BMED 420 Notes PDF
Document Details
Uploaded by Deleted User
Tags
Summary
These notes provide an introduction to biomaterials, focusing on their applications in medicine and their key characteristics. They discuss different types, including metals, polymers, and ceramics, and their biocompatibility properties. The material also touches on biomaterial interactions with the body.
Full Transcript
Introduction to Biomaterials What is a biomaterial? A biomaterial is a material used for medical applications to support, enhance, or replace damaged tissue or a biological function ○ In close contact with living tissue “Bio” = bio-compatible...
Introduction to Biomaterials What is a biomaterial? A biomaterial is a material used for medical applications to support, enhance, or replace damaged tissue or a biological function ○ In close contact with living tissue “Bio” = bio-compatible Can be natural or synthetic Can be therapeutic or diagnostic Generally implanted Applications of biomaterials Medical implants Methods to promote healing ○ Sutures, staples, etc Regenerated human tissues ○ Tissue engineering Biosensors ○ Glucose monitoring devices Drug delivery systems Important attributes of a biomaterial Biocompatibility = no harm to the host body ○ Non toxic ○ Non allergenic ○ Non thrombogenic ○ Non carcinogenic ○ Non mutagenic ○ Non inflammatory Appropriate mechanical properties for the job ○ Aim to match native tissue Four ways a biomaterial can interact with the body Hurt you Dissolve Be surrounded by a protective layer Bond/integrate with tissue Types of biomaterials: Metals ○ Composed of one or more metallic elements and small amounts of nonmetallic elements ○ Metallic bonds ○ Atoms arranged in an orderly structure ○ Good conductors of electricity and heat ○ Ductile (can deform before breaking) ○ Strong, tough, prone to corrosion Polymers ○ Broad class of compounds based on nonmetallic elements Can be natural or synthetic, mostly organic ○ Typically composed of intertwining long chains of large molecules with a carbon backbone and covalent bonds ○ Low conductivity of heat and electricity ○ Ductile, usually soft and compliant, but can be rigid ○ Low strength and mechanical properties, but good corrosion resistance Ceramics ○ Inorganic compound of metallic and nonmetallic elements ○ Can have covalent (shared electrons) or ionic bonds (transferred electrons) ○ Insulator of heat and electricity ○ Thermally stable, good for high temperature applications ○ Brittle, hard, and rigid ○ Resistant to wear, friction, and corrosion Composites ○ Composed of multiple materials Best of both worlds, but expensive ○ Usually synthetic but some natural composites exist Bone (cellulose and lignin polymers Bone (collagen and hydroxyapatite) Summary: Biomaterials must be biocompatible (not cause harm to the body) and have appropriate mechanical properties for the job Biomaterials can hurt you, dissolve, be surrounded by a proactive layer, or integrate with tissue Main types of biomaterials: metals, ceramic, polymers, and composites ○ Metals are ductile and strong but prone to corrosion ○ Ceramics are hard and resistant to corrosion but brittle ○ Polymers are composed of long chains are typically ductile and resistant to corrosion but have lower mechanical properties Metals, Ceramics, and Polymers as Biomaterials Metals as biomaterials Common metals for biomaterials: titanium, stainless steel, cobalt-chromium, nitinol, gold ○ Most other metals are not biocompatible Pros: strong, resistant to fatigue, ductile, easily sterilized Cons: subject to stress shielding, corrosion, and wear Stainless steels, cobalt, and titanium based alloys are used for orthopaedics Nitinol has shape memory Magnesium is biodegradable Metals in orthopaedics Stainless steel ○ Biocompatible and low cost but poor corrosion resistance ○ Applications: temporary devices (fracture plates, screws, nails) and some permanent implants ○ 316L is the most common ○ Can enhance corrosion resistance by alloying and metallurgical processing Cobalt-chromium alloys ○ Superior to stainless steel in corrosion resistance and fatigue strength ○ Expensive, can be toxic/allergenic, and induce stress shielding ○ Applications: load-bearing implants (hip implants, knee and ankle prostheses ○ Can alloy to increase corrosion resistance and strength Titanium ○ Best corrosion resistance and biocompatibility, light weight, less stress shielding, good bone bonding, poor bending ductility, poor wear resistance ○ Applications: permanent devices (hips, pacemakers, dental implants, bone rods, etc) Shape-memory alloys ○ Shape-memory property of nitinol is useful for self-expandable stents, clamps, and clips ○ Mixed studies on biocompatibility, concern over nickel ion release ○ Expends to original shape in body temperature Metals in dental applications Fillings, crowns, bridges, and root replacements used chromium and titanium Orthodontics like braces and retainers use nitinol Biomaterial classifications Biotolerant: accepted but encapsulated with scar tissue Bioinert: direct contact with no tissue reaction Bioactive: bonds/integrates with tissue Biodegradable: dissolves and is replaced with tissue Magnesium alloys Degradable Applications: tissue engineering and resorbable screws Safety concerns related to metal toxicity Ceramics and biomaterials Common materials: alumina, zirconia, bioglass, hydroxyapatite, calcium phosphate Pros: strong, chemically inert, high compressive strength, biodegradable or bioactive Cons: hard to manufacture, brittle, fracture, can loosen, only compressive strength Types of bioceramics Alumina and zirconia are bioinert and used for dental and hip implants Hydroxyapatite and calcium phosphate are bioactive and used for cement, coatings, and scaffolds ○ Also biodegradable and can be replaced by native tissue Polymers as biomaterials Common materials: silicone, PEEK, PTFE, polyethylene, PMMA, hydrogels, nylon, etc Pros: easy to manufacture and modify, can be bendable and biodegradable Cons: poor strength, prone to wear and tear Can be bioinert or biodegradable, usually not bioactive Smart polymers can respond to stimuli Structure of polymers Thermoplastics: linear and branched chains Thermosets: crosslinked and networked chains Composites as biomaterials Can be strong and lightweight, corrosion resistant, high cost, difficult to change shape The Body’s Response to Biomaterials Local responses to biomaterials Response at the site of the implant or device Can be caused by the material or trauma/surgery Inflammation ○ Part of your natural immune response Complex reaction of tissue to local injury Body tries to contain or neutralize injurious agent and heal the site ○ Intensity/duration of inflammation varies on size/shape/properties of material ○ Clinical features: redness, swelling, heat, pain ○ Can be acute or chronic Encapsulation - bad ○ Foreign body response refers to the immune response to the implant, ultimately leading to encapsulation An attempt to protect the body, but may affect implant function ○ Can coat or modify device surface to make this effect temporary Biointert with only a thin layer of collagenous tissue Acute inflammation: immediate and early response ○ Characterized by fluids and plasma exuding into tissue ○ Accumulation of neutrophils Chronic inflammation: longer term response ○ Leads to repair or foreign body response ○ Involves macrophages Thrombosis ○ Blood-material interaction that triggers coagulation pathways Infection ○ Bacteria can attach to the surface Can be introduced through surgery ○ May prevent tissue integration ○ Can affect device function ○ Can lead to systemic infection if it spreads into the bloodstream ○ Can add antimicrobial coatings ○ Most common serious complication of medical devices Tumorigenesis ○ Implanted material can elicit excessive and uncontrolled cell proliferation ○ Can be benign/local or malignant ○ Can invade neighboring tissues or bloodstream to become systemic Foreign body response Proteins adsorb to device surface Neutrophils try to break down the surface Macrophages degrade the surface Giant cells are formed Fibroblasts come to the site Device becomes encapsulated Systemic response to biomaterials Response throughout the body Toxicity ○ Release of wear and corrosion particles ○ Release of nickel and chromium alloys can lead to muscle pain, decline in cognitive function, memory difficulties, etc ○ Additives can be irritants and induce systemic immune reaction ○ Chemical substances in implants can be carcinogens ○ Need to minimize corrosion ○ Corrosion: destruction of metal by electrochemical reactions Galvanic corrosion: two different metals in electrical contact Hypersensitivity ○ Known as intolerance ○ Discussed in relation to metals ○ Undesirable reactions produced by normal immune system, allergies and autoimmunity Thromboembolism ○ Local blood clot breaks off and travels through circulation to block another vein/artery ○ Biggest risk is venous thromboembolism leading to pulmonary embolism Bioactive Materials and Surface Modifications Why modify surfaces? Surface features