Questions and Answers

What is the operating temperature of gravity-type steam instrument sterilizers?

  • 134°C (correct)
  • 55°C
  • 45°C
  • 37°C
  • Which sterilizing method is used for heat-sensitive items such as flexible endoscopes?

  • Gas Plasma
  • Dry Heat
  • Ethylene Oxide (EO) (correct)
  • Gravity-type Steam
  • How long do gravity-type steam sterilizers operate typically?

  • 15 minutes
  • 10 minutes
  • 30 minutes
  • 3–10 minutes (correct)
  • Which of the following is not a step in the Ethylene Oxide process?

    <p>Heat treatment</p> Signup and view all the answers

    What is a significant drawback of using Ethylene Oxide sterilizers?

    <p>High cost of health and safety requirements</p> Signup and view all the answers

    What is the maximum temperature used in Gas Plasma sterilizers?

    <p>55°C</p> Signup and view all the answers

    Which type of sterilizer requires no aeration after the cycle is complete?

    <p>Gas Plasma</p> Signup and view all the answers

    For which type of instruments is the use of gravity-type steam sterilizers primarily intended?

    <p>Invasive and non-heat sensitive instruments</p> Signup and view all the answers

    What is a characteristic of gravity displacement steam sterilizers?

    <p>They are dependent on gravity for operation.</p> Signup and view all the answers

    Which type of sterilizer incorporates a vacuum-assisted air removal stage?

    <p>Porous load steam sterilizers</p> Signup and view all the answers

    What is the typical sequence of operations in a porous load steam sterilization cycle?

    <p>Conditioning, exposure, exhaust, drying</p> Signup and view all the answers

    What type of items are not suitable for gravity displacement steam sterilizers?

    <p>Wrapped items</p> Signup and view all the answers

    What does immediate use steam sterilization (IUSS) refer to?

    <p>Fast sterilization of surgical instruments at the point of use.</p> Signup and view all the answers

    Which statement about flash sterilization is correct?

    <p>It is for non-porous and/or non-hollow surgical instruments.</p> Signup and view all the answers

    What determines the drying time at the end of a sterilization cycle?

    <p>Packaging, device, and sterilizer’s IFU.</p> Signup and view all the answers

    What is a drawback of using gravity displacement steam sterilizers?

    <p>They rely solely on gravity for air removal.</p> Signup and view all the answers

    What is the primary sterilizing agent used in vaporized hydrogen peroxide sterilization?

    <p>Hydrogen Peroxide</p> Signup and view all the answers

    Which of the following is a disadvantage of using dry heat sterilization?

    <p>Long sterilization and cooling times</p> Signup and view all the answers

    What is NOT a step in the vaporized hydrogen peroxide sterilization cycle?

    <p>Injection of Ethylene Oxide</p> Signup and view all the answers

    What type of packaging material is not suitable for use in hydrogen peroxide sterilizers?

    <p>Cellulose-containing materials</p> Signup and view all the answers

    Which sterilization method requires the longest processing time?

    <p>Ozone sterilization</p> Signup and view all the answers

    What is a primary advantage of steam sterilization?

    <p>Highly effective against a broad range of microorganisms</p> Signup and view all the answers

    Which of the following statements about ozone sterilization is TRUE?

    <p>It eliminates the need for purchasing a sterilant.</p> Signup and view all the answers

    What is the sterilization temperature range for ozone sterilization?

    <p>30.8-36 °C</p> Signup and view all the answers

    What is the suitable sterilization temperature for IUSS or flash sterilization?

    <p>134°C</p> Signup and view all the answers

    Which of the following types of items are classified as unsuitable for moist heat sterilizers?

    <p>Heat resistant tools</p> Signup and view all the answers

    What is a characteristic of ethylene oxide sterilization?

    <p>Dry process</p> Signup and view all the answers

    What is the recommended cycle time for gas plasma sterilization?

    <p>45-75 minutes</p> Signup and view all the answers

    What is a disadvantage of ethylene oxide sterilization?

    <p>Toxic and explosive gas</p> Signup and view all the answers

    What is the minimum time for porous load sterilization at 134°C?

    <p>3 minutes</p> Signup and view all the answers

    Which type of sterilizer is deemed unsuitable for long narrow lumened instruments?

    <p>Gas plasma sterilizer</p> Signup and view all the answers

    What is a common requirement for sterilization of heat sensitive reusable medical devices using ethylene oxide?

