Podcast
Questions and Answers
Que é un funcional probador segundo as normativas mencionadas?
Que é un funcional probador segundo as normativas mencionadas?
Os valores de SpO2 obtidos mediante un funcional probador son:
Os valores de SpO2 obtidos mediante un funcional probador son:
Para verificar o rendemento dos dispositivos de SpO2, un funcional probador debe usar:
Para verificar o rendemento dos dispositivos de SpO2, un funcional probador debe usar:
Que papel xoga a curva R na avaliación dos dispositivos de SpO2?
Que papel xoga a curva R na avaliación dos dispositivos de SpO2?
Signup and view all the answers
Cal é a función principal do sensor PPG do dispositivo MAXREFDES103?
Cal é a función principal do sensor PPG do dispositivo MAXREFDES103?
Signup and view all the answers
O que se necesita para establecer o ambiente de proba para o AECG100?
O que se necesita para establecer o ambiente de proba para o AECG100?
Signup and view all the answers
Que tipo de luz usa o dispositivo durante a proba de SpO2?
Que tipo de luz usa o dispositivo durante a proba de SpO2?
Signup and view all the answers
A que se refire a calibración clínica de dispositivos de SpO2?
A que se refire a calibración clínica de dispositivos de SpO2?
Signup and view all the answers
Que aspecto non é regulado polas normativas mencionadas sobre os probadores de función?
Que aspecto non é regulado polas normativas mencionadas sobre os probadores de función?
Signup and view all the answers
Que se busca ao evaluar a capacidade de reprodución do dispositivo de SpO2?
Que se busca ao evaluar a capacidade de reprodución do dispositivo de SpO2?
Signup and view all the answers
Study Notes
Introduction to Pulse Oximeter Testing
- Compliance with ISO 80601-2-61:2011 and YY0784_2010 standards for testing pulse oximeter devices using simulation signal generators referred to as functional test instruments.
- Functional test instruments help determine if pulse oximeters and their probes can perform as designed by manufacturers, but they do not confirm accuracy of performance.
- Accurate blood oxygen values require clinical calibration using CO-oximeters or recalibrated pulse oximeter devices.
Functional Test Instruments
- These instruments provide a preset R value signal, allowing stakeholders to observe the displayed SpO2 values for comparison against expected values for specific pulse oximeter models.
- If a manufacturer knows the R curve associated with a particular model, they can generate an accurate R value corresponding to specified blood oxygen levels (e.g., 85%).
- Assessment of a pulse oximeter's ability to reproduce its internal calibration curve is possible.
Limitations of SpO2 Readings
- Accurate SpO2 readings from functional test instruments do not guarantee accuracy on human subjects.
- Function test instruments can only evaluate the reproduction capability of the pulse oximeter’s curve, which may not be accurate.
Testing Methodology
- Blood oxygen functional test instruments use a designated R value signal to assess the functions and reproduction abilities of pulse oximeters, referred to as the R curve.
- ISO 80601-2-61:2011 and YY0784_2010 mainly focus on clinical test methods; specific protocols for testing using blood oxygen functional test instruments are not detailed.
- The described method for testing reflective SpO2 is based on the principles of transmittance SpO2 testing methods as per the standards.
AECG100 Functional Test Instrument
- Introduction of AECG100 as a functional test instrument to evaluate wearable devices with SpO2 measurement capabilities, using Maxim's MAXREFDES103 as a test subject.
- To set up the testing environment:
- A PC with AECG100 software installed and connected to the AECG100 test system.
- The MAXREFDES103 wearable device.
- A fixture for pairing with MAXREFDES103.
MAXREFDES103 Device Overview
- MAXREFDES103 platform features:
- High-sensitivity green/red/infrared PPG biosensors, power management IC, algorithms in microcontrollers, Bluetooth for wireless communication, and a tri-axis accelerometer.
- Designed for wrist-worn usage, capturing crucial biometric signals for health care.
- Algorithms calculate heart health metrics such as heart rate (HR), SpO2, and heart rate variability (HRV).
Testing Setup for SpO2 Measurement
- Equipment description:
- Diagrams illustrate the structure of MAXREFDES103 and the design of a 3D fixture for connecting AECG100 PPG and MAXREFDES103 to test SpO2.
- The 3D fixture features openings for aligning DUT LED emissions for red/infrared light while blocking green light to facilitate SpO2 testing.
- Proper alignment ensures that PPG LED and DUT photodiodes (PD) pair effectively for accurate testing results.
Studying That Suits You
Use AI to generate personalized quizzes and flashcards to suit your learning preferences.
Description
Este cuestionario explora os estándares ISO80601-2-61:2011 e YY0784_2010 para a proba de equipos de oxímetro de pulso. A función de probas permite verificar se o equipamento se comporta segundo as especificacións do fabricante, pero a calibración clínica é esencial para obter valores de SpO2 precisos. Avalíase a importancia da comparación entre os valores amosados e os valores esperados.