10 Questions
What is the recommended number of sinks for cleaning devices?
Three sinks
What should be the pH range of the mild detergent preferred for manual cleaning?
8.0–11.5
What must never be used for cleaning RMD?
Household soap
How should lumened devices be flushed to avoid splashing and aerosols?
Discharging under water with a jet-gun
What is the purpose of immersing the item in the cleaning solution?
To ensure the cleaning solution reaches all surfaces
Which type of water is recommended for the final rinse of the item?
Warm-to-hot water
What should be done before attempting to remove stubborn staining from devices?
Remove visible soiling first
When should an enzymatic detergent be considered for cleaning devices?
To clean devices with channels or complex parts
What is the reason for using a jet-gun to flush lumened devices?
To avoid splashing and aerosols.
What should be avoided when using a scouring pad for cleaning devices?
Using an abrasive scouring pad.
Explore the decontamination zone, equipment, and processes required for thorough cleaning in healthcare settings. Learn about the importance of systematically decontaminating reusable medical devices after each surgical procedure to prevent infections.