Radiation Safety Measures

Choose a study mode

Play Quiz
Study Flashcards
Spaced Repetition
Chat to Lesson

Podcast

Play an AI-generated podcast conversation about this lesson

Questions and Answers

What is the primary purpose of lead lining in the diagnostic-type protective tube housing?

  • To maintain a constant temperature within the x-ray tube during operation.
  • To protect patients and personnel from off-focus or leakage radiation. (correct)
  • To reduce the overall weight of the x-ray tube housing for easier maneuverability.
  • To enhance image resolution by focusing the x-ray beam.

When a diagnostic x-ray tube is operated at its highest voltage and current, what is the maximum permissible radiation leakage at a distance of 1 meter from the x-ray source?

  • 1.76 mGya/hr
  • 0.88 mGya/hr (correct)
  • 2.64 mGya/hr
  • 0.22 mGya/hr

Which of the following is a characteristic of the control panel or console in an x-ray imaging system, designed for radiation safety?

  • It must use analog displays for mA and kVp readouts to improve accuracy.
  • It must provide a positive indication when the x-ray tube is energized. (correct)
  • It must be easily accessible to patients for adjustments during the procedure.
  • It must be located within the x-ray room to allow for immediate adjustments.

What is the key consideration for the material used in the construction of a radiographic examination table to minimize patient radiation dose?

<p>It must be as radiolucent as possible to minimize radiation absorption. (B)</p> Signup and view all the answers

What is the accuracy requirement for the distance indicator on a Source-to-Image Receptor Distance (SID) indicator?

<p>It must be accurate to within 2% of the SID. (B)</p> Signup and view all the answers

What is the accuracy requirement for centering indicators?

<p>1% within the SID. (A)</p> Signup and view all the answers

What is the primary purpose of collimation in diagnostic radiology?

<p>To limit the x-ray field size to the area of clinical interest, reducing patient dose. (D)</p> Signup and view all the answers

Which of the following is a key benefit of restricting the x-ray field size using collimation?

<p>Significant reduction in patient dose and improved image quality. (C)</p> Signup and view all the answers

What is the function of positive beam limitation (PBL)?

<p>It collimates the radiographic beam to the size of the image receptor. (A)</p> Signup and view all the answers

Which type of X-Ray beam limitation device is used mostly with dental x-ray machines?

<p>Cones. (B)</p> Signup and view all the answers

What is the primary goal of radiographic beam filtration?

<p>To reduce patient skin exposure and absorbed dose by removing low-energy photons. (B)</p> Signup and view all the answers

What material is most commonly utilized in radiographic beam filtration?

<p>Aluminum (C)</p> Signup and view all the answers

What does the half-value layer (HVL) measure in diagnostic radiology?

<p>The thickness of a material required to reduce the intensity of the primary beam by 50%. (C)</p> Signup and view all the answers

How frequently should the half-value layer (HVL) be measured by a radiologic physicist to ensure adequate x-ray beam filtration?

<p>At least once a year and after any x-ray tube replacement or repairs (B)</p> Signup and view all the answers

What is the primary purpose of employing compensating filters in radiography?

<p>To achieve uniform imaging of body parts with varying thicknesses or tissue composition. (B)</p> Signup and view all the answers

Which of the following best describes the mechanism by which compensating filters function?

<p>They partially attenuate x-rays directed toward thinner areas, permitting more x-radiation to strike thicker areas. (C)</p> Signup and view all the answers

What is assessed when evaluating exposure reproducibility in radiographic equipment?

<p>The consistency in output radiation intensity for identical generator settings. (D)</p> Signup and view all the answers

What variance is acceptable for exposure reproducibility?

<p>5% or less. (A)</p> Signup and view all the answers

What does exposure linearity evaluate in the context of x-ray equipment performance?

<p>The consistency in output radiation intensity at selected kVp settings when generator settings are changed. (C)</p> Signup and view all the answers

What variance is acceptable for exposure linearity?

<p>10% or less. (A)</p> Signup and view all the answers

In mobile radiography, what is the general source to image distance (SID) used?

<p>100 cm (40 inches) or even 120 cm (48 inches). (D)</p> Signup and view all the answers

What is the minimum SSD of mobile radiography?

<p>At least 30 cm (12 inches). (D)</p> Signup and view all the answers

Which of the following best describes when mobile radiography is appropriate?

<p>If a patient is unable to be transported to a fixed radiographic installation. (A)</p> Signup and view all the answers

Which of the following is not an example of mobile radiography?

<p>Radiographic room. (B)</p> Signup and view all the answers

What is the purpose of diagnostic imaging systems?

<p>Every diagnostic imaging system must have protective tube housing (A)</p> Signup and view all the answers

Why are diagnostic imaging systems important?

<p>Operates safely to protect: patients &amp; personnel. (A)</p> Signup and view all the answers

How can a tube housing cause harm to the surrounding environment?

<p>Lead lined and metal from off focus radiation can lead to leaking into the the patient. (C)</p> Signup and view all the answers

Why is aluminum chose widely for machines above 70kVp?