    <p>Non-cellulose wrapping materials</p> Signup and view all the answers

    Study Notes

    Steam Sterilization

    • Steam sterilization is commonly used in healthcare settings for medical equipment disinfection.
    • Various types of steam sterilizers exist, each suited for specific instruments and applications.

    Gravity Displacement Steam Sterilizers

    • These sterilizers depend on gravity for air removal, leading to slower steam penetration.
    • Not ideal for wrapped items or instruments with channels, as air trapping hinders effective sterilization.
    • Suitable for solid metal items like non-complex surgical instruments.

    Porous Load (Vacuum Assisted) Steam Sterilizers

    • Employ vacuum-assisted air removal prior to steam admission, ensuring efficient steam penetration.
    • Cycle involves conditioning (steam entry and air removal), exposure (sterilization at predetermined parameters), exhaust (steam removal and filtered air introduction), and drying.
    • Effective for heat-resistant tools, packaged items, and instruments with lumened channels.

    Immediate Use Steam Sterilization (IUSS) or Flash Sterilization

    • Primarily used for non-porous, non-hollow, unwrapped instruments, typically in operating rooms.
    • Instruments are sterilized quickly (134°C for 3–10 minutes), but emerge wet and very hot.

    Ethylene Oxide (EO) Sterilization

    • Low-temperature sterilization (37°C or 55°C), suitable for heat-sensitive items.
    • Commonly used commercially and in some hospitals, but restricted in some countries due to safety regulations.
    • EO is effective for invasive flexible endoscopes, cardiac, and ophthalmic devices, but cycle times are lengthy due to aeration requirements.

    Gas Plasma Sterilization

    • Another suitable method for heat-sensitive items, including endoscopes and laparoscopic instruments.
    • Operates at lower temperatures (45°C to 55°C) than EO, with shorter cycle times (45–75 minutes).
    • No post-sterilization aeration is required, as byproducts are nontoxic.

    Vaporized Hydrogen Peroxide Sterilization

    • Utilizes hydrogen peroxide to kill microorganisms through oxidation.
    • Cycle times are shorter (28-55 minutes) at lower temperatures (less than 50°C).
    • Effective for both lumened and non-lumened instruments, depending on the selected program.

    Ozone Sterilization

    • A low-temperature sterilization system that eliminates the need for sterilant purchase by generating ozone (O3).
    • Cycle takes 4 hours and 30 minutes at 30.8-36 °C.
    • Requires manufacturer approval for the medical device to be sterilized.

    Dry Heat Sterilization

    • Used for glassware, metal items, heat-stable powders, and non-aqueous liquids.
    • Operates at normal pressure with cycle times up to 2 hours.
    • Suitable for non-aqueous fluids and glassware, but not for heat-sensitive items.

    Steam Sterilization Advantages and Disadvantages

    Gravity Displacement Steam Sterilizers

    • Advantages: Inexpensive, highly effective, rapid process
    • Disadvantages: Unsuitable for heat-sensitive items, longer cycle times, items may be packaged

    Porous Load Steam Sterilizers

    • Advantages: Highly effective, rapid process, no toxic residues
    • Disadvantages: Moisture process (moist heat), unsuitable for heat-sensitive items, items may be packaged

    IUSS or Flash Sterilization

    • Advantages: Rapid process
    • Disadvantages: Moisture process (moist heat), unsuitable for heat-sensitive items, items cannot be wrapped

    Gas Plasma Sterilization

    • Advantages: Suitable for heat-sensitive devices, short cycle time, non-cellulose wrapping materials
    • Disadvantages: May not be suitable for long, narrow lumens

    Ethylene Oxide (EO) Sterilization

    • Advantages: Suitable for heat-sensitive reusable medical devices, extremely penetrative suitable for lumened devices, items can be wrapped, dry process, non-damaging
    • Disadvantages: Equipment is expensive, lengthy cycle with aeration and validation, ethylene oxide gas is toxic, explosive and flammable, separate accommodation is recommended, lengthy aeration required to remove toxic residues

    Vaporized Hydrogen Peroxide Sterilization

    • Advantages: Non-toxic
    • Disadvantages: Cellulose-containing packaging material cannot be used

    Studying That Suits You

    Use AI to generate personalized quizzes and flashcards to suit your learning preferences.

    Quiz Team

    Related Documents

    Steam Sterilizers PDF

    Description

    This quiz explores various steam sterilization methods used in healthcare settings for effective disinfection of medical equipment. It covers gravity displacement and vacuum-assisted steam sterilizers, detailing their mechanisms and best use cases. Test your knowledge on the principles and applications of steam sterilization.

    More Like This

    Use Quizgecko on...
    Browser
    Browser