Calibration". NOTE: The GALAXY GX2 cannot test the Chlorine Dioxide sensor. For this sensor, use this Bump Test procedure and/or manually calibrate as described in chapter 4.10 "Calibration". Equipment See accessory chapter for ordering information for these components. Calibration Check Gas Cylinder(s) See chapter 6.3 "Calibration Specifications" for calibration gas target values and appropriate MSA calibration gas cylinders. Demand Flow Regulator(s) Tubing appropriate for the gases to be tested Kits containing tubing and regulators suitable for reactive and non-reactive gases are available from MSA. Performing a Bump Test (1) While the device is turned ON in a clean, fresh air environment, verify that readings indicate no gas is present. (2) From the normal measure screen press the button to display “BUMP TEST?”. (3) Verify the gas concentrations displayed match the Calibration Check Gas Cylinder. If they do not, adjust the values through the Calibration Setup menu. Depending on the sensors installed, there could be one to five separate Bump Tests performed, each with a different cylinder, regulator, and tubing used. (4) Attach the demand regulator (supplied in the calibration kit) to the cylinder providing the indi- cated gases. (5) Connect tubing (supplied in the calibration kit) to the regulator. (6) Attach the other end of tubing to the device pump inlet. (7) Press the q button to start the bump test: the progress bar advances the sensors respond to the gas. The message BUMP TEST PASS indicates a successful Bump Test of the sensors. If any sensor fails the Bump Test: US the message BUMP TEST FAIL appears the failed sensor is indicated. If there are more sensors to be Bump Tested, the next sensor displays and the process repeats from Step 4. If there are no more sensors to be Bump Tested, the tubing can be removed from the device pump inlet. ALTAIR® 5X PID 43 Operation After the Bump Test After all installed sensors pass the Bump Test, the √ symbol displays on the MEASURE page. This √ symbol appears on the display in the upper feature bar If any sensor was not bump tested, or fails the Bump Test, the √ symbol does not display. The color display: temporarily shows the √ symbol at each gas reading for successfully bump tested sensors √ symbol is then replaced by the present gas reading. The √ symbol shows for 24 hours after the Bump Test. If a sensor fails the Bump Test, calibrate the device as described in chapter 4.10 "Calibration". 4.10 Calibration The ALTAIR 5X PID can be calibrated either manually using this procedure or automatically using the GALAXY GX2 test stand. Refer to chapter 11.5 "Calibration Options". The use of the demand regulators listed in chapter 9 "Ordering Information" is recommended. If a new sensor has been installed, the battery pack has been depleted or a new battery pack has been installed allow sensors to stabilize for 30 minutes before calibration is performed. WARNING! Special conditions with toxic gases! If the device is to be checked or calibrated for reactive gases, prerequisites are required; otherwise, incorrect calibration would result in incorrect device operation. Reactive toxic gases (e.g., chlorine, ammonia, chlorine dioxide) have the property of diffusing into the rubber and plastic tubes so that the volume of test gas available in the device would no longer be suffi- cient to correctly perform device calibration. When calibrating the device with toxic gases, certain prerequisites are required, otherwise incorrect calibration could result: A special pressure regulator Shortest possible connection tubes between the pressure regulator and the device Connection tubes made from a material that does not absorb the test gases (e.g., PTFE). NOTE: If using normal tubes and pressure regulators, expose them to the required test gas for an US extended time period. Keep these materials dedicated for use with that test gas only; do not use them for other gases. For example, for chlorine, allow the entire contents of a test gas cylinder to flow through the pres- sure regulator and tubes before using to calibrate the device. Mark these materials for use with chlorine only. ALTAIR® 5X PID 44 Operation Zero Calibration Procedure (1) Press the button for five seconds in Normal Measurement page. ZERO screen displays. To skip the ZERO procedure and move directly to the span calibration proce- dure, push the button. If no button is pushed for 30 seconds, the device prompts user to perform a SPAN calibration before device returns to the Normal Measurement page. To perform ONLY a Fresh Air Setup at this time, press the q button. The device then performs a Fresh Air Setup as described in chapter 4.2 "Turning ON and Fresh Air Setup". When the Fresh Air Setup is complete, the device returns to the normal Measure screen. (2) Press the button to confirm the ZERO screen, i. e. to execute zero calibration. The message "SENSOR REFRESH" displays, followed by the message "ZERO CALIBRATION". The "REFRESH" message does not appear if a catalytic combus- tible sensor is not installed. ZERO calibration starts. A progress bar shows the user how much of the calibration has been