Summary

This document provides a summary and overview of ultrasound technology. It details its history, types, and applications, including various methods of scanning and analysis. The study of ultrasound and its use in medical imaging is well-covered.

Full Transcript

HISTORY OF ULTRASOUND INFRASOUND  1918 - Sound Navigation and  Below 16 Hz Ranging was used  Early 1950’s –Water bath ULTRASOUND GENERATORS immersion technique  Late 1950’s – First contact TYPES OF ULTRASOUND compound B-scann...

HISTORY OF ULTRASOUND INFRASOUND  1918 - Sound Navigation and  Below 16 Hz Ranging was used  Early 1950’s –Water bath ULTRASOUND GENERATORS immersion technique  Late 1950’s – First contact TYPES OF ULTRASOUND compound B-scanner WAVES  1970’s – Gray scale imaging  Longitudinal/Compression  Mid 1970’s – Real time Waves scanning systems  Transverse/Shear Waves  1980’s – Doppler technique  Surface/Rayleigh Waves SOUND ACOUSTIC VARIABLES  A mechanical energy  Period (T): the time taken for  Requires a vibrating object to one complete cycle to occur (s produced or µs)  Cannot travel through a  Wavelength (λ): length of vacuum space over which one cycle occurs (m or mm) ULTRASOUND  Amplitude (Depth): the  High frequency sound waves maximum displacement that  Above 20,000 cycles per occurs in an acoustic variable second (20 kHz)  Frequency: cycle per second  Inaudible to humans (Hz)  Used to scan tissues of the body  Velocity: frequency times  Ultrasound Pulse: 2-10 MHz wavelength  Pulse Duration: 1 microsecond  Pulse Repetition: 1000 PIEZOELECTRIC CRYSTALS times/second  Generate ultrasound waves  Capable of changing electrical AUDIBLE SOUND signals into mechanical  16 Hz to 20,000 Hz (ultrasound) waves PIEZOELECTRIC CRYSTAL AS REAL-TIME TRANSMITTER OF SOUND  Shows movement as it occurs  Converting electrical energy into mechanical energy (sound) M-MODE  Shows movement as a function PIEZOELECTRIC CRYSTAL AS of time TRANSMITTER OF SOUND  Used in cardiac scanning  Converting mechanical energy (sound) into electrical energy DOPPLER ULTRASOUND  Demonstrates and measures NOTE: blood flow  Small crystal diameter o Increased beam DOPPLER EFFECT divergence  The change in apparent  Larger crystal diameter frequency of a wave as a result o Decreased beam of relative motion between the divergence observer and the source  Stationary Reflector: reflected DIFFERENT MODES OF echoes are the same as the ULTRASOUND transmitted waves  Reflector that Moves Closer: A-MODE reflected echoes are higher than  Echoes are shows as peaks the transmitted echoes  Distance between various  Reflector that Moves Away: structures can be measured reflected echoes are lower than  Used to build two-dimensional the transmitted echoes B-mode image BASIC TYPES OF DOPPLER B-MODE ULTRASOUND UNIT  Two-dimensional images in which the echo amplitude is 1.) Continuous Wave Doppler Unit depicted as dots of different  Ultrasound is continuous brightness  Measures high velocities accurately  No depth resolution WAVELENGTH  The length of a single cycle of 2.) Pulsed Wave Doppler Unit the ultrasound wave  Ultrasound is transmitted in  Inversely proportional to the pulses frequency Single element  With good depth resolution  Determines the resolution of  Measures the speed of the the scanner blood in a particular vessel  Higher the frequency, the  Cannot measure high blood shorter the wavelength velocities in deep vessels  High velocities may be wrongly FOCUSING displayed as low velocities  Adjustment of the ultrasound beam 3.) Colour Doppler Unit  To improve resolution  Shows different flow-velocities  May be electronic or by a lens in different colours attached to the transducer 4.) Duplex Doppler System AMPLIFICATION  Combination of a B-mode and  Done by the time-gain- Doppler system compensation (TGC) amplier  Allows the Doppler beam to be  To compensate for ultrasound directed accurately at any attenuation in any part of the particular blood vessels body  To improve the quality of the WAVE PROPAGATION final image  The transmission and spread of ultrasound waves to different BOUNDARIES tissues  The line at the periphery of two  Average Propagation for Soft tissues which propagate Tissues: 1540 m/s ultrasound differently  Average Propagation for Soft  The zone of echoes at the Tissues: 4620 m/s interface PIEZOELECTRIC EFFECT  Low density substance – low  Piezein – “press or pressure” acoustic impedance  Ability of a material to generate  Formula: Z=pc an electrical charge un response o p = density of material to applied pressure (kg/m3) o c = speed of sound (m/s) PIEZOELECTRIC MATERIALS o Z = acoustic impedance  Crystalline materials composed (rayls) of dipolar molecules  Quartz – naturally occurring SUBSTANCE Z SPEED crystals (m/s)  Lead zirconate titanate – man Air 0.0004 330 Fat 1.38 1450 made ceramic Water 1.48 1480  Natural Materials: