X-Ray Circuit Components Quiz
54 Questions
7 Views

Choose a study mode

Play Quiz
Study Flashcards
Spaced Repetition
Chat to lesson

Podcast

Play an AI-generated podcast conversation about this lesson

Questions and Answers

Which device compensates for small variations in incoming line voltage that may cause large variations in the kVp to the x-ray tube?

  • The exposure switch
  • The filament circuit
  • The voltage compensator (correct)
  • The autotransformer
  • What is the primary purpose of the high-voltage circuit?

  • To provide compensation for fluctuations in the incoming line voltage
  • To control amperage in the filament circuit
  • To increase the incoming voltage by the value of the transformer ratio
  • To supply the x-ray tube with voltage high enough to create x-rays (correct)
  • What is the function of the autotransformer in the x-ray circuit?

  • To heat the x-ray tube filament for thermionic emission of electrons
  • To terminate the exposure
  • To provide the means for kVp selection (correct)
  • To supply power to other parts of the x-ray circuit
  • What is the purpose of the exposure switch in the x-ray circuit?

    <p>To allow electric current to flow through the primary side of the step-up transformer</p> Signup and view all the answers

    What device terminates the exposure in the x-ray circuit and is set by the operator on the control console?

    <p>The exposure timer</p> Signup and view all the answers

    What is the primary purpose of the low-voltage circuit in the x-ray circuit?

    <p>To monitor and stabilize the incoming voltage</p> Signup and view all the answers

    What is the function of the filament circuit in the x-ray circuit?

    <p>To heat the x-ray tube filament for thermionic emission of electrons</p> Signup and view all the answers

    What is the purpose of the mA selector in the x-ray circuit?

    <p>To control amperage in the filament circuit</p> Signup and view all the answers

    What is the transformer ratio for the step-up transformer in the x-ray circuit?

    <p>At least 500:1</p> Signup and view all the answers

    Which device compensates for small variations in incoming line voltage that may cause large variations in the kVp to the x-ray tube?

    <p>The voltage compensator</p> Signup and view all the answers

    What is the primary purpose of the high-voltage circuit?

    <p>To supply the x-ray tube with voltage high enough to create x-rays</p> Signup and view all the answers

    What is the function of the autotransformer in the x-ray circuit?

    <p>To provide the means for kVp selection</p> Signup and view all the answers

    What is the purpose of the exposure switch in the x-ray circuit?

    <p>To allow electric current to flow through the primary side of the step-up transformer</p> Signup and view all the answers

    What device terminates the exposure in the x-ray circuit and is set by the operator on the control console?

    <p>The exposure timer</p> Signup and view all the answers

    What is the primary purpose of the low-voltage circuit in the x-ray circuit?

    <p>To monitor and stabilize the incoming voltage</p> Signup and view all the answers

    What is the function of the filament circuit in the x-ray circuit?

    <p>To heat the x-ray tube filament for thermionic emission of electrons</p> Signup and view all the answers

    What is the purpose of the mA selector in the x-ray circuit?

    <p>To control amperage in the filament circuit</p> Signup and view all the answers

    What is the transformer ratio for the step-up transformer in the x-ray circuit?

    <p>At least 500:1</p> Signup and view all the answers

    Which subcircuit is located between the AC power supply and the primary side of the step-up transformer?

    <p>The low-voltage circuit</p> Signup and view all the answers

    What is the function of the voltage compensator?

    <p>To provide compensation for fluctuations in the incoming line voltage</p> Signup and view all the answers

    What are the three functions of the autotransformer?

    <p>To provide compensation for fluctuations in the incoming line voltage, to supply power to other parts of the x-ray circuit, and to increase the incoming voltage</p> Signup and view all the answers

    What is the function of the exposure switch?

    <p>To induce current flow through the x-ray tube</p> Signup and view all the answers

    What is the function of the exposure timer?

    <p>To terminate the exposure</p> Signup and view all the answers

    What is the purpose of the filament circuit?

    <p>To heat the x-ray tube filament for thermionic emission of electrons</p> Signup and view all the answers

    Where does the high-voltage circuit begin and end?

    <p>With the secondary side of the step-up transformer</p> Signup and view all the answers

    What is the primary purpose of the high-voltage circuit?

    <p>To supply the x-ray tube with voltage high enough to create x-rays</p> Signup and view all the answers

    What is the function of the step-up transformer?

    <p>To increase the incoming voltage</p> Signup and view all the answers

    What is the primary function of the autotransformer in the x-ray circuit?

    <p>To provide means for kVp selection</p> Signup and view all the answers

    What is the purpose of the exposure switch in the x-ray circuit?