completed. During the first moments of a ZERO calibration, the combustible sensor reading may be replaced by a moving display of “PLEASE WAIT”. This is normal. After the ZERO calibration is completed the device displays either “ZERO CALIBRATION PASS” or “ZERO CALIBRATION FAIL”. US Only if the device passes the zero calibration the SPAN screen displays. Span Calibration To skip the Span calibration procedure, push the button. If the SPAN calibration of the combustible sensor is skipped after a successful ZERO cali- bration, the combustible sensor reading may be replaced with a moving display of “PLEASE WAIT” for a few moments. This is normal, and the device is fully operational once a combustible gas reading reappears. ALTAIR® 5X PID 45 Operation If no button is pushed for 30 seconds, span calibration is skipped. Because of the different possible combinations of gases that are possible, skipping a Span calibration could advance the user to the Span calibration of another installed sensor, or back to Measuring mode. When calibrating a 0-2000 ppm PID sensor, use 100 ppm isobutylene balanced in air. Isobutylene calibration gas other than 100 ppm is not allowed due to long term affect on the CO sensor. (1) Connect one end of tubing to the cylinder regulator (supplied in the calibration kit). (2) Connect the other end of the tubing to the pump inlet. (3) Press the button to calibrate (span) the device. “SPAN CALIBRATION“ flashes SPAN calibration starts. A progress bar shows the user how much of the calibration has already been completed. After the SPAN calibration is completed, the device displays either “SPAN CALIBRATION PASS” or “SPAN CALIBRATION FAIL” The device returns to Measuring mode. US ALTAIR® 5X PID 46 Operation If a sensor is nearing its end-of-life, the "PASS" display is followed by the Sensor Life indicator ♥ display. While the sensor is still fully functional, this warning gives the user time to plan for a replacement sensor to minimize downtime. The ♥ indicator blinks as the device returns to Measure mode. After 15 seconds, the blinking stops, but the ♥ indicator continues to display during ongoing oper- ations as a reminder of a sensor’s pending end-of-life. If a span calibration fails: The Sensor Life Indicator ♥ blinks to show a sensor has reached its end-of-life and should be replaced. The device remains in the Sensor Life alarm condition until the button is pressed. After the alarm is cleared, the device enters Measure mode and the Sensor Life indicator ♥ blinks during ongoing operations until the sensor is replaced and/or successfully calibrated. Span calibration can fail for reasons other than a sensor at the end of its life. If a span calibration failure occurs, verify items such as: sufficient gas remaining in the calibration cylinder gas expiration date integrity of calibration tubing/fittings, etc. Reattempt the span calibration before replacing the sensor. Finishing Successful Calibration (1) Remove the calibration tube from pump inlet. The calibration procedure adjusts the span value for any sensor that passes the calibration test. Sensors that fail calibration are left unchanged. On the color display, each successfully calibrated sensor temporarily shows a √ symbol at its gas reading. These √ symbols remain visible for a few moments and are then replaced by the present gas reading. Since residual gas may be present, the device may briefly go into an exposure alarm after the calibra- tion sequence is completed. (2) Press the button to reset the alarm as necessary. Following a PID sensor calibration, VOC gas readings may be slightly elevated (< 5 ppm) for several US minutes. This is normal behavior as the isobutylene is purged from the device. A √ symbol displays on the MEASURE page.This √ symbol appears on the: color display in the upper feature bar The √ symbol displays for 24 hours after the calibration and then turns off. If the horn alarm is turned OFF, the calibration √ symbol does not appear on the color display. ALTAIR® 5X PID 47 Operation Calibration with an Automated Test System The device can be calibrated using the GALAXY GX2 Automated Test System - contact MSA for a list of compatible gases and concentrations. Similar to the successful (manual) calibration described in chapter "Finishing Successful Calibration", a √ symbol displays on the MEASURE page after successful GALAXY GX2 calibration. This √ symbol appears on the display in the upper feature bar. The √ symbol displays for 24 hours after the calibration and then turns off. If the horn alarm is turned OFF, the calibration √ symbol does not appear on the color display. 4.11 Time of Day Testing This feature permits the device to be automatically calibrated on a user-defined interval. The most common use of this feature allows the user to configure the ALTAIR 5X PID and the GALAXY GX2 System to automatically calibrate a device prior to the start of work-shift. See the GALAXY GX2 Oper- ating Manual ("Automated Testing Features" section) for a complete description of how to configure the GALAXY GX2 for this mode. On ALTAIR 5X PID devices, the following settings must be configured using either MSA Link or the GALAXY GX2 → Instrument Setup page: For automated calibration testing, Calibration Due must be enabled and a non-zero Calibration Interval must be entered for automatic calibration testing For automated bump testing, Bump Due must be enabled and a non-zero Bump Interval must be entered for automatic bump testing Carefully follow all GALAXY GX2 set-up directions as described in the GALAXY GX2 Operating Manual for proper setup. 