have the most impact on the biocompatibility of a material Bulk material governs mechanical properties, durability, and functionality ○ Want to maintain build properties, but modify the surface on a nanometer scale Enhance biocompatibility, increase hardness, enhance corrosion resistance Controlling protein adsorption Implanted material is immediately coated with proteins Cells attach to the proteins and interact with them Control over protein adsorption gives you control over cellular processes ○ Can prevent a foreign body response Main types of surface modifications Surface coatings ○ Increase attachment strength at tissue-implant interface ○ Increase surface area ○ Decrease friction, increase corrosion and wear resistance ○ Decrease infection Surface patterning ○ Nanoscale topographical features that give the body physical cues that allow for normal cell adhesion to form a base for normal tissue growth Surface roughening ○ Increased surface area leading to increased bioactivity Plasma modification ○ Change surface energy There are others Hydrophobic and hydrophilic surfaces Hydrophobic: water fearing, surface will not interact with water ○ High contact angle ○ Want to shed as much water as possible ○ Bind to proteins more firmly than hydrophilic surfaces ○ Self-cleaning Repel fluids Surgical tools, urine cups, diagnostic devices, drug delivery ○ Reduces risk of infection Hydrophilic: water loving, surface will interact with water ○ Low contact angle ○ Liquid evenly spreads across surface ○ Lubricity reduces force required to manipulate intravascular medical devices Can decrease frictional force, reducing risk of damage to blood vessel walls Catheters, guide wires, etc ○ Reduces thrombogenicity Determined how material interacts with the body Mechanical Properties of Materials How can we design for success? Choose materials with appropriate mechanical properties for the intended use ○ Mechanical properties for a stent should differ from a hip implant ○ Conduct mechanical testing of device under simulated anatomical conditions Choose materials with properties closest to native material properties Stress shielding and Wolff’s Law Wolff’s Law: bone will remodel based on the loads placed upon it Stress shielding: a device that is too strong with weaken surrounding bone Mechanical properties important to biomaterials Young’s modulus Ultimate tensile strength Fracture toughness Elongation at break Ultimate compressive strength Biomaterial Fatigue and Failure Fatigue Most biomaterials will need to withstand repetitive loading in the body Fatigue testing involves applying cyclic loading to determine fatigue lids and crack growth data, identify critical locations, and demonstrate safety Fatigue fracture is the main cause of premature failure in biomedical implants ○ Can arise from tiny defects that grow Non homogeneity in microstructure Manufacturing defects Under cyclic loading, a material can fracture below its UTS or yield strength Fatigue is the progressive structural damage that occurs under cyclic loading ○ Under static loading fracture can only occur after UTS Failure Tend to be associated with tensile and bending instead of compressive loads Damage is irreversible Influenced by temperature, surface finish, oxidizing chemicals, etc Fretting fatigue = cyclic stress + friction ○ Normal stress and shear stress between components against bone Corrosion fatigue = cyclic stress + corrosion ○ Development of brittle cracks leading to growth and failure Wear Caused by friction Can lead to aseptic loosening ○ Generation of tiny particles around the device ○ Particles that attract macrophages ○ Dying macrophages break down and release acidic enzymes that can cause erosion Concerns over toxicity of metal wear particles and ions Wear resistance is closely controlled by hardness ○ Can improve with surface modification like zirconia coating Hardness Measures local deformation to an indenter ○ Hard materials are resistant to wear and direction Most popular tests are Brinell and Rockwell Hardness is roughly proportional to tensile strength ○ Can estimate tensile strength without a tensile test Brinell hardness measures the diameter of penetration indenter Rockwell hardness measures the depth of penetration of diamond indenter In Vitro Testing of Biomaterials Evaluation of biomaterials: Must validate: ○ Material properties Material testing ○ Biological effects Bench (in vitro) Animal models (in vivo) Clinical trials Post-implantation follow-up studies In vitro: in a test tube or outside an organism In