Questions and Answers

What is the recommended number of sinks for cleaning devices?

  • One sink
  • Two sinks
  • Three sinks (correct)
  • Four sinks

What should be the pH range of the mild detergent preferred for manual cleaning?

  • 8.0–11.5 (correct)
  • 12.0–14.0
  • 4.0–7.0
  • 1.0–3.5

What must never be used for cleaning RMD?

  • Enzymatic detergent
  • Household soap (correct)
  • Scouring pad
  • Detergent

How should lumened devices be flushed to avoid splashing and aerosols?

<p>Discharging under water with a jet-gun (A)</p> Signup and view all the answers

What is the purpose of immersing the item in the cleaning solution?

<p>To ensure the cleaning solution reaches all surfaces (C)</p> Signup and view all the answers

Which type of water is recommended for the final rinse of the item?

<p>Warm-to-hot water (A)</p> Signup and view all the answers

What should be done before attempting to remove stubborn staining from devices?

<p>Remove visible soiling first (D)</p> Signup and view all the answers

When should an enzymatic detergent be considered for cleaning devices?

<p>To clean devices with channels or complex parts (D)</p> Signup and view all the answers

What is the reason for using a jet-gun to flush lumened devices?

<p>To avoid splashing and aerosols. (C)</p> Signup and view all the answers

What should be avoided when using a scouring pad for cleaning devices?

<p>Using an abrasive scouring pad. (B)</p> Signup and view all the answers

Flashcards are hidden until you start studying

Related Documents

More Like This

SFD.002 - Cancer Reduction Initiative
26 questions
Decontamination SOP Quiz
16 questions

Decontamination SOP Quiz

WellMadeFreedom5205 avatar
WellMadeFreedom5205
Decontamination Point Zone Quiz
24 questions
Use Quizgecko on...
Browser
Browser