Chapter 38 Designing for Radiation Protection PDF
Document Details
Uploaded by CheapestArtNouveau
University of Perpetual Help System JONELTA
Stewart C. Bushong
Tags
Summary
This chapter discusses designing for radiation protection. It covers key aspects of radiographic protection features, filtration, linearity, and operator shields. The material also describes various aspects of fluoroscopic protection features, dose area product, and barrier design.
Full Transcript
CHAPTER 38 DESIGNING FOR RADIATION PROTECTION RADIOGRAPHIC PROTECTION FEATURES Beam Alignment Each radiographic tube should be provided Protective X-ray Tube Housing...
CHAPTER 38 DESIGNING FOR RADIATION PROTECTION RADIOGRAPHIC PROTECTION FEATURES Beam Alignment Each radiographic tube should be provided Protective X-ray Tube Housing with a mechanism It must be contained within protective o Purpose: to ensure proper alignment housing of the x-ray beam & the IR It reduces leakage radiation Filtration Leakage radiation must be less than 100 Total Filtration: mR/hr at a distance of 1 m from the protective o 2.5 mm Al – operated above 70 kVp housing! o 1.5 mm Al – operated b/n 50-70 kVp o 0.5 mm Al – operated below 50 kVp Control Panel HVL: it measures filtration It must indicate the condition of exposure It must positively indicate when the x-ray Mammography tube is energized Total Filtration: 30 μm Mo or 60 μm Rh o kVp & mA indicators Reproducibility X-ray beam on must be positively & clearly For any radiographic technique, the output indicated to the radiologic technologist! radiation intensity should be constant form one exposure to another Source-to-Image Receptor Distance Indicator It must be provided The variation in x-ray intensity should not o Tape measure attached to the tube exceed 5%! housing o Lasers Linearity Ability of a radiographic unit to produce The SID indicator must be accurate to within constant radiation output for various 2% of the indicated SID! combinations of mA & exposure time Radiation Intensity: mR/mAs Collimation Light-localized, variable-aperture The maximum acceptable variation in rectangular collimators should be provided linearity is 10% from one mA station to an Cones & Diaphragms: for special adjacent mA station! examination Operator Shield The x-ray beam & the light beam must RT may be in the exposure room during coincide to within 2% of the SID! exposure, but only if protective apparel is worn Positive-Beam Limitation (PBL) Exposure Control: fixed, not a long cord 125 Automatic, light-localized, variable-aperture collimators Mobile X-ray Imaging System Lead Apron: should be assigned Page The PBL must be accurate to within 2% of Exposure Switch: at least 2 m from x-ray SID! tube during exposure STEWART C. BUSHONG SUMMARIZED BY: MEYNARD Y. CASTRO CHAPTER 38 DESIGNING FOR RADIATION PROTECTION FLUOROSCOPIC PROTECTION FEATURES Cumulative Timer It produces an audible sound when the Source-to-Skin Distance fluoroscopic time has exceeded 5 minutes Increased SSD: reduces entrance skin Designed to ensure that the radiologist is exposure (ESE) aware of the relative beam-on time during Stationary Fluoroscopes: not less than 38 each procedure cm Mobile Fluoroscopes: less than 30 cm Dose Area Product (DAP) A quantity that reflects not only the dose but Primary Protective Barrier also the volume of tissue irradiated Fluoroscopic IR Assembly: Expressed in: R-cm2 o It serves as a primary protective Better indicator of risk than dose barrier X-ray Beam Intensity o It must be 2 mm Pb equivalent o Table Top: