Questions and Answers

What is the primary purpose of normalizing the linear attenuation coefficient to water when calculating Hounsfield Units (HU)?

  • To directly correlate HU values with electron density, facilitating dose calculations in radiation therapy.
  • To amplify the differences in tissue density for improved contrast.
  • To establish a consistent reference point, minimizing variability due to equipment and environmental factors. (correct)
  • To correct for beam hardening artifacts by calibrating against water's known attenuation.

A CT scan of a patient reveals a tissue with a linear attenuation coefficient very close to that of water. According to the HU scale, which value would be closest to that tissue?

  • -1000
  • 0 (correct)
  • 1000
  • -100

Why do CT scanner X-ray tubes have higher power ratings compared to those used in conventional radiography?

  • To allow for continuous or near-continuous operation, accommodating the demands of helical scanning. (correct)
  • To generate a more penetrating X-ray beam, compensating for increased tissue absorption in dense anatomical regions.
  • To utilize lower tube current, reducing the risk of overheating the anode during continuous operation.
  • To enable shorter exposure times, minimizing motion artifacts in dynamic imaging sequences.

How does the parallel arrangement of the anode-cathode axis to the z-axis of the CT scanner mitigate potential mechanical issues?

<p>It minimizes gyroscopic torque on the rotating anode, crucial for maintaining stable gantry rotation at high angular velocities. (B)</p> Signup and view all the answers

What is the primary rationale for employing higher aluminum (Al) filtration levels in CT scanners when compared to general radiography?

<p>To harden the X-ray beam, reducing beam-hardening artifacts and improving image uniformity. (D)</p> Signup and view all the answers

In the context of CT scanning, how does the persistent 'ON' state of the X-ray tube during helical scans impact tube design and heat management strategies?

<p>It requires more efficient cooling mechanisms and robust anode designs to dissipate the continuously generated heat. (C)</p> Signup and view all the answers

Which statement accurately explains how a bow-tie filter improves image quality and reduces patient dose in CT scanning?

<p>It compensates for the elliptical shape of the body, reducing the radiation intensity at the edges and providing a more uniform signal at the detectors. (B)</p> Signup and view all the answers

How does the use of a bow-tie filter in CT imaging inherently balance image quality with patient radiation dose?

<p>By reshaping the X-ray beam to uniformly expose the detector array, which reduces the need for excessive radiation to penetrate thicker body sections. (C)</p> Signup and view all the answers

In Single-Slice CT (SSCT), how is the slice thickness determined, and what physical component of the scanner is responsible for this selection?

<p>By adjusting the collimator width, which directly controls the portion of the X-ray beam incident on the patient. (B)</p> Signup and view all the answers

In modern Multi-Detector CT (MDCT) systems with >64 detectors, how is the final slice thickness determined, considering the system's ability to record detector array signals and reconstruct images?

<p>The slice thickness is governed by the detector size and configuration, but can be adjusted post-acquisition by combining signals from adjacent detectors during image reconstruction. (C)</p> Signup and view all the answers

How does adjusting the collimator width impact the functionality of Multi-Slice CT (MSCT) scanners, particularly concerning the number of detector rows utilized?

<p>It directly controls the number of detector rows exposed to the X-ray beam, effectively determining the volume of tissue scanned per rotation. (D)</p> Signup and view all the answers

In an axial CT scan, the X-ray tube state is described as 'ON only during image acquisition, OFF during table motion.' How does this intermittent operation impact the heat load on the X-ray tube compared to helical scanning?

<p>It allows for more efficient heat dissipation between acquisitions, resulting in a lower average heat load compared to the continuous operation in helical scanning. (B)</p> Signup and view all the answers

In helical CT scanning, why is image reconstruction described as requiring 'interpolation for pitch >1,' and how does this impact the resultant image quality?

<p>Because the data is acquired at different anatomical locations due to table movement, requiring estimation of data points to create axial images, potentially reducing image accuracy. (C)</p> Signup and view all the answers

In the context of helical CT scanning, what is the precise definition of 'pitch,' and how does it quantitatively relate table movement to beam collimation?

