10 Questions
What is the essential requirement for the sink used in the immersion method?
Receptacle that can fully immerse the item to be cleaned
What is crucial when selecting brushes for cleaning both external and internal device surfaces?
Selecting nondamaging brushes in different diameters and lengths
Why is it recommended to use brushes under water during cleaning?
To avoid splashing and aerosols
What should be done to manually washed devices that are not going through the washer-disinfector?
Place them in a drying cabinet after cleaning
Why is single-use of brushes desirable during the decontamination process?
To ensure high cleanliness standards
What is essential for effective manual cleaning according to the provided text?
Following written procedures including wearing PPE
Why are thermostatic mixer taps preferred in wash and rinse sinks?
To control temperature in the sinks
What does the manufacturer need to provide regarding brush size?
Information about the appropriate brush size
What should be used if a drying cabinet is not available after manually washing devices?
A clean, disposable lint-free, absorbent wipe
What is recommended for cleaning brushes used for manual cleaning if they are not single-use?
Decontaminate them at least daily
Learn about the decontamination zone and the processes and equipment required for thorough cleaning. Understand the importance of systematically decontaminating all RMDs to prevent infection after each surgical procedure.