Anesthesia Equipment Maintenance
12 Questions
5 Views

Anesthesia Equipment Maintenance

Created by
@ConsiderateZinnia

Podcast Beta

Play an AI-generated podcast conversation about this lesson

Questions and Answers

A positive pressure leak could be caused by a leakage in, or, at ______ components.

device

Dow Corning ______ is to be used to lubricate O-rings.

111

The pressure relief valve is designed to open when the pressure within the breathing circuit is between +20 and +50 ______ of pressure measured on the water gage.

millimeters

The EPDM breathing hoses use ______ Rings.

<p>Bayonet</p> Signup and view all the answers

Only ______ 400 is to be used to fill the refillable cartridge.

<p>DRAGERSORB</p> Signup and view all the answers

The BG-4 constant dosage must be 1.5 to 1.9 ______ per min.

<p>Liters</p> Signup and view all the answers

The drain valve should not open at less than ______ embar.

<p>10</p> Signup and view all the answers

A fully filled steel oxygen cylinder holds ______ liters of medical oxygen.

<p>440</p> Signup and view all the answers

The minimum valve provides greater than ______ L/min flow.

<p>80</p> Signup and view all the answers

The breathing bag has a ______ liter volume.

<p>5.5</p> Signup and view all the answers

The first low pressure warning occurs when the pressure drops to approximately ______ psi.

<p>700</p> Signup and view all the answers

Medium pressure in the BG-4 is between ______ psi and 64 psi.

<p>58</p> Signup and view all the answers

Study Notes

Oxygen Cylinder and Equipment Specifications

  • A fully filled steel oxygen cylinder contains 440 liters of medical oxygen.
  • It is critical to ensure the drain valve does not open at pressures below 10 embar.
  • The BG-4 breathing apparatus weighs 15 kg (33 lbs) when fully charged.
  • The maintenance of the BG-4's battery is crucial; it should be replaced every 6 months.
  • Avoid replacing batteries in potentially explosive environments.

Pressure Measurement and Flow Rates

  • Sentinel pressure measurement accuracy is ±2% of the final value.
  • The first low pressure warning triggers at approximately 700 psi, indicating about 75% oxygen used.
  • The last low pressure warning occurs at approximately 145 psi, with roughly 95% oxygen depleted.
  • Medium pressure in the BG-4 ranges between 58 psi and 64 psi.
  • The minimum valve provides flow rates greater than 80 L/min.

Testing Procedures and Guidelines

  • Conduct supply checks on the Sentinel every 15 minutes.
  • During the constant dosage test, the RZ-25 tester should indicate a flow of 1.5 to 1.9 LPM.
  • The breathing bag has a volume of 5.5 liters.
  • Proper testing involves adjusting the zero point on the RZ-25 tester before conducting tests.
  • Use a test adapter to connect the BG-4 apparatus to the RZ-25 tester.

Maintenance and Safety

  • Rubber components must be shielded from radiation and avoid exposure to solvents like acetone and alcohol.
  • Ensure belt and harness are dry before storage to prevent mold and fungus growth.
  • A failed self-check on the device is indicated by symbol X with an error code.

Alarm and Warning Systems

  • The yellow panic button initiates the main alarm on the Sentinel.
  • First low pressure alarm emits intermittent sounds for about 30 seconds with a flashing red LED.
  • The pressure relief valve operates between +20 and +50 millimeters (+2 embar to +5 embar) on the water gauge.

Component Care and Troubleshooting

  • Use Dow Corning 111 to lubricate O-rings.
  • To prevent leaks, check the inhalation and exhalation valves for defects.
  • The proper functioning of the exhalation valve should show -10 embar and +10 embar for inhalation cycles on the pressure gauge.
  • During a positive pressure leak test, ensure pressure drop is less than 1 embar within one minute.

Cartridge and Chemical Usage

  • Only DRAGERSORB 400 should fill the refillable cartridge, which remains good for 4 years from the manufacture date.
  • Factory-packed cartridges are also viable for the full four years.
  • A positive pressure within the breathing circuit is essential to stop ambient air infiltration.

Regulatory Compliance

  • The BG-4 system is approved for use with both factory and refillable cartridges.
  • Steel cylinders are considered full at 3135 psi when a + sign is marked post hydro test.

Studying That Suits You

Use AI to generate personalized quizzes and flashcards to suit your learning preferences.

Quiz Team

Description

This quiz covers the maintenance and testing of anesthesia equipment, including procedures for replacing batteries, lubricating O-rings, and adjusting pressure relief valves. It also reviews the use of specific devices and tools, such as the RZ-25 tester and BG-4 apparatus. Test your knowledge of anesthesia equipment maintenance and troubleshooting.