    <p>To induce current flow through the secondary side of the step-up transformer</p> Signup and view all the answers

    What does the exposure timer do in the x-ray circuit?

    <p>Terminates the exposure</p> Signup and view all the answers

    What is the function of the filament circuit in the x-ray circuit?

    <p>To heat the x-ray tube filament for thermionic emission of electrons</p> Signup and view all the answers

    What is the primary purpose of the high-voltage circuit in the x-ray circuit?

    <p>To supply the x-ray tube with voltage high enough to create x-rays</p> Signup and view all the answers

    What is the primary function of the low-voltage circuit in the x-ray circuit?

    <p>To supply power to other parts of the x-ray circuit</p> Signup and view all the answers

    What is the purpose of the mA selector in the x-ray circuit?

    <p>To determine the number of available electrons at the x-ray tube filament</p> Signup and view all the answers

    What is the purpose of the rectifier unit in the x-ray circuit?

    <p>To convert AC to DC</p> Signup and view all the answers

    What is the purpose of the voltage compensator in the x-ray circuit?

    <p>To compensate for fluctuations in the incoming line voltage</p> Signup and view all the answers

    What is the function of the autotransformer in the x-ray circuit?

    <p>It provides compensation for fluctuations in the incoming line voltage</p> Signup and view all the answers

    What is the purpose of the rectifier unit in the x-ray circuit?

    <p>To supply the x-ray tube with voltage high enough to create x-rays</p> Signup and view all the answers

    What is the primary purpose of the low-voltage circuit in the x-ray circuit?

    <p>To monitor and stabilize the incoming voltage</p> Signup and view all the answers

    What is the purpose of the exposure switch in the x-ray circuit?

    <p>To allow electric current to flow through the primary side of the step-up transformer</p> Signup and view all the answers

    What is the function of the filament circuit in the x-ray circuit?

    <p>To heat the x-ray tube filament for thermionic emission of electrons</p> Signup and view all the answers

    What is the purpose of the mA selector in the x-ray circuit?

    <p>To control the amperage in the filament circuit</p> Signup and view all the answers

    What is the purpose of the exposure timer in the x-ray circuit?

    <p>To terminate the exposure</p> Signup and view all the answers

    What is the purpose of the voltage compensator in the x-ray circuit?

    <p>To compensate for fluctuations in the incoming line voltage</p> Signup and view all the answers

    What is the function of the step-up transformer in the x-ray circuit?

    <p>To increase the incoming voltage by the value of the transformer ratio</p> Signup and view all the answers

    What is the primary function of the autotransformer in the x-ray circuit?

    <p>To provide compensation for fluctuations in the incoming line voltage</p> Signup and view all the answers

    What is the purpose of the exposure timer in the x-ray circuit?

    <p>To terminate the exposure</p> Signup and view all the answers

    What is the primary purpose of the high-voltage circuit in the x-ray circuit?

    <p>To supply the x-ray tube with voltage high enough to create x-rays</p> Signup and view all the answers

    What is the function of the filament circuit in the x-ray circuit?

    <p>To heat the x-ray tube filament for thermionic emission of electrons</p> Signup and view all the answers

    What is the purpose of the mA selector in the x-ray circuit?

    <p>To control amperage in the filament circuit</p> Signup and view all the answers

    What is the purpose of the rectifier unit in the x-ray circuit?

    <p>To convert AC to DC for the x-ray tube</p> Signup and view all the answers

    What is the purpose of the low-voltage circuit in the x-ray circuit?

    <p>To supply power to other parts of the x-ray circuit</p> Signup and view all the answers

    What is the purpose of the voltage compensator in the x-ray circuit?

    <p>To provide compensation for fluctuations in the incoming line voltage</p> Signup and view all the answers

    What is the function of the step-up transformer in the x-ray circuit?

    <p>To increase the incoming voltage by the value of the transformer ratio</p> Signup and view all the answers

    Study Notes

    Overview of X-Ray Circuit Components

    • The low-voltage circuit is the subcircuit between the AC power supply and the primary side of the step-up transformer.
    • The control console contains all devices in the low-voltage circuit except the step-up transformer.
    • The incoming voltage is monitored and stabilized by a voltage compensator due to small variations in the incoming line voltage that may cause large variations in the kVp to the x-ray tube.
    • The autotransformer serves three functions: it provides the means for kVp selection, it provides compensation for fluctuations in the incoming line voltage, and it supplies power to other parts of the x-ray circuit.
    • The exposure switch allows electric current to flow through the primary side of the step-up transformer, inducing current flow through the secondary side and creating voltage across the x-ray tube.
    • The exposure timer is a device that terminates the exposure and is set by the operator on the control console.
    • The filament circuit supplies a low current to heat the x-ray tube filament for thermionic emission of electrons.
    • The primary side of the filament circuit begins and ends with the contacts on the autotransformer.
    • The mA selector controls amperage in the filament circuit, determining the number of available electrons at the x-ray tube filament and thus determining the mA in the high-voltage circuit.
    • The high-voltage circuit begins and ends with the secondary side of the step-up transformer and includes the x-ray tube and the rectifier unit.
    • The step-up transformer increases the incoming voltage by the value of the transformer ratio, which is at least 500:1.
    • The primary purpose of the high-voltage circuit is to supply the x-ray tube with voltage high enough to create x-rays, and it is located in a tank near the x-ray machine with oil surrounding the transformers inside the tank for heat dissipation.

