What must the user do when the software indicates decontamination is needed?
What must the user do when the software indicates decontamination is needed?
- Prime the instrument three times
- Click on instrument to be decontaminated (correct)
- Click on decontamination
- Choose the reagent containers to be decontaminated
True or false: The Ventana Benchmark Ultra must be decontaminated quarterly and annually.
True or false: The Ventana Benchmark Ultra must be decontaminated quarterly and annually.
False (B)
What is the purpose of this SOP?
What is the purpose of this SOP?
The purpose of this SOP is to ensure the Ventana Benchmark Ultra is free from any contaminants/bacteria that could affect the quality staining of the immunohistochemistry (IHC) and insitu hybridization slides (ISH) slides.
______ the software indicates decontamination is needed, the user must perform the following steps: Click on instrument to be decontaminated, click on decontamination, and choose the reagent containers to be decontaminated.
______ the software indicates decontamination is needed, the user must perform the following steps: Click on instrument to be decontaminated, click on decontamination, and choose the reagent containers to be decontaminated.
How often is the calibration of the Ventana Benchmark Ultra performed?
How often is the calibration of the Ventana Benchmark Ultra performed?
The calibration of the Ventana Benchmark Ultra is performed ______.
The calibration of the Ventana Benchmark Ultra is performed ______.
True or false: Stage 1 of the decontamination process involves rinsing the system with distilled water.
True or false: Stage 1 of the decontamination process involves rinsing the system with distilled water.
What type of equipment is used for the decontamination process?
What type of equipment is used for the decontamination process?
Quality control for the Ventana Benchmark Ultra is performed by ______.
Quality control for the Ventana Benchmark Ultra is performed by ______.
True or false: Personal Protective Equipment (PPE) is required for decontamination.
True or false: Personal Protective Equipment (PPE) is required for decontamination.
How often is the calibration of the Ventana Benchmark Ultra performed?
How often is the calibration of the Ventana Benchmark Ultra performed?
What type of solution is used during Stage 1 of the decontamination procedure?
What type of solution is used during Stage 1 of the decontamination procedure?
What type of maintenance is performed quarterly to ensure the Ventana Benchmark Ultra is functioning correctly?
What type of maintenance is performed quarterly to ensure the Ventana Benchmark Ultra is functioning correctly?
True or false: Quality control is done by software.
True or false: Quality control is done by software.
Waste must be transported to the waste room in the Histology Laboratory ______.
Waste must be transported to the waste room in the Histology Laboratory ______.
What steps must the user perform when the software indicates decontamination is needed?
What steps must the user perform when the software indicates decontamination is needed?
The Ventana Benchmark Ultra must be decontaminated ______.
The Ventana Benchmark Ultra must be decontaminated ______.
What must be done before the Ventana Benchmark Ultra is used by a pathologist?
What must be done before the Ventana Benchmark Ultra is used by a pathologist?
True or false: The pathologist is responsible for priming the instrument three times before testing.
True or false: The pathologist is responsible for priming the instrument three times before testing.
What must be done with the waste after decontamination?
What must be done with the waste after decontamination?
Study Notes
- The purpose of this SOP is to ensure the Ventana Benchmark Ultra is free from any contaminants/bacteria that could affect the quality staining of the immunohistochemistry (IHC) and insitu hybridization slides (ISH) slides.
- The equipment used for the decontamination process is a diluted Lysol ICB, bulk reagent carboys, distilled water (DI water), temperature verification slides, QC log quarterly maintenance (AP-HIS.A509), waste containers, large funnel, sharpie pen, and personal protective equipment (PPE).
- The calibration of the Ventana Benchmark Ultra is performed quarterly and annually.
- Quality control for the Ventana Benchmark Ultra is performed by software that notifies the user of maintenance needed for the decontamination and the preventative maintenance.
- The Ventana Benchmark Ultra decontamination procedure includes three stages: Stage 1, Stage 2, and Stage 3.
- In Stage 1, a sterilizing solution is pumped throughout the system and let soak.
- In Stage 2, the solution is rinsed out with distilled water.
- In Stage 3, fresh bulk fluids are primered onto the system.
- When the software indicates decontamination is needed, the user must perform the following steps: Click on instrument to be decontaminated, click on decontamination, and choose the reagent containers to be decontaminated.
- Waste must be transported to the waste room in the Histology Laboratory 441-13.
- The Ventana Benchmark Ultra must be decontaminated before use by a pathologist.
- The instrument must be primed three times before testing.
- The pathologist will review antibody test results and sign the decontamination form.