<p>Aluminum is the metal most widely selected because it removes low-energy (soft) x-rays . (B)</p> Signup and view all the answers

What are examples of what the tube requires for inherent filtration?

<p>Glass window and metal housing (B)</p> Signup and view all the answers

What is the goal of using carbon in a radiolucent material for an x-ray examination table?

<p>To reduce the material amount that will absorb radiation. (C)</p> Signup and view all the answers

What features have digital readouts, permitting the operator to assess exposure conditions?

<p>The kVp &amp; mA values. (A)</p> Signup and view all the answers

If a hospital does not have a variable-aperture rectangular collimator, which tool can they use?

<p>The current, predominant x-ray beam limitation device. (C)</p> Signup and view all the answers

What is a benefit of collimating the radiographic beam to the IR size?

<p>Improves the overall image quality. (B)</p> Signup and view all the answers

What is the end purpose of collimating the radiographic beam to the IR size?

<p>Supports law that collimation cannot be larger than IR (D)</p> Signup and view all the answers

What is the purpose of filtration?

<p>To reduce absorbed dose to patients body by removing low energy. (B)</p> Signup and view all the answers

How is aluminum filtration connected to added filtration?

<p>aluminum filtration = added filtration (D)</p> Signup and view all the answers

To what extend does a radiographer want their table to be radiolocent?

<p>Must be as radiolucent as possible (C)</p> Signup and view all the answers

What is the primary goal of a compensating filter?

<p>To accomplish uniform imaging of body parts. (D)</p> Signup and view all the answers

Why does SSD needs to be increased by 100cm or 120cm?

<p>There is a more uniform distribution of exposure throughout the pt. (B)</p> Signup and view all the answers

Flashcards

Radiation Safety Features

Measures ensuring radiographic equipment operates safely to protect patients and personnel.

Protective Tube Housing

A shield that prevents excessive radiation emission, protecting from off-focus or leakage radiation.

Functioning Control Panel

Displays radiation exposure settings and confirms X-ray tube energization.

Lead-lined & metal

To protect from off-focus or leakage radiation.

Signup and view all the flashcards

Emission Restriction

Restricting X-rays to the useful area for patients.

Signup and view all the flashcards

Radiation Leakage Limit

Ensures radiation leakage doesn't exceed 0.88 mGya/hr at 1 m from source.

Signup and view all the flashcards

Control Panel Location

Must be behind a protective barrier with a radiation-absorbent window.

Signup and view all the flashcards

mA and kVp Readouts

A visible display of mA and kVp to assess exposure conditions.

Signup and view all the flashcards

Audible Sound

Indicates X-ray tube is energized and ionizing radiation emitted.

Signup and view all the flashcards

Table Strength

Must adequately support patients over 300 pounds.

Signup and view all the flashcards

Floating Table

Floating tabletop to make it easier to maneuver the patient

Signup and view all the flashcards

Uniform Thickness

Uniformity ensures consistent image results.

Signup and view all the flashcards

Radiolucent Table

Material that minimizes radiation absorption.

Signup and view all the flashcards

Carbon vs Metal

Used for easy maneuvering and support.

Signup and view all the flashcards

SID Indicator

Ensures correct Source-to-Image distance is maintained.

Signup and view all the flashcards

Distance Accuracy

Must have accuracy within 2% of the SID.

Signup and view all the flashcards

Centering Accuracy

Ensuring accurate alignment, must be within 1% of the SID.

Signup and view all the flashcards

Primary X-Ray Beam

Adequately collimated to be no larger than the IR size.

Signup and view all the flashcards

Rectangular Collimator

Device that adjusts the size/shape of the X-ray beam.

Signup and view all the flashcards

Restricting X-ray Field Size

Reduces patient dose and improves image quality.

Signup and view all the flashcards

Positive Beam Limitation (PBL)

Collimates the radiographic beam to the IR size.

Signup and view all the flashcards

X-Ray Cones

Used mostly with dental X-ray machines.

Signup and view all the flashcards

Purpose of Filtration

Reduces dose by removing low energy photons.

Signup and view all the flashcards

Types of Filtration

Inherent + Added = Total filtration

Signup and view all the flashcards

Aluminum Filtration

Most widely used metal filtration; removes soft X-rays.

Signup and view all the flashcards

Half-Value Layer (HVL)

Thickness needed to reduce beam intensity by 50%.

Signup and view all the flashcards

HVL Measurement

Should be measured annually & after X-ray tube changes.

Signup and view all the flashcards

compensating filters

Made of aluminum, lead-acrylic, or other suitable materials

Signup and view all the flashcards

Exposure linearity

Maintained when generator settings are changed from one mA and time combination to another.

Signup and view all the flashcards

Exposure Reproducibility

Intensity for identical generator settings from one individual exposure to subsequent exposures

Signup and view all the flashcards

Minimum SSD for Mobile Radiography

SSD should be at least 30 cm.

Signup and view all the flashcards

Study Notes

  • Measures ensure radiographic equipment operates safely for patients and personnel.
  • Every diagnostic imaging system must have a protective tube housing and a correctly functioning control panel.
  • Radiographic examination tables and accessories must be designed to reduce patient radiation dose.

