What is the primary purpose of normalizing the linear attenuation coefficient to water when calculating Hounsfield Units (HU)?
What is the primary purpose of normalizing the linear attenuation coefficient to water when calculating Hounsfield Units (HU)?
- To directly correlate HU values with electron density, facilitating dose calculations in radiation therapy.
- To amplify the differences in tissue density for improved contrast.
- To establish a consistent reference point, minimizing variability due to equipment and environmental factors. (correct)
- To correct for beam hardening artifacts by calibrating against water's known attenuation.
A CT scan of a patient reveals a tissue with a linear attenuation coefficient very close to that of water. According to the HU scale, which value would be closest to that tissue?
A CT scan of a patient reveals a tissue with a linear attenuation coefficient very close to that of water. According to the HU scale, which value would be closest to that tissue?
- -1000
- 0 (correct)
- 1000
- -100
Why do CT scanner X-ray tubes have higher power ratings compared to those used in conventional radiography?
Why do CT scanner X-ray tubes have higher power ratings compared to those used in conventional radiography?
- To allow for continuous or near-continuous operation, accommodating the demands of helical scanning. (correct)
- To generate a more penetrating X-ray beam, compensating for increased tissue absorption in dense anatomical regions.
- To utilize lower tube current, reducing the risk of overheating the anode during continuous operation.
- To enable shorter exposure times, minimizing motion artifacts in dynamic imaging sequences.
How does the parallel arrangement of the anode-cathode axis to the z-axis of the CT scanner mitigate potential mechanical issues?
How does the parallel arrangement of the anode-cathode axis to the z-axis of the CT scanner mitigate potential mechanical issues?
What is the primary rationale for employing higher aluminum (Al) filtration levels in CT scanners when compared to general radiography?
What is the primary rationale for employing higher aluminum (Al) filtration levels in CT scanners when compared to general radiography?
In the context of CT scanning, how does the persistent 'ON' state of the X-ray tube during helical scans impact tube design and heat management strategies?
In the context of CT scanning, how does the persistent 'ON' state of the X-ray tube during helical scans impact tube design and heat management strategies?
Which statement accurately explains how a bow-tie filter improves image quality and reduces patient dose in CT scanning?
Which statement accurately explains how a bow-tie filter improves image quality and reduces patient dose in CT scanning?
How does the use of a bow-tie filter in CT imaging inherently balance image quality with patient radiation dose?
How does the use of a bow-tie filter in CT imaging inherently balance image quality with patient radiation dose?
In Single-Slice CT (SSCT), how is the slice thickness determined, and what physical component of the scanner is responsible for this selection?
In Single-Slice CT (SSCT), how is the slice thickness determined, and what physical component of the scanner is responsible for this selection?
In modern Multi-Detector CT (MDCT) systems with >64 detectors, how is the final slice thickness determined, considering the system's ability to record detector array signals and reconstruct images?
In modern Multi-Detector CT (MDCT) systems with >64 detectors, how is the final slice thickness determined, considering the system's ability to record detector array signals and reconstruct images?
How does adjusting the collimator width impact the functionality of Multi-Slice CT (MSCT) scanners, particularly concerning the number of detector rows utilized?
How does adjusting the collimator width impact the functionality of Multi-Slice CT (MSCT) scanners, particularly concerning the number of detector rows utilized?
In an axial CT scan, the X-ray tube state is described as 'ON only during image acquisition, OFF during table motion.' How does this intermittent operation impact the heat load on the X-ray tube compared to helical scanning?
In an axial CT scan, the X-ray tube state is described as 'ON only during image acquisition, OFF during table motion.' How does this intermittent operation impact the heat load on the X-ray tube compared to helical scanning?
In helical CT scanning, why is image reconstruction described as requiring 'interpolation for pitch >1,' and how does this impact the resultant image quality?
In helical CT scanning, why is image reconstruction described as requiring 'interpolation for pitch >1,' and how does this impact the resultant image quality?
In the context of helical CT scanning, what is the precise definition of 'pitch,' and how does it quantitatively relate table movement to beam collimation?
In the context of helical CT scanning, what is the precise definition of 'pitch,' and how does it quantitatively relate table movement to beam collimation?
In helical CT, a pitch value of less than 1 (Pitch < 1) is often described as 'over-scanning the patient.' How does this technique influence image quality and patient radiation exposure?
In helical CT, a pitch value of less than 1 (Pitch < 1) is often described as 'over-scanning the patient.' How does this technique influence image quality and patient radiation exposure?
How does increasing the pitch in a helical CT scan directly impact the quality of the reconstructed images, particularly concerning the trade-off between scan speed and data integrity?
How does increasing the pitch in a helical CT scan directly impact the quality of the reconstructed images, particularly concerning the trade-off between scan speed and data integrity?
What is the relationship between pitch selection in helical CT scanning and the degree of interpolation required during image reconstruction, and how does this impact the diagnostic quality of the scan?
What is the relationship between pitch selection in helical CT scanning and the degree of interpolation required during image reconstruction, and how does this impact the diagnostic quality of the scan?
How does decreasing the pitch affect patient dose and image quality in a helical CT scan, assuming all other parameters remain constant?
How does decreasing the pitch affect patient dose and image quality in a helical CT scan, assuming all other parameters remain constant?