4.12 Device Shutdown For device shutdown press and hold the q button. US Fig. 11 Shutdown The device displays a blinking "HOLD BUTTON FOR SHUTDOWN" and a progress bar shows the user how much longer to hold the button to complete the shutdown. ALTAIR® 5X PID 48 Maintenance 5 Maintenance If irregularities occur during operation, use the displayed error codes and messages to determine appropriate next steps. WARNING! Repair or alteration of the device beyond the procedures described in this manual or by anyone other than a person authorized by MSA, could cause the unit to fail to perform properly. Use only genuine MSA replacement parts when performing any maintenance procedures described in this manual. Substitution or incorrect installation of components can seriously impair performance of the unit, alter intrinsic safety characteristics or void agency approvals. Failure to follow this warning can result in serious personal injury or death. Refer to EN 60079-29-2 (Guide for the selection, installation, use and maintenance of apparatus for the detection and measurement of combustible gases or oxygen) and EN 45544-4 (Guide for the selection, installation, use and maintenance of electrical apparatus used for the direct detection and direct concentration measurement of toxic gases and vapors). US ALTAIR® 5X PID 49 Maintenance 5.1 Troubleshooting Error State Details Recommended Action Alternating display ADC ERROR Analogue measurement error Contact MSA MEM ERROR Memory error Contact MSA PROG ERROR Program error Contact MSA RAM ERROR RAM error Contact MSA BT ERROR Bluetooth error Contact MSA LOW BATTERY Battery Warning repeats Remove from service as soon as possible and every 30 seconds recharge or replace battery (flashing) Device is no longer sensing gas; Remove from BATTERY ALARM Battery is completely discharged service and recharge or replace battery. Remove from service as soon as possible and Device does not turn ON Battery fully discharged recharge or replace battery pack. SENSOR MISSING Sensor damaged or missing Replace sensor Device must have at least one sensor enabled at NO SENSORS No sensors are enabled all times Sensor warning Sensor is near the end of its life Sensor has reached the end of its life and cannot Sensor alarm be calibrated. Replace sensor and recalibrate. (flashing) Pump malfunction or flow path Check flowpath for blockage. If error persists, PUMP ERROR blockage remove from service. INVALID CONFIGURA- Sensor(s) installed in incorrect Install sensors as shown in Figure Fig. 13 location. "Possible positions for sensor replacement". US TION 5.2 Verifying Pump Operation Users can check operation of the sampling system any time during operation by blocking the sampling system to generate a pump alarm. When the pump inlet, sample line or probe is blocked, the pump alarm must activate. Once gas readings are displayed, plug the free end of the sampling line or probe. The pump motor shuts down and an alarm sounds. PUMP ERROR will flash on the display. (1) Press the button to reset the alarm and restart the pump. If the alarm does not activate: Check the sample line and probe for leaks. Once leak is fixed, recheck pump alarm by blocking the flow. (2) Press the button to reset the alarm and restart the pump. ALTAIR® 5X PID 50 Maintenance WARNING! Do not use the device, sample line, or probe unless the pump alarm activates when the flow is blocked. Lack of an alarm is an indication that a sample may not be drawn to the sensors, which could cause inaccurate readings. If a sample line or probe is installed and the pump alarm does not activate, remove the line or probe and repeat the test. This will provide information on where the blockage is located.Failure to follow the above can result in serious personal injury or death.Never let the end of the sampling line touch or go under any liquid surface. If liquid is drawn into the device, readings will be inaccurate and device could be damaged. We recommend the use of an MSA sample probe containing a special membrane filter, permeable to gas but impermeable to water, to prevent such an occurrence. During operation, a pump alarm may occur when the: Flow system is blocked Pump is inoperative Sample lines are attached or removed. To Clear Pump Alarm (1) Correct any flow blockage. (2) Press the button. The Pump will now restart. 5.3 Replacing the Battery WARNING! Never replace the battery in a hazardous area. This could result in an explosion. 