What is the applicable temperature range for calibration when it is being performed on-site?
What is the applicable temperature range for calibration when it is being performed on-site?
What is the acceptable range for Relative Humidity during calibration?
What is the acceptable range for Relative Humidity during calibration?
In what type of calibration is the PAB logo affixed in the calibration certificate?
In what type of calibration is the PAB logo affixed in the calibration certificate?
When would Limited Calibration be carried out?
When would Limited Calibration be carried out?
Signup and view all the answers
What happens when a unit undergoes Conditional Calibration?
What happens when a unit undergoes Conditional Calibration?
Signup and view all the answers
What is the purpose of the procedure mentioned in the text?
What is the purpose of the procedure mentioned in the text?
Signup and view all the answers
What is the responsibility of all personnel involved in calibration?
What is the responsibility of all personnel involved in calibration?
Signup and view all the answers
What does CMTS stand for in the context of the text?
What does CMTS stand for in the context of the text?
Signup and view all the answers
What does OOT stand for based on the terms provided in the text?
What does OOT stand for based on the terms provided in the text?
Signup and view all the answers
Which document provides the general requirements for the competency of testing and calibration laboratories?
Which document provides the general requirements for the competency of testing and calibration laboratories?
Signup and view all the answers
What is the acceptable range for temperature maintenance inside the Calibration Laboratory according to the procedure?
What is the acceptable range for temperature maintenance inside the Calibration Laboratory according to the procedure?
Signup and view all the answers
Study Notes
Environmental Monitoring
- The laboratory manager is responsible for maintaining the environmental condition of the CMTS calibration laboratory.
- The applicable environmental condition for on-site calibration must be: • 25 ± 10 °C (15°C to 35°C) for Temperature • 55 ± 25 % (30 % to 80%) for Relative Humidity
- CMTS Lab personnel must verify the suitability of the environment before proceeding with calibration.
- If the temperature or relative humidity is out of the limit, necessary corrective action must be initiated, and calibration activities will be suspended.
Type of Calibration Service
- Accredited Calibration: • Done in a manner traceable to national and international standards • In accordance with PNS ISO/IEC 17025:2017 • Report must be signed by the approved signatory, and the PAB logo is affixed in the calibration certificate
- Traceable Calibration: • Instrument tested using manufacturer or applicable procedures • Processes ensure testing with traceability to national and international standards, but do not necessarily fulfill all the requirements of PNS ISO/IEC 17025:2017 • PAB logo is not affixed in the calibration certificate
Unit Calibrated with Limited Calibration
- Limited Calibration is carried out due to: • Customer's request • Lack of facility or resources in the calibration laboratory
- "Not tested specifications or ranges" are clearly mentioned in the calibration report
- Agreement with the customer for the Limited Calibration status is indicated in the Calibration Service Request, CMTS-FM-02
Unit Calibrated with Conditional Calibration
- Conditional Calibration is carried out when: • One or two test/s in the result of calibration has an OOT or non-compliance with specification • Adjustment is not possible
- Agreement with the customer for the Conditional Calibration status is indicated through: • DUT Service Report, CMTS-FM-57 • Onsite DUT Service Report, CMTS-FM-58 for Onsite Calibration
- Specify in the remarks section in the calibration certificate, the Out of Tolerance function of the DUT
- If 3 or more tests result of calibration has an OOT or non-compliance and adjustment is not possible, the DUT is to be remarked as defective