Blood 1.61 1570 o Quartz Kidney 1.62 1560 o Tourmaline Soft Tissue 1.63 1540 o Rochelle Salt Liver 1.65 1550  Synthetic Materials: Muscle 1.70 1580 o Lead zirconate titanate Bone 7.80 3500 PZT (crystal) 30 3870 (PZT) o Barium titanate ACOUSTIC IMPEDANCE AND o Lead metaniobate REFLECTION o Ammonium dihydrogen  Substances with same phosphate acoustic impedance: o Lithium sulphate o 100% energy transmission ACOUSTIC IMPEDANCE o No reflection  Property of a substance  Substances with a small  Describes how the particles of difference in acoustic that substance behave when impedance: subjected to pressure wave o 95% energy transmission  High density substance – high o 5% reflection acoustic impedance  Substances with a large  Best suited to image deep lying difference in acoustic structures impedance:  3.5 MHz o 1% energy transmission o 99% reflection 3.) Convex Transducer  Wide fan-shaped TRANSDUCER/PROBE  Useful for all parts of the body  A device which converts one  Except for specialized form of energy to another echocardiography  Converts electrical energy into ultrasound waves and vice 4.) Phased Array Transducer versa  Flat faced transducer  Contains piezoelectric crystals  Wide field of view o Transmit ultrasound  useful in cardiac and cranial beam ultrasound o Receive reflected echoes COMPONENTS AND TRANSDUCERS/SCANNING CONSTUCTION OF A TYPICAL PROBES TRANSDUCER  The most expensive part of any ultrasound unit 1.) PHYSICAL HOUSING  Contains all individual 1.) Linear Array Transducer components  Parallel scan lines  Provides the necessary  Rectangular field of view structural support  Vascular, small parts and  Acts as an electrical and musculoskeletal applications acoustic insulator  Above 4 MHz 2.) ELECTRICAL 2.) Sector/Curvilinear Array CONNECTIONS Transducer  Formed in front and back of the  Provides wide field of view crystal  Most useful in abdominal and  Made of thin film of gold or obstetric scanning silver 3.) PIEZOELECTRIC  Controls the length of ELEMENTS vibrations from the front face  Crystalline minerals that  Improves axial resolution generate voltages when  Materials: subjected to a mechanical force o Plastic or epoxy resin  Piezein – “to press or squeeze” o Cork  Piezoelectric Effect – o Rubber discovered by Jacques and o Araldite loaded with Pierre Curie tungsten powder  Thinner Piezoelectric Materials o Higher resonant 5.) ACOUSTIC LENS frequencies  Reduce the beam width of the transducer FREQUENCY  Improve image resolution  Affects the quality the  Width of the Beam: ultrasound image determines lateral resolution  Higher Frequency  Lateral Resolution: the ability o Shorter wavelength to resolve structure across or o Better Resolution perpendicular to the beam axis o Lower Penetration  Materials: o Higher Absorption o Aluminum  Lower Frequency o Perspex o Longer wavelength o Polystyrene o Poor Resolution o Higher Penetration 6.) IMPEDANCE MATCHING o Lower Absorption LAYER o Sandwich between the 4.) BACKING/DAMPING piezoelectric crystal and the MATERIALS patient  Shortens the ultrasound pulse o Chosen to improved length transmission into the body  Eliminates the vibrations from the back face BANDWIDTH 3.) Pediatric Ultrasound o The range of frequencies  5.0 MHz transducer: for contained within an ultrasound children pulse  Focused at 5-7 cm o Wide Bandwidth:  Sector transducer of 7 MHz: o Shorter spatial pulse o Neonatal brain scans length o For adult testis and neck o Wider range of  Focused at 4-5 cm frequency o Narrow Bandwidth: ULTRASOUND BEAM o Longer spatial pulse  Area through which the sound length energy emitted from the o Narrower range of ultrasound transducer frequency  Three dimensional and symmetrical around its central CHOOSING THE APPOPRIATE axis TRANSDUCER TWO REGIONS OF 1.) Obstetric Ultrasound ULTRASOUND BEAM  Linear or convex transducer 1.) Near Field/Fresnel zone  3.5 MHz: better in later 2.) Far Field/Fraunhofer zone pregnancy  Increasing Frequency  5.0 MHz: best during early o Longer near field pregnancy o Less far field divergence  Focused at 7-9 cm  Narrow Crystal Diameter o Narrower near field o More far field divergence 2.) General Purpose Ultrasound  Thin Crystal  Sector or convex transducer o Decreased near field  3.5 MHz o Increased far field  Focused at 7-9 cm  Thick Crystal o Increased near field o Decreased far field BEAM INTENSITY SPATIAL RESOLUTION  The power (measured in watts)  Detail Resolution flowing through a unit area  The ability to display two structures situated close SIDE LOBES/GRATING LOBES together as separate images  Lobes at various angles to the  Higher Frequency: main beam o Better resolution  Approximately 15% of the o Lower penetrability energy in the beam o Higher absorption  Cause a degradation of lateral  Lower Frequency: resolution o Poor resolution o Higher penetrability BEAM WIDTH o Lower