<p>Pitch is the ratio of table travel per gantry rotation to the beam width, indicating the degree of helical overlap or extension. (B)</p> Signup and view all the answers

In helical CT, a pitch value of less than 1 (Pitch < 1) is often described as 'over-scanning the patient.' How does this technique influence image quality and patient radiation exposure?

<p>Improving image quality through increased data redundancy, but increasing patient radiation exposure due to overlapping scans. (B)</p> Signup and view all the answers

How does increasing the pitch in a helical CT scan directly impact the quality of the reconstructed images, particularly concerning the trade-off between scan speed and data integrity?

<p>It allows for faster scan times, but necessitates increased interpolation during reconstruction, which may lead to reduced image accuracy and potential artifacts. (D)</p> Signup and view all the answers

What is the relationship between pitch selection in helical CT scanning and the degree of interpolation required during image reconstruction, and how does this impact the diagnostic quality of the scan?

<p>Higher pitch values necessitate more interpolation, potentially compromising image accuracy and increasing the risk of diagnostic errors. (C)</p> Signup and view all the answers

How does decreasing the pitch affect patient dose and image quality in a helical CT scan, assuming all other parameters remain constant?

<p>It increases patient dose and improves image quality due to less need for interpolation. (A)</p> Signup and view all the answers

In the design and operation of CT scanners, what inherent trade-off exists between achieving higher image quality and minimizing patient radiation dose, and how is this balance typically managed?

<p>There is an inverse relationship; attempts to improve one aspect often compromise the other. Manufacturers typically use techniques like automatic exposure control and iterative reconstruction to find an acceptable balance. (C)</p> Signup and view all the answers

Considering the interplay between pitch, patient dose, and image quality in helical CT, which of the following strategies balances these factors for optimal clinical utility?

<p>Selecting a moderate pitch value which balances scanning time, interpolation requirements, and radiation exposure, while optimizing tube current and voltage based on patient size and tissue density. (C)</p> Signup and view all the answers

Flashcards

What is Hounsfield Unit (HU)?

HU or CT number is a normalised (to that of water) linear attenuation coefficient.

Typical CT numbers for air, water, and bone?

On a typical scale of -1000 to +1000, CT number are: Air: -1000, Water: 0, Bone: +1000

Power of X-ray tubes in CT scanners

X-ray tubes in CT scanners have much more power than tubes used for radiography or fluoroscopy.

Orientation of X-Ray tube in CT gantry

The X-ray tube is mounted onto the gantry such that the anode-cathode axis runs parallel to the z-axis of the scanner.

Signup and view all the flashcards

X-ray tube state in helical CT

In a helical scan, the X-ray tube is always ON; hence, significantly higher heat output is generated

Signup and view all the flashcards

Bow-tie filter in CT and its purpose?

A beam shaping filter, known as a bow-tie filter, is used on whole-body CT scanners because Most body areas scanned by a CT scanner are approximately circular/elliptical in cross-section, this gives rise to a high x-ray fluence at the edges compared to the centre.

Signup and view all the flashcards

Slice selection in Single-Slice CT (SSCT)

For SSCT, the slice width is selected by adjusting the collimator width.

Signup and view all the flashcards

Slice selection in Multi-Detector CT (MDCT)

For a multidetector CT, the thickness of the slice is governed by detector size and configuration.

