MRI Artifacts and Compensations PDF

Document Details

TopUnity

Uploaded by TopUnity

University of Hilla

Dr.Hayder Jasim Taher

Tags

medical imaging magnetic resonance imaging MRI artifacts medical technology

Summary

This document provides an overview of artifacts in MRI, covering different types like ghosting, aliasing, and chemical shift. It also discusses the causes of these artifacts and corrective measures. The information is likely geared toward medical students or professionals.

Full Transcript

Artifacts and their compensations Dr.Hayder Jasim Taher PhD of Medical Imaging Outline of my presentation ✓Artifacts. ✓Ghosts/Motion Artifacts. ✓Aliasing/ Wraparound. ✓Chemical Shift Artifacts. ✓Gibbs or Truncation artifacts. ✓Magnetic Suscepti...

Artifacts and their compensations Dr.Hayder Jasim Taher PhD of Medical Imaging Outline of my presentation ✓Artifacts. ✓Ghosts/Motion Artifacts. ✓Aliasing/ Wraparound. ✓Chemical Shift Artifacts. ✓Gibbs or Truncation artifacts. ✓Magnetic Susceptibility Artifact. ✓Zipper Artifacts. ✓Shading Artifacts. ✓Cross Excitation. Artifacts Artifacts may be defined as the false features in the image produced during the imaging process. Artifacts can be rectified easily when the causes are known. Ghosts/Motion Artifacts Artifacts can be classified into different categories and these include the following: 1-Ghosts/Motion Artifacts: Ghosts are replica of something in the image. Ghosts are produced by body part moving along a gradient during pulse sequence resulting into phase mis mapping. Ghosts can originate from any structure that moves during the acquisition of data (Fig. 1). Periodic movement such as respiratory, cardiac and vessel pulsation causes ghosts while nonperiodic movement causes a smearing of the image. Ghosts/Motion Artifacts - Axis: ghosts almost always seen along phase encoding axis. - Corrective measures: A-Patient motion: Make patient lie comfortably, stabilize, with straps and cushions, it is important not to use excessively long sequences, as movement for a brief period spoil all the images. B-Cardiac motion This type of artifact is caused by the contraction and relaxation of heart (chest) while the scanning is going on. To avoid this type of artifact, cardiac gating is mandatory during the procedure. C-Respiratory motion This type of artifact is caused by respiration during the scanning. This can be avoided by respiratory gating and respiratory compensation. Ghosts/Motion Artifacts D-Blood flow motion This type of artifact is caused by the flow of blood throughout the cardiac cycle. The artifact is prominent in axial images. An effective remedy for blood flow motion artifact is ‘Spatial Presaturation (SAT) which is a new technique for suppressing these artifacts. Ghosts/Motion Artifacts In this technique a spectrally shaped radiofrequency pulse has been applied to selectively saturates spins located in regions outside the image volume. Note: Another way to reduce the effect of the motion artifacts is by the using of saturation band. - Saturation Band It is also called REST slab or SAT band and is used to suppress the signal from a part of the FOV (Fig. 2). A 90 degrees RF pulse is sent immediately prior to proper sequence tilting the magnetization in transverse plane in the region of the band. When proper excitation pulse follows there is no longitudinal magnetization in this region to be tilted. Hence this region will not have any signal. Artifacts can be classified Fig.1: Ghosting/ Fig. 2: Image showing saturation Movement artifacts band anterior to the spine. Aliasing/ Wraparound 2-Aliasing/ Wraparound: In aliasing, anatomy that exists outside the FOV appears within the image and on the opposite side (Fig. 3). When the imaging field of view is smaller than the anatomy being imaged, aliasing occurs. -Axis: aliasing can occur along any axis. Aliasing along frequency encoding axisis called frequency wrap and along phase encoding axis is called phase wrap. Aliasing can occur along slice selection axis in 3D imaging. -Corrective measures: A-Increase FOV. B-Filtering the frequency encoded direction. C-Phase wrap can be corrected by increasing FOV along phase encoding direction. Aliasing/ Wraparound Fig 3: Aliasing artifact Chemical Shift Artifacts 3-Chemical Shift Artifacts: Chemical shift artifacts appear at the interfaces between water and fat because the precessional frequency of protons is slightly different in these two substances. This difference in precessional