Podcast
Questions and Answers
Where should the HAZ hardness impressions be located?
Where should the HAZ hardness impressions be located?
- Entirely within the fusion boundary
- Entirely within the base metal and close to the fusion boundary
- Entirely within the weld metal and close to the fusion boundary
- Entirely within the heat-affected zone and close to the fusion boundary (correct)
What are the requirements for impact energy values of repair weld specimens?
What are the requirements for impact energy values of repair weld specimens?
- Maximum average and minimum individual values shall not be less than specified for qualification of the production weld
- There are no specific requirements for impact energy values
- Minimum average and minimum individual values shall be less than specified for qualification of the production weld
- Minimum average and minimum individual values shall not be less than specified for qualification of the production weld (correct)
Who should make the repair weld?
Who should make the repair weld?
- A mechanical engineer
- A welding inspector
- A qualified welder experienced in methods used for repair of a defective weld (correct)
- An unqualified welder with basic welding knowledge
What are the qualifications required for a repair welder?
What are the qualifications required for a repair welder?
What is the formula to determine the minimum distance between the source and object in radiographic testing procedures for welds?
What is the formula to determine the minimum distance between the source and object in radiographic testing procedures for welds?
What must IQIs conform to in radiographic testing procedures for welds?
What must IQIs conform to in radiographic testing procedures for welds?
What should the selection of IQIs be based on in radiographic testing procedures for welds?
What should the selection of IQIs be based on in radiographic testing procedures for welds?
Who can interpret production weld images in radiographic testing procedures?
Who can interpret production weld images in radiographic testing procedures?
What must radiographers do if defects are observed in the images of welds?
What must radiographers do if defects are observed in the images of welds?
What is the exposure geometry evaluated at for in-motion imaging in radiographic testing procedures for welds?
What is the exposure geometry evaluated at for in-motion imaging in radiographic testing procedures for welds?
What material should IQIs be made of in radiographic testing procedures for welds?
What material should IQIs be made of in radiographic testing procedures for welds?
What does the symbol 'D' represent in the formula $D = \frac{St}{k}$?
What does the symbol 'D' represent in the formula $D = \frac{St}{k}$?
What standard should IQIs conform to in radiographic testing procedures for welds?
What standard should IQIs conform to in radiographic testing procedures for welds?
What is the requirement for interpreting production weld images in radiographic testing procedures?
What is the requirement for interpreting production weld images in radiographic testing procedures?
What factor determines the minimum distance between source and object in radiographic testing procedures for welds?
What factor determines the minimum distance between source and object in radiographic testing procedures for welds?
What do radiographers need to do with observed defects in images of welds?
What do radiographers need to do with observed defects in images of welds?
What is the minimum length of repair weld required for necessary deposit for destructive testing?
What is the minimum length of repair weld required for necessary deposit for destructive testing?
Under what conditions may a single completed weld qualify more than one type of repair?
Under what conditions may a single completed weld qualify more than one type of repair?
What happens if a welder fails the qualification test?
What happens if a welder fails the qualification test?
What is required if essential variables are changed for repair welders?
What is required if essential variables are changed for repair welders?
How are repaired areas required to be inspected and evaluated?
How are repaired areas required to be inspected and evaluated?
What is the specific requirement for radiographic test methods?
What is the specific requirement for radiographic test methods?
What is the specific maximum hardness values provided for?
What is the specific maximum hardness values provided for?
What type of testing is required for repair welds?
What type of testing is required for repair welds?
Who must supervise the repairs according to the text?
Who must supervise the repairs according to the text?
What must be recorded and maintained regarding repair welder qualification?
What must be recorded and maintained regarding repair welder qualification?
What is the requirement for the segment on which the repair weld is deposited?
What is the requirement for the segment on which the repair weld is deposited?
What is the specific requirement for the length of repair weld for necessary deposit for destructive testing?
What is the specific requirement for the length of repair weld for necessary deposit for destructive testing?
What is responsible for evaluating whether the weld meets the requirements of Section 9?
What is responsible for evaluating whether the weld meets the requirements of Section 9?
How should unexposed films be stored?
How should unexposed films be stored?
What must be established and agreed upon by the company and the NDT contractor for magnetic particle testing?
What must be established and agreed upon by the company and the NDT contractor for magnetic particle testing?
What is required for film viewing equipment?
What is required for film viewing equipment?
How long should film or other imaging media be ensured interpretability for, if requested by the company?
How long should film or other imaging media be ensured interpretability for, if requested by the company?
Who is responsible for the protection and monitoring of individuals working with or near radiation sources?
Who is responsible for the protection and monitoring of individuals working with or near radiation sources?
What must be agreed upon by the company and the ultrasonic contractor for ultrasonic testing of new and/or in-service circumferential butt welds?
What must be agreed upon by the company and the ultrasonic contractor for ultrasonic testing of new and/or in-service circumferential butt welds?
What should film viewing facilities provide to avoid reflections, shadows, or glare on the radiograph?
What should film viewing facilities provide to avoid reflections, shadows, or glare on the radiograph?
What must be established and agreed upon by the company and the NDT contractor for liquid penetrant testing?
What must be established and agreed upon by the company and the NDT contractor for liquid penetrant testing?
What should film density requirements for both transparent-based and opaque-based films have?
What should film density requirements for both transparent-based and opaque-based films have?