vivo: in an organism Ex vivo: in tissues explained from an organism In situ: in original position In vitro testing Use testing standards (ISO and ASTM) ○ Ensures safety and effectiveness ○ Quick turnover, high throughput screening, low cost Compare with SRM (standard reference material/control) ○ SRM should be clinical gold standard biomaterial for that application Cytotoxicity testing ISO 10993-5: Tests for Cytotoxicity ○ Most important in vitro test Prescribed for every type of medical device Good first step, gives initial biocompatibility Tests are sensitive, cheap, and quick ○ Negative: free of harmful leachables, or insufficient amount to cause acute effect ○ Positive: warning sign, material could have harmful leachables, need more tests Cell monolayers are grown to 70-80% confluence and exposed to test and control samples ○ Contact method: apply material directly, used for patches/contact devices ○ Elution method: uses a solvent to extract one material from another, more relevant Signs of cell toxicity: morphological changes and cell death Hemocompatibility ISO 1993-4: Selection of Tests for Interaction with Blood ○ No standard methods that can predict hemocompatibility across devices and applications ○ Based on blood contact category, it recommends tests for thrombosis, coagulation, platelet function, hematology, and immunology Cytotoxicity to blood can cause significant harm ○ Hemolysis: breakdown of red blood cells impairs oxygen transport ○ Adverse reactions with white blood cells impair body’s ability to fight pathogens ○ Adverse reactions with cells involved in clotting can lead to blood device Genotoxicity testing ISO 10993-3: Tests for Genotoxicity, Carcinogenicity, and Reproductive Toxicity ○ Required for devices in contact with the body for >30 days ○ Assay are done to ensure material is not a: Genotoxic mutagen: agent that changes genetic material (DNA) Carcinogen: agent that causes uncontrolled proliferation of host cells Chromosomal aberration and DNA effects tests Can be in vitro or in vivo Chromosomal aberration tests: detects chromosomal damage DNA effects tests: detects DNA damage Cell adhesion testing Want adhesion of some cells Don’t want adhesion of inflammatory cells Cells do not interact with material, they interact with proteins adsorbed to the material surface Cells culture on biomaterials can be visualized using microscopy In vivo testing In vitro assays do not reproduce the complexity of an in vivo situation ○ Might find that a new biomaterial coating does not improve cell attachment in vitro but does in vivo Both in vitro and in vivo testing must be done In Vivo Testing of Biomaterials In vivo testing Must follow testing standards and compare with SRMs Goal is to take measurements of cell and tissue-level responses to an implanted biomaterial Must: ○ Provide an environment that closely matches the biological environment in which the material will be used ○ Provide quantifiable outcome measures ○ Detect and precinct clinically relevant differences across materials tested Implantation testing ISO 10993-6: Tests for Local Effects Following Implantation Implant material in muscle and examine local reaction at site ○ Presence of inflammatory cells ○ Extent of fibrosis ○ Scored quantitatively using ISO’s scoring recommendations Set implant intervals to time of human exposure or show the reaction has stabilized Irritation and sensitivity testing ISO 10993-10: Tests for Irritation and Sensitization ○ 3 in vivo tests Primary skin test (material placed on shaved back of rabbits for 4-24 hours) Ocular irritation test (material placed in rabbit’s eye) Intracutaneous injection (material injected into rabbit’s skin) System toxicity testing ISO 10993-11: Tests for Systemic Toxicity Acute toxicity of metal ions is tested using animal models ○ Most injection test (sensitization) ○ Oral and dermal tests (irritation/inflammation) ○ Pyrogen test (fever response) Subchronic testing: tissue exposure or repeated treatment up to 10% of animal lifespan ○ Necessary for devices in contact with patients for 1-30 days Chronic testing: tissue exposure or repeated treatment for less than 10% of animal lifespan Selection of animal models Select animal model based on device location and function ○ Sheep and goats closest to human bone pathology Variables in animal testing Species, age, gender ○ Ovariectomized rodents are used to mimic endocrine status of postmenopausal women Diet ○ Calcium-deficient diet mimics postmenopausal nutritional state in women that leads to osteoporosis Size