Module 8: Decontamination Process Overview

HardyLeprechaun avatar
HardyLeprechaun
·
·
Download

Start Quiz

Study Flashcards

10 Questions

What is the recommended number of sinks for cleaning devices?

Three sinks

What should be the pH range of the mild detergent preferred for manual cleaning?

8.0–11.5

What must never be used for cleaning RMD?

Household soap

How should lumened devices be flushed to avoid splashing and aerosols?

Discharging under water with a jet-gun

What is the purpose of immersing the item in the cleaning solution?

To ensure the cleaning solution reaches all surfaces

Which type of water is recommended for the final rinse of the item?

Warm-to-hot water

What should be done before attempting to remove stubborn staining from devices?

Remove visible soiling first

When should an enzymatic detergent be considered for cleaning devices?

To clean devices with channels or complex parts

What is the reason for using a jet-gun to flush lumened devices?

To avoid splashing and aerosols.

What should be avoided when using a scouring pad for cleaning devices?

Using an abrasive scouring pad.

Explore the decontamination zone, equipment, and processes required for thorough cleaning in healthcare settings. Learn about the importance of systematically decontaminating reusable medical devices after each surgical procedure to prevent infections.

Make Your Own Quizzes and Flashcards

Convert your notes into interactive study material.

Get started for free

More Quizzes Like This

Decontamination SOP Quiz
16 questions

Decontamination SOP Quiz

WellMadeFreedom5205 avatar
WellMadeFreedom5205
Decontamination & Hemodialysis Concepts Quiz
15 questions
Maximizing Level C Decontamination Procedures
10 questions
Decontamination Zone Procedures Quiz
8 questions
Use Quizgecko on...
Browser
Browser