Diagnostic-Type Protective Tube Housing

  • Requirements include lead-lined & metal construction.
  • Designed to protect patients and imaging personnel from off-focus or leakage radiation.
  • Restricts x-ray emission to the useful or primary beam area.
  • Radiation leakage through the housing must not exceed 0.88 mGya/hr, measured 1 meter from the x-ray source.
  • Leakage is measured when the tube operates at its highest voltage and current for continuous operation.

Control Panel or Console

  • Must be located behind a suitable protective barrier with a radiation-absorbent window.
  • Must provide positive indication when the x-ray tube is energized.
  • Should have visible milliampere (mA) and kilovoltage peak (kVp) digital readouts.
  • Readouts permit operator assessment of exposure conditions.
  • A tone is emitted when exposure begins, stopping when exposure terminates.
  • Audible sound indicates when the x-ray tube is energized and emitting radiation.

Radiographic Examination Table

  • Must be strong enough to adequately support patient weights exceeding 300 pounds.
  • Often features a floating tabletop for easy patient maneuvering.
  • Must be of uniform thickness.
  • Needs to be constructed with a radiolucent material to minimize radiation absorption for patient safety.
  • Frequently made of carbon fiber material.

Source-to-Image Receptor Distance (SID) Indicator

  • Provides a way to measure the distance from the anode focal spot to the IR.
  • Ensures that the correct SID is maintained.
  • Often is a simple device, like a tape measure, attached to the collimator or tube housing.
  • Lasers can measure SID.
  • Distance must be accurate to within 2% of the SID.
  • Centering indicators must be accurate to within 1% of the SID.

X-Ray Beam Limitation Devices

  • The primary x-ray beam should be adequately collimated.
  • Size should be no larger than the size of the image receptor (IR) being used for the examination.
  • Light-localizing variable-aperture rectangular collimator adjusts x-ray beam size and shape automatically or manually.
  • Collimators are the most common x-ray beam limitation device.
  • Other types include cones and cylinders.
  • Restricting x-ray field size reduces patient dose and improves image quality.

Positive Beam Limitation (PBL)

  • Collimates the radiographic beam to the IR size.
  • Supports the law that collimation cannot be larger than the IR.

Cones

  • Often used with dental x-ray machines.
  • Sometimes used for examinations of specific areas, such as the head.
  • Features flared metal tubes and straight cylinders.

Filtration

  • Purpose is to reduce absorbed dose to the patient by removing low-energy photons.
  • Types of filtration: inherent + added = total filtration.
  • Minimum total filtration required for machines at 70kVp or higher: 2.5 mm Al/Eq.
  • Aluminum is widely selected for its effectiveness in removing low-energy x-rays from a polyenergetic x-ray beam.
  • It removes them without severely decreasing x-ray beam intensity.
  • A diagnostic x-ray beam must always be adequately filtered.

Half-Value Layer (HVL)

  • The thickness of a designated absorber, such as aluminum, reduces primary beam intensity by 50% of its initial value.
  • Needs to be measured to verify adequate filtration of the x-ray beam.
  • Measurement should be taken by a radiologic physicist at least once a year.
  • Measurement is also taken after an x-ray tube is replaced or after repairs.
  • Expressed in millimeters of aluminum.
  • It measures beam quality or effective energy of the x-ray beam.
  • A minimal HVL is required at a given kVp.

Compensating Filters

  • Filters are made of aluminum, lead-acrylic, or other suitable materials.
  • Used to accomplish uniform imaging of body parts with varying thickness or tissue composition.
  • Partially attenuates x-rays directed toward the thinner areas of the filter.
  • Permits more x-radiation to pass through the thicker areas.
  • Examples: wedge filter and trough (bilateral wedge) filter.

Exposure Reproducibility

  • Is the consistency in output in radiation intensity, for identical generator settings.
  • Requires one individual exposure to subsequent exposures
  • May be verified by making a series of repeated exposures using the same technical factors.
  • Monitor radiation intensity variation with a calibrated ion chamber.
  • Variance of 5% or less is acceptable.

Exposure Linearity

  • Is the consistency in output radiation intensity at selected kVp settings.
  • Linearity changes generator settings from one mA and time combination to another.
  • Variance of 10% or less is acceptable.

Mobile Radiography

  • Only for patients who cannot be transported to a fixed radiographic installation.
  • The SID generally used for mobile radiography is 100 cm (40 inches) or 120 cm (48 inches).
  • The minimum SSD is at least 30 cm (12 inches).
  • Remember for stationary machines minimum SSD is at least 38cm (15 inches).
  • Increased SSD such as 100 or 120 cm yields a more uniform distribution of exposure.
  • Examples: portable machines and C-arm machines.

Studying That Suits You

Use AI to generate personalized quizzes and flashcards to suit your learning preferences.

Quiz Team

Related Documents

More Like This

Radiography Essentials and Patient Safety
27 questions
X-Ray Radiation Safety
37 questions

X-Ray Radiation Safety

OptimalBalalaika3628 avatar
OptimalBalalaika3628
Use Quizgecko on...
Browser
Browser