    Overview of X-Ray Circuit Components

    • The low-voltage circuit is the subcircuit between the AC power supply and the primary side of the step-up transformer.
    • The control console contains all devices in the low-voltage circuit except the step-up transformer.
    • The incoming voltage is monitored and stabilized by a voltage compensator due to small variations in the incoming line voltage that may cause large variations in the kVp to the x-ray tube.
    • The autotransformer serves three functions: it provides the means for kVp selection, it provides compensation for fluctuations in the incoming line voltage, and it supplies power to other parts of the x-ray circuit.
    • The exposure switch allows electric current to flow through the primary side of the step-up transformer, inducing current flow through the secondary side and creating voltage across the x-ray tube.
    • The exposure timer is a device that terminates the exposure and is set by the operator on the control console.
    • The filament circuit supplies a low current to heat the x-ray tube filament for thermionic emission of electrons.
    • The primary side of the filament circuit begins and ends with the contacts on the autotransformer.
    • The mA selector controls amperage in the filament circuit, determining the number of available electrons at the x-ray tube filament and thus determining the mA in the high-voltage circuit.
    • The high-voltage circuit begins and ends with the secondary side of the step-up transformer and includes the x-ray tube and the rectifier unit.
    • The step-up transformer increases the incoming voltage by the value of the transformer ratio, which is at least 500:1.
    • The primary purpose of the high-voltage circuit is to supply the x-ray tube with voltage high enough to create x-rays, and it is located in a tank near the x-ray machine with oil surrounding the transformers inside the tank for heat dissipation.

    Overview of X-Ray Circuit Components

    • The low-voltage circuit is the subcircuit between the AC power supply and the primary side of the step-up transformer.
    • The control console contains all devices in the low-voltage circuit except the step-up transformer.
    • The incoming voltage is monitored and stabilized by a voltage compensator due to small variations in the incoming line voltage that may cause large variations in the kVp to the x-ray tube.
    • The autotransformer serves three functions: it provides the means for kVp selection, it provides compensation for fluctuations in the incoming line voltage, and it supplies power to other parts of the x-ray circuit.
    • The exposure switch allows electric current to flow through the primary side of the step-up transformer, inducing current flow through the secondary side and creating voltage across the x-ray tube.
    • The exposure timer is a device that terminates the exposure and is set by the operator on the control console.
    • The filament circuit supplies a low current to heat the x-ray tube filament for thermionic emission of electrons.
    • The primary side of the filament circuit begins and ends with the contacts on the autotransformer.
    • The mA selector controls amperage in the filament circuit, determining the number of available electrons at the x-ray tube filament and thus determining the mA in the high-voltage circuit.
    • The high-voltage circuit begins and ends with the secondary side of the step-up transformer and includes the x-ray tube and the rectifier unit.
    • The step-up transformer increases the incoming voltage by the value of the transformer ratio, which is at least 500:1.
    • The primary purpose of the high-voltage circuit is to supply the x-ray tube with voltage high enough to create x-rays, and it is located in a tank near the x-ray machine with oil surrounding the transformers inside the tank for heat dissipation.

    Studying That Suits You

    Use AI to generate personalized quizzes and flashcards to suit your learning preferences.

    Quiz Team

    Description

    Test your knowledge on the fundamental components of an X-ray circuit with this quiz! From the low-voltage circuit to the high-voltage circuit, learn about the autotransformer, exposure switch, exposure timer, filament circuit, and more. This quiz will challenge your understanding of the function and purpose of each component in an X-ray circuit. Sharpen your skills and test your knowledge today!

    More Like This

    Electric Circuits and X-Rays
    5 questions

    Electric Circuits and X-Rays

    JawDroppingWalrus1974 avatar
    JawDroppingWalrus1974
    X-ray Machine Classification
    0 questions
    X-ray Machine Classification
    0 questions
    ELE243 Chapter 4: Cathode Ray Oscilloscope (CRO)
    30 questions
    Use Quizgecko on...
    Browser
    Browser