frequencies of protons in water and fat is called ‘chemical shift’. It is expressed in parts permillion (ppm). The frequency of water protons is about 3.5 ppm greater than that of fat protons. This chemical shift of 3.5 ppm causes water protons top recess at a frequency 220 Hz higher than that of fat proton at 1.5 Tesla. This leads to misregistration of the signals. They are displayed by the equipment as dark region of signal void on one side of water containing tissue and a region of bright signal at the other end of the water fat interface due to super imposition of fat and water signals on the frequency encoding direction. The chemical shift artifacts are commonly noticed in the abdomen, spine and orbits where fat and other tissues from boarders. This artifact is greater at higher field strength.(fig.4). -Axis: Frequency encoded direction. -Corrective measures: The only way to eliminate this artifact is to use a fat suppression technique. Chemical Shift Artifacts Fig 4: Chemical shift artifact Gibbs or Truncation artifacts 4-Gibbs or Truncation artifacts: Gibbs or Truncation artifacts are bright and dark lines that are seen parallel and adjacent to boarders of abrupt intensity change, as many be seen at CSF, spinal cord, fat and muscle. (fig.5) -Axis: Phase encoding direction. -Corrective measures: Gibbs artifact can be reduced by increasing the matrix and using a filter. Gibbs or Truncation artifacts Fig 5: Gibbs artifact Magnetic Susceptibility Artifact 5 -Magnetic Susceptibility Artifact: Magnetic susceptibility is the ability of a substance to become magnetized. Some tissues magnetize to different degree than other, resulting into differences in precessional frequency and phase. This causes dephasing at the interface of these tissues and signal loss. For example, magnetic susceptibility difference between soft tissues and air is about 10 ppm. This causes signal loss and distortion of the boundaries of the brain near air sinuses. Other common causes of magnetic susceptibility artifacts include metal. (fig.6) -Axis: Frequency encoding and phase encoding. -Corrective measures: Use of SE sequence and remove the metal. Magnetic Susceptibility Artifact Fig. 6: Susceptibility artifact Zipper Artifacts 6-Zipper Artifacts: This artifact is caused by external RF entering the roomat a certain frequency and interfering with inherently weak signal coming from the patient. There are various causes for zipper artifacts in images. Most of them are related to hardware or software problems. The zipper artifacts that can be controlled easily are those due to RF entering thescanning room when the door is open during acquisition of images. RF from radio transmitters will cause zipper artifacts that are oriented perpendicular to the frequency axis of the image. Fig.7 -Axis: Perpendicular to the frequency axis of the image. -Corrective measures: System generated artifacts should be reported service engineer. Zipper artifact Fig 7: Zipper artifact Shading Artifacts 7 -Shading Artifacts: In shading artifact image has uneven contrast with loss of signal intensity in one part of the image (Fig. 8). The causes includeeuneven excitation of nuclei within the patient due to RF pulses applied at flip angles other than 90- and 180-degree, abnormal loading of coil orcoupling of coil and inhomogeneity of magnetic field. -Axis: Frequency and phase encoding. -Corrective measures: 1. Load the coil correctly. 2. Shimming to reduce the inhomogeneity of the magnetic field. Shading artifact Fig. 8: Shading artifact: T2-w axial image of the brain shows comparatively less signal in the frontal regions. This was because of the improper loading (connection) of the coil in the anterior part. Cross Excitation 8 -Cross Excitation: An RF excitation pulse is not exactly square. As a result, nuclei in slices adjacent to the one excited by RF pulse may also receive energy and be excited. This energy flips NMV of these nuclei into transverse plane. When they are excited by their own RF excitation pulse, they do not have enough longitudinal magnetization to be tilted. This results in reduced signal intensity in the adjacent slices. This phenomenon is called cross excitation (Fig.9). -Axis: Slice selection gradient. -Corrective measures: Increase interslice gap. Cross Excitation Fig 9: Cross Excitation: Diagram shows excitation of nuclei in the adjacent slices because of parabolic shape of the slice excitation Cross Excitation

Use Quizgecko on...
Browser
Browser