RT5FIAPIA: Film Screen Image Acquisition, Processing, and Image Analysis PDF
Document Details
Uploaded by Deleted User
Tags
Summary
This document provides information on film processing, including a detailed look into the different types of darkroom constructions, locations, equipment requirements, and safety precautions. The content covers topics like darkroom equipment, ventilation, and safe lighting.
Full Transcript
8/24/24 RT5FIAPIA: FILM - PRELIM SCREEN IMAGE, Processing Room and Design ACQUISITION, PROCESSING AND...
8/24/24 RT5FIAPIA: FILM - PRELIM SCREEN IMAGE, Processing Room and Design ACQUISITION, PROCESSING AND Processing Room Equipments and Accesories IMAGE ANALYSIS Film Processing 1 2 Radiographic LOCATION Centrally located Processing Room Serviced by hatches from the adjacent imaging room Dark room is a place where the necessary Away from damp or hot areas handling and processing of films can be carried out safely and Accessible in terms of power and water supply efficiently. DARKROOM / Adjoining viewing room PROCESSING ROOM – is a room from which Completely light proof – no windows normal light excluded. 3 4 1 8/24/24 CENTRALLY LOCATED DARK ROOM Complete blackout is unnecessary. Safe illumination is required to facilitate darkroom procedures. It must exclude all outside light and provide safe artificial light. 5 6 PASS BOX / CASETTE HATCH Two light-tight and x-ray proof doors Both cannot be opened at the same time. 7 8 2 8/24/24 Darkroom construction Darkroom Construction Darkroom illumination Minimum floor area is 10 sq m Ceiling Height must be 2.5 m or 3 Darkroom equipment and its arrangement meters Size may reduce depending upon Health and safety in processing area the darkroom needs 9 10 Darkroom floor Relative humidity is maintained at 40- 60%. Non-slip flooring Room temperature maintained between 18-20 degrees Celsius Durable and easy to clean VENTILATION Light colored (low-light AND HEATING A minimum of 10 air changes per hour working conditions) All these conditions can be achieved using a good air conditioning system. 11 12 3 8/24/24 Satisfactory working conditions for the staff Adequate processing conditions X-ray apparatus in the Radioactive adjacent room should VENTILATION be properly placed – materials should be AND HEATING Efficient automatic processor performance stored as remotely primary beam never be directed at the as possible from darkroom wall. the darkroom. Adequate removal of stale humid air and supply of fresh air Intake and extract fans can be used 13 14 TYPE OF ENTRANCE RADIATION PROTECTION Walls adjacent to the radiographic rooms should be Single door system shielded with correct thickness of the lead - 1.6 – 2 mm lead or its equivalent: Double door system Ø25mm high quality barium plaster Ø225 mm thick single brick wall Maze and Labyrinth Ø150 mm thick concrete Revolving door system 15 16 4 8/24/24 SINGLE DOOR ENTRANCE D/A: This type of door can Latch or lock from inside cause Capable of releasing accidental the lock from outside in case of emergency exposure of Ideally no one should film to white work alone in a dark room light. 17 18 DOUBLE DOOR DOUBLE DOOR ENTRANCE It has two doors One leading in and other out with a short hallway in between. 19 20 5 8/24/24 Maze Entrance MAZE TYPE ENTRANCE Has no doors People can enter and exit at any moment 21 22 LABYRINTH TYPE ENTRANCE Advantages of labyrinth: Black paint for the interior of the passages Easy access to darkroom Matte black paint Vertical height- 2m No hazard to single handed operator Length – not less than 3 m Width – not more than 700mm Safelight fitted along the Allows air circulation passageway 23 24 6 8/24/24 REVOLVING DOOR ENTRANCE The most eficient but most expensive type of door. 3 feet in diameter. Lightproof access to the room. 25 26 For inspection & maintenance of DARK ROOM ILLUMINATION cassettes & screens Cleaning of work surfaces WHITE Servicing of equipment, LIGHT changing solutions Sited close to the ceiling ; White lighting Safe lighting Preferably centrally placed Moderate in intensity (60w tungsten, 30w fluorescent ) 27 28 7 8/24/24 Used in conjunction SAFE LIGHT with a 25 W lamp A source of light which will not fog the films and still provide adequate illumination under processing conditions. SAFE Extremes of heat and moisture deteriorates Consists of a lamp fixture in which a pearl bulb of 25 W is used with a colored filter placed in front of the LIGHT the filter bulb However no safe light is completely FILTERS Should be cleaned safe – depends on wavelength periodically /color , intensity of light and duration of exposure. 29 30 SAFELIGHT FILTERS Direct Safelighting Light from the safe lamp directly falls Sheet of gelatin dyed on the working surface in appropriate colour Minimum distance and sandwich of 1.2 m from the between two sheets working surface glass for protection Best for loading and unloading of films 31 32 8 8/24/24 Indirect Safelighting Directs the light towards When the white light passed the ceiling which reflects through the filters of safelight, it corresponds to the color of the light back into the room. How does filters to be transmitted. Intented to provide general illumination to the safelight darkroom work Therefore, correct matching of Suspended atleast 2.1 safelight filter to film is above the floor level necessary. 33 34 WITH NO SAFELIGHTING SAFELIGHTING 35 36 9 8/24/24 SAFELIGHTING HANDLING TIME Identification of Maximum time a film can be exposed to safelight during a procedure without respective causing any appreciable degree of fogging switches is average time: 20-45 secs important. 37 38 DARKROOM 1. MANUAL LAYOUT AND PROCESSING EQUIPMENTS PROCESSING APPARATUS 2.AUTOMATIC PROCESSING 39 40 10 8/24/24 MANUAL PROCESSING UNIT MATERIAL – PVC /Stainless steel 41 42 AUTOMATIC PROCESSOR 1. DRY BENCH LOADING BENCH 2. WET BENCH 43 44 11 8/24/24 DRY BENCH Wet Bench The dry bench is where the cassettes are unloaded and The wet bench is where the processing of the loaded with fresh film. films is carried out. The top of the dry bench must be large enough to accomodate the large cassettes in use when opened out. The usual method is to use a set of tanks holding The top surface must be should either be with wood or a developer, rinse water and fixer, and a large linoleum. tank for washing of films. It is usual to store film boxes, especially those current use, beneath the dry bench , either a cupboard (protected if near x-ray set), or in a film hopper. 45 46 47 48 12