X-ray Generator Fixed Installation Diagnosis
16 Questions
0 Views

Choose a study mode

Play Quiz
Study Flashcards
Spaced Repetition
Chat to lesson

Podcast

Play an AI-generated podcast conversation about this lesson

Questions and Answers

What is the aim of the provided information and procedures?

  • To provide information on how to use an X-ray generator
  • To provide information about the history of X-ray generators
  • To provide general information about medical equipment in hospitals
  • To provide information and procedures for diagnosing or repairing a problem with an X-ray generator (correct)
  • What does Section 1 of the contents focus on?

  • Preparing for radiographic exposure (correct)
  • Repairing high-tension problems
  • Testing X-ray tubes
  • Diagnosing a problem with the X-ray generator
  • What does the control system do before permitting preparation to begin?

  • It starts the exposure process immediately
  • It checks the X-ray tube
  • It performs high-tension tests
  • It makes a number of safety tests (correct)
  • What happens if all tests are satisfied during the preparation mode?

    <p>The system signals ready for an exposure</p> Signup and view all the answers

    Which section focuses on diagnosing or repairing a problem with an X-ray generator?

    <p>Part 1: Nothing happens</p> Signup and view all the answers

    What are flow charts provided for in the content?

    <p>To indicate logical steps when diagnosing a problem with an X-ray generator</p> Signup and view all the answers

    What could be the reason for no sound of X-ray tube anode rotation?

    <p>Fault in the generator</p> Signup and view all the answers

    When would a warning light indicate a fault in the generator?

    <p>Incorrect setting of the X-ray control</p> Signup and view all the answers

    What would cause a warning light indicating wrong exposure factors?

    <p>kV set above the maximum X-ray tube rating</p> Signup and view all the answers

    What happens if the combination of a high value of mA and a low value of kV is selected?

    <p>X-ray tube filament overheats</p> Signup and view all the answers

    Why might selecting a very short exposure time combined with a low mA prevent operation?

    <p>Exceeds X-ray tube anode heat capacity</p> Signup and view all the answers

    What should be done if the X-ray control microprocessor has calculated the amount of heat in the X-ray tube anode to be exceeded?

    <p>Wait a few minutes for the anode to cool down</p> Signup and view all the answers

    What do modern microprocessor-controlled generators have in special display modes?

    <p>'Reduce mAs' message on the control panel</p> Signup and view all the answers

    What would cause a warning light to indicate operator error?

    <p>Incorrect setting of X-ray control</p> Signup and view all the answers

    When might a small symbol indicate excessive anode heat on the control panel?

    <p>'Anode heat exceeded' symbol displayed</p> Signup and view all the answers

    What type of message might be displayed with microprocessor-controlled generators?

    <p>Direct message 'Reduce mAs'</p> Signup and view all the answers