1 US 2 Fig. 12 Battery Replacement 1 Captive screw 2 Battery pack ALTAIR® 5X PID 51 Maintenance (1) Unscrew the two captive screws on the rear of the device. (2) Pull the battery pack out of the device by gripping the sides and lifting it up and away from the device. (3) When replacing the battery, be sure to accurately align screws and battery with housing. (4) Screws should be tightened and torqued to 5.5 in lb. 5.4 Maintenance Procedure - Replacing or Adding a Sensor Any factory-installed Series 20 sensor may be removed or replaced with a like type. Any XCell sensor may be removed or replaced according to the positions allowed in table below Figure Fig. 13 "Possible positions for sensor replacement". WARNING! The PID sensor may be replaced with an authorized MSA PID replacement sensor. If the PID sensor range is different than the previously installed sensor, the new configuration MUST be selected in the Instrument Options → Sensor Setup screen prior to sensor replacement. Otherwise incorrect readings could occur and persons relying on this product for their safety could sustain serious personal injury or death. WARNING! Remove and reinstall sensors carefully, ensuring that the components are not damaged; otherwise device intrinsic safety may be adversely affected, wrong readings could occur, and persons relying on this product for their safety could sustain serious personal injury or death. NOTICE Before handling the PC board, ensure you are properly grounded; otherwise, static charges from your body could damage the electronics. Such damage is not covered by the warranty. Grounding straps and kits are available from electronics suppliers. While device case is open, do not touch any internal components with metallic/conduc- US tive objects or tools.Damage to the device can occur. ALTAIR® 5X PID 52 Maintenance 4 2 3 5 1 Fig. 13 Possible positions for sensor replacement 1 Combustible sensor 4 See table below or a sensor plug 2 O2 sensor 5 PID sensor 3 See table below SENSOR OPERATIONAL ONLY IN POSITION XCell combustible sensor 1 XCell O2 sensor 2 XCell SO2, Cl2, NH3, H2S, H2S-LC sensor 4 XCell Two-Tox, SO2, Cl2, NH3, CO-HC 3 Series 20 sensor 3 US PID sensors 5 (1) Verify that the device is turned OFF. (2) Remove the battery pack. (3) Remove the two remaining case screws, and remove the case front. (4) Gently remove the sensor to be replaced. (5) Carefully align the new sensor contact pins with the sockets on the printed circuit board. (6) Press the new sensor into place. (7) Note the position restrictions in the table above. Adapter (part no. 10110183) is required for XCell usage in position 3. If a sensor is removed and will not be replaced, be sure to install a sensor plug in its place in order to maintain correct device function. The plug for XCell positions is P/N 10105650.The Series 20 plug is P/N 10088192. (8) Visually inspect the green gasket, assuring that it is seated properly in the front housing. ALTAIR® 5X PID 53 Maintenance (9) Attach front case and tighten two case screws using 5.5 in-lbs of torque. (10) Attach the battery pack and tighten the two battery pack screws using 5.5 in-lbs of torque. If a change in XCell Sensor configuration is detected during the device turn-ON process: The "ACCEPT?" prompt appears on the display The button accepts the sensor configuration The button rejects the sensor configuration; the device is not operational. When an XCell sensor is replaced, the device automatically enables the sensor after the change has been accepted. If a Series 20 or PID sensor is replaced, it must be manually enabled (→ chapter 4.10 "Calibration", SETTING SENSOR OPTIONS). If the oxygen sensor was replaced, see chapter 4.3 "Special Consideration for Oxygen Sensor" regarding the oxygen reading display. (11) Allow sensors to stabilize at least 30 minutes before calibration. (12) Calibrate device before use. WARNING! Calibration is required after a sensor is installed; otherwise, the device will not perform as expected and persons relying on this product for their safety could sustain serious personal injury or death. 5.5 Replacing the Pump Filter (1) Turn OFF the device. (2) Unscrew the two captive screws from the clear filter cover on the back of the device to access the filter. (3) Carefully lift out the O-ring and the filter disk(s). (4) Use both the paper-like filter and the fibrous dust filter (the thicker disk) as supplied in the Main- tenance Kit if the device is NOT configured to use a reactive toxic gas sensor (does not have a Cl2, ClO2, or NH3 sensor). Use ONLY the paper filter supplied in the Reactive Gas Maintenance Kit if the device IS config- ured to use a reactive toxic gas sensor (Cl2, ClO2, or NH3). (5) Place the new paper-like filter into the recess in the back of the device. If it is to be used, place the fibrous dust filter into the clear filter cover. WARNING! US Use of the fibrous dust filter or the incorrect paper filter for the measurement of reactive gases could cause erroneous readings. (6) Replace the O-ring in the recess. (7) Re-install the clear filter cover on the back of the device. 5.6 Cleaning the Device Exterior Clean the exterior of the device regularly using only a damp cloth. Do not use cleaning agents as many contain silicones which will damage the combustible sensor. 5.7 Storage When not in use, store the device in a safe, dry place between 18 °C (65 °F) and 30 °C (86 °F). After