of defect ○ Critical-sized defect that won’t heal on its own, varies among species and sites Ethical issues Increasing concern for well-being of animals Duty to ensure animals are treated humanely Research centers have committees to maintain approval for animal trials While R&D can begin in vitro, animal studies are required to assess safety and efficacy The 3 R’s Replace ○ Replace animal studies with other methods ○ Full replacement: avoids using any animals Includes use of human volunteers, tissues, cells, and computer models ○ Partial replacement: includes use of animals that are not capable of suffering Reduce ○ Reduce number of trials on animals ○ Appropriately design experiments to ensure robust and reproducible findings ○ Maximize information gathered per animal Balance against additional suffering caused by repeated use ○ Sharing data and resources between research groups Refine ○ Minimize stress of animals ○ Use humane housing, feeding protocols, handling, etc ○ Use appropriate anesthesia to minimize pain ○ Train animals to cooperate with procedures ISO 10993-2: Animal Welfare Requirements ○ Specifies minimum requirements to ensure that proper provision has been made for welfare of animals in biocompatibility testing ○ Offers guidance on the 3 R’s Degradable biomaterial considerations Often used as scaffold material for tissue engineering Must be biocompatible and degradation products should be non-toxic Degradation rate should be tuned to match formation rate of functional tissue The soluble biomaterial degradation products are transported via lymphatic system to kidneys then excreted Degradation of polymers Can degrade due to heat, light, or chemicals ○ Degradation is often intentional Long polymer chains break down to shorter segments and lose structural/chemical properties, leading to cracking and disintegration Hydrolytic degradation Hydrolysis: breakdown of material due to reaction with water ○ Bonds are cleaved Occurs in polymers with hydrolyzable groups along the main chains Polyesters are the most common biodegradable polymers that undergo hydrolysis Benefits of biodegradable polymers No requirement of second surgery Eliminate chronic inflammation Deliver drugs through degradation Useful for tissue engineering Factors that affect rate of degradation Nature of chemical bond pH Polymer composition Extent of water uptake Hydrophilicity of material ○ Hydrophobic has slower degradation, less water diffusion Molecular weight of polymer Bulk vs. surface degradation Degradation leads to erosion Surface erosion: polymer degrades from exterior surface ○ Water erodes surface Bulk erosion: degradation occurs throughout whole material equally ○ Strength decreases more quickly overtime Many materials undergo a combination Ceramic degradation Most common: calcium phosphates and bioactive glass ○ Unique bone-bonding properties ○ Can modulate degradation profile by mixing different ceramics Metal degradation Magnesium is the main biodegradable material ○ Light, biocompatible ○ Degrades too fast for some applications; can apply calcium phosphate coatings to control degradation rates to match the kinetics of bone healing Regulatory Pathways FDA organization Office of Medical Products and Tobacco ○ CBER: Center for Biologics Evaluation and Research ○ CDER: Center for Drug Evaluation and Research ○ CDRH: Center for Devices and Radiological Health Classes of medical devices Class I: low risk ○ Bandages, stethoscopes, gloves, masks, tongue depressors, toothbrushes, etc Class II: moderate risk ○ Blood pressure cuffs, pregnancy tests, syringes, blood transfusion devices, contacts, etc Class III: high risk ○ Stents, implants, pacemakers, etc FDA approval pathways Exempt devices 93% of Class I devices and 8% of Class II devices are exempt from 510(k) Manufacturer is responsible for: ○ Good manufacturing processes (GMP) ○ Quality assurance (QA) ○ Documentation 510(k) premarket notification Most common pathway 92% of Class II devices are subject to 510(k) Manufacturer is responsible for: ○ Demonstrating substantial equivalence through performance testing Mechanical testing Sometimes bench or animal testing, only some require clinical data Special controls include device specific performance standards, labeling, data, guidelines, etc Must demonstrate the device is as safe and effective as an existing legally marketed device FDA decides within 90 days if it’s substantially equivalent (SA) ○ Otherwise, goes through pre-market approval (PMA) route Device testing for 510(k) Common: ○ Bench testing