In the design and operation of CT scanners, what inherent trade-off exists between achieving higher image quality and minimizing patient radiation dose, and how is this balance typically managed?
In the design and operation of CT scanners, what inherent trade-off exists between achieving higher image quality and minimizing patient radiation dose, and how is this balance typically managed?
Considering the interplay between pitch, patient dose, and image quality in helical CT, which of the following strategies balances these factors for optimal clinical utility?
Considering the interplay between pitch, patient dose, and image quality in helical CT, which of the following strategies balances these factors for optimal clinical utility?
Flashcards
What is Hounsfield Unit (HU)?
What is Hounsfield Unit (HU)?
HU or CT number is a normalised (to that of water) linear attenuation coefficient.
Typical CT numbers for air, water, and bone?
Typical CT numbers for air, water, and bone?
On a typical scale of -1000 to +1000, CT number are: Air: -1000, Water: 0, Bone: +1000
Power of X-ray tubes in CT scanners
Power of X-ray tubes in CT scanners
X-ray tubes in CT scanners have much more power than tubes used for radiography or fluoroscopy.
Orientation of X-Ray tube in CT gantry
Orientation of X-Ray tube in CT gantry
Signup and view all the flashcards
X-ray tube state in helical CT
X-ray tube state in helical CT
Signup and view all the flashcards
Bow-tie filter in CT and its purpose?
Bow-tie filter in CT and its purpose?
Signup and view all the flashcards
Slice selection in Single-Slice CT (SSCT)
Slice selection in Single-Slice CT (SSCT)
Signup and view all the flashcards
Slice selection in Multi-Detector CT (MDCT)
Slice selection in Multi-Detector CT (MDCT)
Signup and view all the flashcards
Pitch (Helical CT)
Pitch (Helical CT)
Signup and view all the flashcards
High Pitch Effect
High Pitch Effect
Signup and view all the flashcards
Low Pitch Effect
Low Pitch Effect
Signup and view all the flashcards
Study Notes
CT Number Definition
- HU or CT number represents a normalized linear attenuation coefficient relative to water
- HU = [µ-µw/µw] x1000 , where µ is the linear attenuation coefficient and µw is the coefficient for water
- Typical CT numbers are: Air = -1000, Water = 0, Bone = +1000
CT Number for Fat Tissue
- Given a linear attenuation coefficient of fat = 0.09 cm^-1 and water = 0.1 cm^-1, the CT number calculation is:
- CT number = [0.09-0.1 /0.1] x 1000 = -100
X-Ray Tube in CT Scanners
- X-ray tubes in CT scanners have higher power compared to those used in radiography or fluoroscopy
- The X-ray tube rotates at high angular velocities within the CT gantry, up to 5 rotations per second
- The anode-cathode axis is parallel to the z-axis of the scanner, thus aligning the heel effect along the patient's long axis
- CT scanners use 80-140 kVp typically
- Higher aluminum filtration (5-10mm Al) is used compared to general radiography, resulting in hardened X-ray spectra
- In helical scans, the X-ray tube is always ON, generating significantly higher heat output
- Modern tubes have a rotating body for higher heat capacity, rather than only a rotating anode
Bow Tie Filter Function
- Most body areas scanned by CT are approximately circular or elliptical in cross-section
- X-ray fluence is higher at the edges compared to the center due to the shape
- A bow-tie filter is a beam shaping filter used on whole-body CT scanners
- It attenuates more towards the periphery, resulting in a uniform signal level at the detector
- Bow tie filters reduce patient dose without sacrificing image quality
Slice Selection in SDCT and MDCT
- Single Detector CT (SDCT): Slice width is selected by adjusting the collimator width
- Multi-Detector CT (MDCT): Slice thickness is governed by detector size and configuration
- Modern CT scanners (>64 detectors) record detector array signals, enabling highest resolution scans
- Signals from adjacent detectors combine during reconstruction to create varying slice thicknesses e.g., 64-slice CT with 0.25mm width can reconstruct slices from 0.25mm to 16mm
Axial vs. Helical Scans
Feature | Axial | Helical |
---|---|---|
X-ray tube | ON during image acquisition, OFF during table motion | Always ON, continuous image acquisition |
Table | Stationary during image acquisition | Moving during image acquisition |
Speed | Fixed speed | Faster |
Dose | High | Less patient dose possible |
Reconstruction | No interpolation required | Interpolation required for pitch >1 |
Pitch Definition (Helical Scan)
- Pitch is the ratio of table travel per rotation to beam width
Pitch Values
- Pitch < 1 means over-scanning the patient
- Pitch = 1 is similar to the axial scan
- Pitch > 1 means under-scanning the patient
Pitch Calculation Example
- Given table moves 100mm/sec, gantry rotation time is 0.5sec, and beam width is 25mm
- Table moves 50mm per rotation (0. 5 sec x 100mm/sec)
- Pitch = 50/25 = 2
Pitch with Faster Gantry Rotation
- Assuming gantry rotation time of 0.25 seconds, the table moves 25mm per rotation
- Pitch = 25/25 = 1
Pitch Effects on Dose and Image Quality
- Higher Pitch: Faster scan, smaller patient dose, but potential missing data
- Higher Pitch requires more interpolation during reconstruction, which can lead to inaccurate data and lower image quality
- Lower Pitch: Slower scan, higher patient dose, requires less interpolation, and gives better image quality