What must the image processing area and accessories be kept as at all times?
What must the image processing area and accessories be kept as at all times?
What should be prevented during interpretation by film viewing equipment?
What should be prevented during interpretation by film viewing equipment?
What must be agreed upon by the company and the radiographic contractor before production radiography?
What must be agreed upon by the company and the radiographic contractor before production radiography?
How many exposures are required for radiographic testing of a complete weld when the radiographic source is outside but not more than 1/2 in. from the weld surface?
How many exposures are required for radiographic testing of a complete weld when the radiographic source is outside but not more than 1/2 in. from the weld surface?
What procedure may be used when the OD of the piping containing the weld is 3.500 in. or less?
What procedure may be used when the OD of the piping containing the weld is 3.500 in. or less?
What should be done when smaller diameter, thicker wall pipe is radiographed?
What should be done when smaller diameter, thicker wall pipe is radiographed?
What is included in the film radiography procedure?
What is included in the film radiography procedure?
What is required in the procedure for radiography using imaging media other than film?
What is required in the procedure for radiography using imaging media other than film?
What is the minimum number of exposures required for radiographic testing of a complete weld when the radiographic source is outside and more than 1/2 in. from the weld surface?
What is the minimum number of exposures required for radiographic testing of a complete weld when the radiographic source is outside and more than 1/2 in. from the weld surface?
What is the frequency of use of radiographic testing determined by?
What is the frequency of use of radiographic testing determined by?
What must the radiographic testing procedures be compatible with?
What must the radiographic testing procedures be compatible with?
What is the requirement for evaluating images in radiographic testing procedures?
What is the requirement for evaluating images in radiographic testing procedures?
What is the requirement for radiographic testing procedures using X-rays and gamma rays?
What is the requirement for radiographic testing procedures using X-rays and gamma rays?
What is the responsibility of the company and the radiographic contractor in radiographic testing procedures?
What is the responsibility of the company and the radiographic contractor in radiographic testing procedures?
Study Notes
Radiographic Inspection and NDT Methods in Welding
- The radiographer is responsible for evaluating whether the weld meets the requirements of Section 9, while the company determines the final disposition of the weld.
- Images used for inspection must be clearly identified with lead numbers, letters, markers, or other identification methods, as per the company's specification.
- Unexposed films should be stored in a clean, dry place, and if their condition is in question, a test film should be processed to check for fog.
- Film density requirements for both transparent-based and opaque-based films are specified, with corresponding minimum and maximum density limits.
- Film viewing equipment must be capable of viewing film densities within the specified range and should prevent light interference during interpretation.
- Viewing facilities should provide subdued background lighting to avoid reflections, shadows, or glare on the radiograph.
- Film or other imaging media should be processed, handled, and stored in a way that ensures interpretability for at least three years after production, if requested by the company.
- The image processing area and accessories must be kept clean at all times.
- The radiographer is responsible for the protection and monitoring of individuals working with or near radiation sources, complying with applicable regulations.
- For magnetic particle testing and liquid penetrant testing, detailed written procedures meeting specific standards (ASTM E709 for magnetic particle testing and ASTM E165 for liquid penetrant testing) must be established and agreed upon by the company and the NDT contractor.
- Ultrasonic testing for inspection of new and/or in-service circumferential butt welds requires a detailed procedure, agreed upon by the company and the ultrasonic contractor, to produce acceptable and accurate results for production testing.
- The use of ultrasonic testing and the scope of its use shall be at the option of the company, with the company requiring the ultrasonic contractor to use approved procedures for production testing.
Radiographic Testing Procedures for Pipeline Welds
- Criteria for evaluating images include acceptable image quality, prescribed IQI and wire diameter, satisfactory identification system, acceptable technique and setup, and compatibility with acceptance standards
- All requirements for image quality apply equally to X-rays and gamma rays
- Use of radiographic testing and its frequency is at the option of the company
- The company and the radiographic contractor must agree on procedures prior to production radiography
- Details of each radiographic procedure must be recorded and furnished to the company
- Film radiography procedure must include radiation source, intensifying screens, film, exposure geometry, exposure conditions, processing, materials, IQIs, and heat shields
- Procedure for radiography using imaging media other than film must include radiation source, image collection and processing systems, exposure geometry, exposure conditions, materials, IQIs, and heat shields
- When a radiographic source is centered in the pipe for exposing a butt weld, one exposure is adequate for radiographic testing of the complete weld
- When the radiographic source is outside but not more than 1/2 in. from the weld surface, at least three exposures separated by 120° shall be made for radiographic testing of a complete weld
- When the radiographic source is outside and more than 1/2 in. from the weld surface, at least four exposures separated by 90° shall be made for radiographic testing of a complete weld
- When the OD of the piping containing the weld is 3.500 in. or less, a DWE/DWV procedure may be used
- When smaller diameter, thicker wall pipe is radiographed, additional exposures should be made to minimize the distortion of imperfection images at the ends of the radiographs
Studying That Suits You
Use AI to generate personalized quizzes and flashcards to suit your learning preferences.
Related Documents
Description
Test your knowledge of radiographic inspection and non-destructive testing (NDT) methods in welding with this quiz. Explore topics such as radiographic testing procedures for pipeline welds, image evaluation criteria, film processing and storage requirements, and detailed procedures for magnetic particle testing, liquid penetrant testing, and ultrasonic testing.