Electrical/software Mechanical ○ Biocompatibility ○ Sterility and shelf life Much less common ○ Animal and clinical testing Types of FDA review decisions Substantially equivalent (SE) ○ Clearance to market a new device Unable to determine ○ May be required to submit additional information Not substantially equivalent (NSE) ○ Sponsor may appeal decision ○ Submit a PMA ○ Submit a new 510(k) ○ Initiate De Novo Process Pre-market approval (PMA) Almost all Class III devices Manufacturer is responsible for: ○ Demonstrating reasonable assurance of safety and effectiveness ○ Performance testing ○ Bench, animal, and clinical testing ○ Post-approval requirements De Novo classification If 510(k) review leads to NSE decision, it is automatically Class III Criteria: clinical benefit, simple technology, no predicate, low-moderate risk Regulatory guidance Several resources exist to help guide you through the process to safely and effectively develop a medical device Main documents: FDA CFR 21, ISO 13485:2016, etc Things can go wrong Infections, corrosion, mechanical failure, etc May lead to litigation cases and recalls ○ Medical device records (MDR): information on recalls ○ MAUDE database: individual medical reports All medical devices carry some level of risk ○ Important to identify root cause of risk Device factors, surgical factors, patient factors, use factors, etc Key takeaways FDA classifies device as I, II, or III ○ Most class I devices are exempt; FDA registered/listed ○ Most class II devices go through the 510(k) pathway; FDA clearance ○ Most class III devices need pre-market approval (PMA); FDA approval Orthopaedics: Anatomy and Pathologies Musculoskeletal system Made up of bones, muscles, cartilage, tendons, ligaments, joints, and other connective tissue that attached tissue and holds organs together ○ Tendons connect muscle to bone ○ Ligaments connect bone to bone Which fails first? The majority of musculoskeletal problems are related to our joints Common problems with aging: ○ Cartilage starts to wear away Osteoarthritis ○ Tissues become stiffer and less flexible ○ Fluid in joints and bone density decrease Composition of musculoskeletal tissues Cells surrounded by ECM ECM is made up of different amounts of: ○ Collagen (types I-V) (protein) ○ Proteoglycan (non-collagenous protein) ○ Water ○ Elastin (extensible tissues) ○ Hydroxyapatite (bone only) Structure of bone Cortical (compact) ○ 80% of total bone mass of the body ○ 3-4% porosity Trabecular (cancellous/spongy) ○ 30-90% porosity ○ More active in remodeling than cortical bone Bone cells Osteoclasts resorb bone Osteoblasts synthesize bone ○ Service from osteogenic stem cells Osteocytes are formed from osteoblasts and maintain bone Types of cartilage Elastic cartilage (most flexible) ○ Ear, auditory tube, larynx Hyaline cartilage (second most flexible) ○ Articular cartilage in joints, cartilage end plate in spine, nose, lungs Fibrocartilage (least flexible) ○ Annulus fibrosus (disc), between pubic bones, resists tension and compression Types of joints Fibrous: Cartilaginous: Synovial: Disc anatomy Resist loading in many directions Lamellae layered at alternating 30 degree angles Annulus fibrosus (outer region) ○ 15% type I and II collagen ○ 70% water ○ 5% proteoglycan ○ Encases nucleus and resist multi-axial loading on the spine Nucleus pulposus (center of disc) ○ 4% type II collagen ○ 77% water ○ 14% proteoglycan ○ Proteoglycan aggregates draw water into disc creating pressure to maintain height and resist compressive loading Cartilage endplate ○ 25% type I collagen ○ 55% water ○ 8% proteoglycan ○ 0.1-2 mm thick ○ Resists compression and allows for nutrient transport Largest avascular structure in the body ○ Needs to get nutrients from surroundings Back pain The most common musculoskeletal condition, affects 80% of people 90% of back pain is nonspecific Lack of accurate diagnosis has implications for treatment Most commonly stems from intervertebral disc pathologies Disc pathologies Degenerated: Bulging: Herniated: Thinning: Osteophytes: Hip anatomy and pain Ball and socket joint ○ Head of femur fits into acetabulum Hip fracture ○ Often caused by osteoporosis Osteoarthritis ○ Most common reason for hip replacement Knee anatomy and pain Femur, tibia, and fibula ○ Connected by ACL and PCL ACL tear is the most common knee injury Meniscus damage and wear Shoulder anatomy and pain Humerus attached to scapula (shoulder blade) and clavicle (collar bone) Shoulder pain causes: ○ Rotator cuff tear ○ Calcific tendonitis ○ Bicep tendonitis ○ Shoulder osteoarthritis Orthopaedics: Hardware Solutions Types of orthopaedic hardware Screws, plates, pins Wires, rods, nails, suture anchors Spinal cages and joint replacements Implants used in spinal fusion Screws and rods ○ Rods stabilize vertebral segments ○ Screws act as anchor points for rods Placed at 2 more consecutive spine segments ○ Most screws and rods are permanent, but some can be removed after bone graft grows Cages/disc spacers ○ Cages have grooves, holes, or pores to allow bony ingrowth ○ Sometimes add bone graft material Hip implant combinations Metal on metal ○ Concerns with metal debris from corrosion Metal/ceramic on polyethylene ○ More wear and aseptic loosening Ceramic on ceramic ○ Inert, but expensive and brittle Suture anchors Arthroscopy: minimally invasive surgical technique that uses an endoscope and small instruments to treat soft tissues through a series of small incisions Suture anchors: used to fasten ligaments and tendons to bones via sutures First generation: stainless steel and titanium ○ Issues: loosening and migration, interference with MRI, cartilage injury Next generation: bioabsorbable polymers ○ Issues: low mechanical strength, breakage, degrade too quickly, implant debris Next generation: biodegradable composites Best of all worlds: PEEK, not biodegradable PEEK (polyetheretherketone) Strong (high holding strength) Soft ○ Less risk of cartilage damage from wear ○ Can be drilled through in event of revision surgery ○ Better manufacturability than metals Radiolucent: transparent to x-rays Carbon-reinforced PEEK anchors combine benefits of radiolucency and mechanical properties similar to bone Vented suture anchors Can promote bone marrow flow and bony ingrowth Smith & Nephew HEALICOIL PK Orthopaedics: Grafts and Tissue Engineering Tissue grafts Autograft: tissue graft of patient’s own tissue ○ Pros: No immune response or risk for rejection Promotes osteogenesis Stronger than tissue that has been sterilized ○ Cons: Cell culture takes time Requires second surgical site Often inadequate amount of tissue can be harvested Allograft: tissue graft from donor of same species ○ Pros: Unlimited supply No size limitation Shorter operation time ○ Cons: Immune response Delayed incorporation Sterilization makes tissue weaker Xenograft: tissue graft from donor of different species Allogeneic response Potent immune response that can cause allograft rejection T cells = part of immune system that focus on foreign particles ○ Cells that attack already-infected cells (transplanted cells) Autografts for spinal fusion Bone graft, in addition to disc spacer, provides foundation for new bone to grow (fusion) Gold standard = autograft ○ Greater fusion success ○ Harvest from iliac crest, rib, or spine ○ Can use a combination of trabecular bone and cortical Allografts for spinal fusion Generally frozen or freeze dried to limit chance of graft rejection Do not contain living bone cells, not as effective at stimulating fusion Bone graft substitutes/extenders for spinal fusion Demineralized bone matrix (DBM) ○ Allograft bone where mineral content has been removed to expose bone-forming proteins Ceramic-based bone graft extenders ○ Ceramics provide matrix fusion ○ Mixed with regular bone Bone morphogenetic protein ○ Found in bone; produced in larger amounts by genetic engineering ○ Can be added to promote new bone growth Osteoconductive: provides structural matrix for bone growth Osteoinductive: stimulates new bone growth ACL grafts Autografts ○ Patellar tendon - gold standard ○ Hamstring tendon ○ Quadricep tendon Allograft ○ Achilles tendon ○ Hamstring tendon ○ Patellar tendon Z-lig xenograft ○ Porcine derived, no live cells ○ Highly scalable ○ ACL xenografts are largely unsuccessful Smart Implants and Bio-Inspired Design Smart devices Smart devices are creating a paradigm shift from episodic monitoring to continuous sensing and integrated healthcare ○ Detect problems early and provide minimally invasive management ○ Enable people to be in control of their own health Bi-directional data can be used to close the loop and tailor therapies according to feedback Smart implants take advantage of implanting hardware in the body Bio-inspired design Bio-inspired design (biomimetics): translation of knowledge obtained from the natural world into new innovations 3 main applications: ○ Implanted materials inspired by native human bone/tissue ○ Implanted materials inspired by other animal tissues/features ○ Non-medical materials inspired by native biomaterials but used for other purposes