Humidity & Bland Aerosol Therapy PDF

Summary

This document provides an overview of humidity and bland aerosol therapy. It discusses concepts like absolute and relative humidity and their significance in respiratory care. It also explains the objective of humidity therapy and various factors that influence it.

Full Transcript

Lesson 3 1 2  Absolute Humidity  Relative Humidity  The amount of water in a given volume of gas  A ratio between the amount of...

Lesson 3 1 2  Absolute Humidity  Relative Humidity  The amount of water in a given volume of gas  A ratio between the amount of water in a given  Expressed in mg/L volume of gas (content) and the amount it is capable of holding at that temperature (capacity)  Expressed as a percent and measured with a hygrometer  RH = (Absolute Humidity/Capacity) x 100 3 4  Involves adding water vapor and sometimes  Objective heat to the inspired gas  Compensate for water loss that occurs when dry  Water in a gaseous state gas is delivered or when the upper airway is  Also called molecular water or invisible bypassed in order to maintain normal physiologic moisture conditions in the lower airways 5 6 1  Nose  Warms, humidifies and filters inspired air  Pharynx, trachea and bronchial tree  Also warm and humidifies inspired air  As inspired gas moves into the lung it eventually achieves BTPS (100% RH at 37 degrees C)  Occurs at about 5 cm below carina (isothermic saturation boundary) 7 8  Above the ISB, temperature and humidity  Factors shifting ISB deeper into the lungs decrease during inspiration and increase  Breathing through the mouth during expiration  Breathing cold, dry air  Below the ISB, temperature and RH remain  When upper airway is bypassed constant  When minute ventilation is higher than normal  When a distal shift in the ISB compromises the body’s normal heat and humidification, humidity therapy may be indicated 9 10  When inspired air is fully saturated (100%) at  Relative humidity at body temperature body temperature  Expressed as a percentage  It holds 44 mg/L of water  Body Humidity = (Absolute Humidity / 44 mg/L) x 100  Exerts a water vapor pressure of 47 mmHg 11 12 2  Actual moisture deficit between the inspired air and the needs of the body  May be expressed in mg/L or as a %  Humidity Deficit = 44 mg/L - absolute humidity  When expressed as a %  (Humidity deficit / 44 mg/L) x 100 13 14  Primary  Medical gases at flows >4 lpm causes heat  Humidify dry medical gases and water loss to upper airway  Overcoming humidity deficit created when upper  Decreased ciliary motility airway is bypassed  Airway irritation  Secondary  Increased mucus production that can become  Managing hypothermia (must heat) thick and inspissated  Treating bronchospasm caused by cold air (must heat) 15 16  If upper airway is bypassed, the hazard of breathing dry gas is even greater  Breathing dry gas through and ET tube can cause damage to tracheal epithelium within minutes 17 18 3  Management of hypothermia  Heating and humidifying gases is one technique to raise core temperature  Temperatures over 40 degrees Celsius will burn the airway  Alleviation of bronchospasm caused by cold air 19 20  Humidifiers add molecular water to a gas  Most important factor affecting humidifier  Physical principles affecting humidifier performance function  If enough heat can be added, the effects of a small  Temperature surface area or short contact time can usually be  Surface area overcome  Contact time  The greater the temperature the more water vapor it can hold  Heated humidifiers always outperform non- heated humidifiers 21 22  The greater the surface area of contact between water and gas, the more opportunity for evaporation to occur 23 24 4  The longer the gas is in contact with the  Bubble humidifiers water, the greater the humidification  Passover humidifiers  Low flows provide more contact time, therefore  Simple provide more humidity than high flows  Wick  Membrane  Heat Moisture Exchangers 25 26  Breaks an underwater gas stream into small bubbles  Commonly used with oronasal O2 delivery systems  Provides about 25% body humidity  Becomes less effective as flow increases  limited effectiveness at flows > 10 lpm 27 28  Don’t heat – condensation obstructs tubing  Direct gas over a water surface  Incorporate a simple pressure relief valve  Can deliver 100% body humidity (pop-off) that release pressure above 2 psi  3 types  Simple passover  Wick  Membrane-type 29 30 5  Wick is placed upright in a water reservoir and surrounded by a heating element.  As gas enters the chamber, it flows around the wick, picking up moisture and leaving the chamber fully saturated 31 32  Separates water from the gas stream by  Maintain saturation at high flows means of a hydrophobic membrane.  Add little or no resistance to spontaneous  Water vapor molecules can easily pass breathing circuits through this membrane, but liquid cannot 33 34  Also known as artificial nose  Captures exhaled heat and moisture and uses it to heat and humidify the next inspiration  Should be used for short term humidification ( 30 mg/L of water vapor (can be accomplished with temperature set 35°C ± 2°C. 43 44  Liquid particles suspended in a gas 45 46  Sterile water  Improve the mobilization of respiratory  Hypotonic saline - 0.9% tonicity to prevent or relieve bronchospasm or inflammatory response  Hydrate airways of tracheostomy patient  Induce cough for sputum collection 47 48 8  Large volume jet nebulizer  Impeller nebulizer (spinning disk or room humidifier)  Ultrasonic nebulizer 49 50  Most common  Generates aerosol by gas passing at high velocity through a small “jet” orifice  Utilizes Bernoulli principle entrain room air and to draw water up the capillary tube into the gas stream to produce aerosol  Length, kinks, or water in tubing will decrease air entrainment and increase FiO2 51 52  If flows provided are not high enough, connect 2 or more nebulizers together with a “wye” connector 53 54 9  Disk rotates rapidly, drawing water up from the reservoir and throwing it through a slotted baffle, reducing the size of the particles  Popular for home use but does not produce clinically adequate aerosol output  Difficult to keep clean  High source of contamination 55 56  Piezoelectric transducer in couplant chamber  Frequency of the transducer is preset by the of the unit is electrically charged and factory and cannot be changed produced high frequency vibrations  Frequency controls the particle size  Vibrations break up the medications in the  Amplitude determines the volume of the cup into small particles, which are delivered aerosol output by altering the transducers to the patient vibrational energy 57 58  Has the highest output range without heating  Aerosol mask  90% of aerosol particles produced fall within  Face tent the 0.5 to 3 micron range  Tracheostomy tube (T-bar, Briggs adapter)  Tracheostomy mask  Mist tents 59 60 10  Short-term application of hypertonic saline  Cross-contamination and infection (3%-10%) to assist in mobilizing secretions  Environmental safety (for for sampling immunosuppressed and TB patients)  Need to use an aerosol device that delivers  Inadequate aerosol output high-density mists  The increase volume in surface fluid and  Inadequate input flow, siphon tube obstruction, irritation to airway produces a cough reflex jet orifice misalignment  Box 39-6 (page 835)  USN - check electric supply, gas flow through device, amplitude setting, couplant chamber for proper fill level, dirt or debris 61 62  Overhydration  Bronchospasm  Risk highest in infants and small children  If this occurs, stop therapy, give O2, administer  Risk also high in patients with preexisting fluid or bronchodilator electrolyte imbalances  If physician wants to continue with aerosol  Normal daily water gain for an adult is 200 ml/day treatment, it is appropriate to give because aerosol compensates for insensible water bronchodilators at specific intervals loss  Swelling of secretions 63 64 11

Use Quizgecko on...
Browser
Browser