CMTS Calibration Laboratory Control Procedure
11 Questions
12 Views

Choose a study mode

Play Quiz
Study Flashcards
Spaced Repetition

Podcast

Play an AI-generated podcast conversation about this lesson

Questions and Answers

What is the applicable temperature range for calibration when it is being performed on-site?

  • 25 ± 20 °C
  • 20 ± 10 °C
  • 20 ± 5 °C
  • 25 ± 15 °C (correct)
  • What is the acceptable range for Relative Humidity during calibration?

  • 60 ± 30%
  • 55 ± 25% (correct)
  • 45 ± 25%
  • 50 ± 20%
  • In what type of calibration is the PAB logo affixed in the calibration certificate?

  • Traceable Calibration
  • Limited Calibration
  • Accredited Calibration (correct)
  • Conditional Calibration
  • When would Limited Calibration be carried out?

    <p>Upon customer's request or resource limitation</p> Signup and view all the answers

    What happens when a unit undergoes Conditional Calibration?

    <p>There are one or two tests with out-of-tolerance results</p> Signup and view all the answers

    What is the purpose of the procedure mentioned in the text?

    <p>To provide a system of control for the calibration process</p> Signup and view all the answers

    What is the responsibility of all personnel involved in calibration?

    <p>To adhere to the process described in the document</p> Signup and view all the answers

    What does CMTS stand for in the context of the text?

    <p>Cebu Metrologyx Technical Services</p> Signup and view all the answers

    What does OOT stand for based on the terms provided in the text?

    <p>Out-of-Tolerance</p> Signup and view all the answers

    Which document provides the general requirements for the competency of testing and calibration laboratories?

    <p>Document No.Title PNS ISO/IEC 17025:2017</p> Signup and view all the answers

    What is the acceptable range for temperature maintenance inside the Calibration Laboratory according to the procedure?

    <p>$23 ± 5$ °C (18°C to 28°C)</p> Signup and view all the answers

    Study Notes

    Environmental Monitoring

    • The laboratory manager is responsible for maintaining the environmental condition of the CMTS calibration laboratory.
    • The applicable environmental condition for on-site calibration must be: • 25 ± 10 °C (15°C to 35°C) for Temperature • 55 ± 25 % (30 % to 80%) for Relative Humidity
    • CMTS Lab personnel must verify the suitability of the environment before proceeding with calibration.
    • If the temperature or relative humidity is out of the limit, necessary corrective action must be initiated, and calibration activities will be suspended.

    Type of Calibration Service

    • Accredited Calibration: • Done in a manner traceable to national and international standards • In accordance with PNS ISO/IEC 17025:2017 • Report must be signed by the approved signatory, and the PAB logo is affixed in the calibration certificate
    • Traceable Calibration: • Instrument tested using manufacturer or applicable procedures • Processes ensure testing with traceability to national and international standards, but do not necessarily fulfill all the requirements of PNS ISO/IEC 17025:2017 • PAB logo is not affixed in the calibration certificate

    Unit Calibrated with Limited Calibration

    • Limited Calibration is carried out due to: • Customer's request • Lack of facility or resources in the calibration laboratory
    • "Not tested specifications or ranges" are clearly mentioned in the calibration report
    • Agreement with the customer for the Limited Calibration status is indicated in the Calibration Service Request, CMTS-FM-02

    Unit Calibrated with Conditional Calibration

    • Conditional Calibration is carried out when: • One or two test/s in the result of calibration has an OOT or non-compliance with specification • Adjustment is not possible
    • Agreement with the customer for the Conditional Calibration status is indicated through: • DUT Service Report, CMTS-FM-57 • Onsite DUT Service Report, CMTS-FM-58 for Onsite Calibration
    • Specify in the remarks section in the calibration certificate, the Out of Tolerance function of the DUT
    • If 3 or more tests result of calibration has an OOT or non-compliance and adjustment is not possible, the DUT is to be remarked as defective