Signup and view all the flashcards

Pitch (Helical CT)

Pitch = table travelled per rotation/beam width

Signup and view all the flashcards

High Pitch Effect

Higher the Pitch: the faster the scan, the smaller the patient's dose, but interpolation required during reconstruction may result in inaccurate data, i.e., poor image quality

Signup and view all the flashcards

Low Pitch Effect

Lower the Pitch; the slower the scan, the higher the patient dose.Less interpolation is needed, hence better image quality

Signup and view all the flashcards

Study Notes

CT Number Definition

  • HU or CT number represents a normalized linear attenuation coefficient relative to water
  • HU = [µ-µw/µw] x1000 , where µ is the linear attenuation coefficient and µw is the coefficient for water
  • Typical CT numbers are: Air = -1000, Water = 0, Bone = +1000

CT Number for Fat Tissue

  • Given a linear attenuation coefficient of fat = 0.09 cm^-1 and water = 0.1 cm^-1, the CT number calculation is:
  • CT number = [0.09-0.1 /0.1] x 1000 = -100

X-Ray Tube in CT Scanners

  • X-ray tubes in CT scanners have higher power compared to those used in radiography or fluoroscopy
  • The X-ray tube rotates at high angular velocities within the CT gantry, up to 5 rotations per second
  • The anode-cathode axis is parallel to the z-axis of the scanner, thus aligning the heel effect along the patient's long axis
  • CT scanners use 80-140 kVp typically
  • Higher aluminum filtration (5-10mm Al) is used compared to general radiography, resulting in hardened X-ray spectra
  • In helical scans, the X-ray tube is always ON, generating significantly higher heat output
  • Modern tubes have a rotating body for higher heat capacity, rather than only a rotating anode

Bow Tie Filter Function

  • Most body areas scanned by CT are approximately circular or elliptical in cross-section
  • X-ray fluence is higher at the edges compared to the center due to the shape
  • A bow-tie filter is a beam shaping filter used on whole-body CT scanners
  • It attenuates more towards the periphery, resulting in a uniform signal level at the detector
  • Bow tie filters reduce patient dose without sacrificing image quality

Slice Selection in SDCT and MDCT

  • Single Detector CT (SDCT): Slice width is selected by adjusting the collimator width
  • Multi-Detector CT (MDCT): Slice thickness is governed by detector size and configuration
  • Modern CT scanners (>64 detectors) record detector array signals, enabling highest resolution scans
  • Signals from adjacent detectors combine during reconstruction to create varying slice thicknesses e.g., 64-slice CT with 0.25mm width can reconstruct slices from 0.25mm to 16mm

Axial vs. Helical Scans

Feature Axial Helical
X-ray tube ON during image acquisition, OFF during table motion Always ON, continuous image acquisition
Table Stationary during image acquisition Moving during image acquisition
Speed Fixed speed Faster
Dose High Less patient dose possible
Reconstruction No interpolation required Interpolation required for pitch >1

Pitch Definition (Helical Scan)

  • Pitch is the ratio of table travel per rotation to beam width

Pitch Values

  • Pitch < 1 means over-scanning the patient
  • Pitch = 1 is similar to the axial scan
  • Pitch > 1 means under-scanning the patient

Pitch Calculation Example

  • Given table moves 100mm/sec, gantry rotation time is 0.5sec, and beam width is 25mm
  • Table moves 50mm per rotation (0. 5 sec x 100mm/sec)
  • Pitch = 50/25 = 2

Pitch with Faster Gantry Rotation

  • Assuming gantry rotation time of 0.25 seconds, the table moves 25mm per rotation
  • Pitch = 25/25 = 1

Pitch Effects on Dose and Image Quality

  • Higher Pitch: Faster scan, smaller patient dose, but potential missing data
  • Higher Pitch requires more interpolation during reconstruction, which can lead to inaccurate data and lower image quality
  • Lower Pitch: Slower scan, higher patient dose, requires less interpolation, and gives better image quality

Studying That Suits You

Use AI to generate personalized quizzes and flashcards to suit your learning preferences.

Quiz Team

Related Documents

More Like This

CT Number and Loading Factors Accuracy
24 questions
CT Motorcycle Crossword Puzzle Quiz
31 questions
ARRT CT Registry Flashcards
18 questions

ARRT CT Registry Flashcards

ResponsiveKazoo9793 avatar
ResponsiveKazoo9793
Use Quizgecko on...
Browser
Browser