absorption  The dimension of the beam in the scan plane TWO COMPONENTS OF  Affects the spatial resolution SPATIAL RESOLUTION  Narrow Beam Width o Better spatial resolution 1.) AXIAL RESOLUTION  Longitudinal, Linear, Depth or SLICE THICKNESS Range  Three dimensional volume  The ability to distinguish two displayed as a two dimensional objects parallel to the image ultrasound beam  Depends upon the spatial pulse RESOLUTION length and wavelength  The ability of an imaging  Short Spatial Pulse Length: system to differentiate between good axial resolution structures  Longer Spatial Pulse Length:  Spatial Resolution: resolution poor axial resolution in space  Contrast Resolution: 2.) LATERAL RESOLUTION resolution of gray shades  Azimuthal, Transverse,  Temporal Resolution: Angular or Horizontal resolution in time  The ability to distinguish two waves as they pass through objects perpendicular to the tissues ultrasound beam  Unit: decibels per centimeter  Depends upon the beam diameter FIVE MAIN PROCESSES THAT  Smaller Beam Width: better CAUSE ATTENUATIONS lateral resolution  Larger Beam Width: poor 1.) ABSORPTION lateral resolution  Occurs when ultrasound energy is lost to tissues by its CONTRAST RESOLUTION conversion to heat  The ability of the imaging  Main factor causing attenuation system to differentiate between  Higher Frequency: body tissue and display them as o Greater amount of different shades of gray absorption  Optimized by using the correct  Bone: higher absorption overall gain coefficient  Increasing protein content gives TEMPORAL RESOLUTION increasing absorption  Frame Rate o Blood –> Fat –> Nerve –  The ability of the imaging > Muscle –> Skin –> system to display events which Tendon occurs at different times as –> Cartilage –> Bone separated images  Best Absorption: tendon,  Higher Frame Rate: better ligament, fascia, joint capsule temporal resolution & scar tissue ULTRASOUND INTERACTIONS 2.) REFLECTION AND ATTENUATIONS  Occurs when two large structure of significantly ATTENUATION different acoustic impedance  Decrease in the intensity and form an interface amplitude of the ultrasound  Occurs when a sound wave 1.) REVERBERATION strikes an object that is larger  Comet tail than the wavelength  The production of spurious or false echoes due to repeated 3.) SCATTERING reflections between two  Occurs when an ultrasound interfaces with a high acoustic wave strikes a boundary or impedance mismatch interface between two small  The presence of two or more structures strong reflecting surfaces  Occurs when a sound wave  Often occur at: strikes an object that is equal to o Skin-transducer interface or smaller than the wavelength o Gas surface and transducer 4.) REFRACTION  Prevention/Elimination:  Occurs when the beam o Increase the amount of encounters an interface between gel used two different tissues at an o Used a stand-off gel pad oblique angle o Reduce the gain  The beam will be deviated as it o Move the position of the travels through the tissue transducer  Occurs due to difference in wave velocity across an 2.) ACOUSTIC SHADOWING interface between two materials  Caused by highly attenuating structure 5.) DIVERGENCE  Often occur at:  Occurs when the beam travels o Soft tissue and gas through tissue and it will o Soft tissue and bone or diverge due to diffraction calculus effects o Calcified mass ULTRASOUND ARTIFACTS 3.) ACOUSTIC ENHANCEMENT  A structure in an image which  Caused by weakly attenuating does not directly correlate with structures actual tissue being scanned  Often occur at: have arisen form the central o Distal to fluid-filled axis of the main lobe urinary bladder,  Appearance can give rise to a gallbladder or cyst false diagnosis o Fluid-filled mass  Inherent characteristic of the transducer 4.) EDGE SHADOWING  Combination of refraction and 8.) MIRROR IMAGE ARTIFACT reflection occurring at the  Caused by specular reflection edges of rounded structures of the beam at a large smooth interface 5.) BEAM WIDTH ARTIFACT  Often seen in:  Variations of all echoes o Fluid-air interface returning to the transducer o Diaphragm  Prevention/Elimination: o Correct positioning of 9.) DOUBLE IMAGE ARTIFACT the focal zone  Caused by refraction of the beam 6.) SLICE THICKNESS  Often occur at: ARTIFACT o Rectus abdominis muscle  Occurs due to the thickness of  Prevention/Elimination: the beam o Move the transducer  Dependent upon beam slightly to one side to angulation avoid the junction of  Often seen in: rectus abdominis muscle o Transverse view of the urinary bladder 10.) EQUIPMENT-GENERATED  Inherent characteristic of the ARTIFACT transducer  Caused by incorrect use of the equipment control 7.) SIDE LOBE ARTIFACT  Echoes generated by side lobes “Passing the Board Exam is your assumed by the transducer to main purpose to learn” 05/13/14

Use Quizgecko on...
Browser
Browser