What is the purpose of consulting the X-ray imaging manufacturer regarding the system’s default configuration for calibration?
What is the purpose of consulting the X-ray imaging manufacturer regarding the system’s default configuration for calibration?
- To streamline the testing process and ensure correct data scaling. (correct)
- To determine the pulse rate for high-output mode.
- To select the FOV for calibration.
- To understand how to program every procedure into the system.
How does selecting high-output mode impact the pulse rate in the IXE system?
How does selecting high-output mode impact the pulse rate in the IXE system?
- It decreases the pulse rate compared to low or normal mode.
- It maintains a constant pulse rate across all modes.
- It eliminates the need for pulse rate adjustment.
- It increases the pulse rate compared to low or normal mode. (correct)
What is recommended in selecting procedures for acceptance testing of an IXE system?
What is recommended in selecting procedures for acceptance testing of an IXE system?
- Avoid selecting any procedures as it can be time-consuming.
- Select procedures based on the system's primary function. (correct)
- Choose every available procedure for thorough evaluation.
- Opt for procedures related to neurological angiography only.
Which factor influences the default FOV for calibration in an IXE system?
Which factor influences the default FOV for calibration in an IXE system?
Why is it essential to understand the numerous control parameters that affect radiation output?
Why is it essential to understand the numerous control parameters that affect radiation output?
What is depicted on the right side of Figure 8 in the described geometry?
What is depicted on the right side of Figure 8 in the described geometry?
What conditions should be considered when measuring actual patient entrance air kerma (AKR)?
What conditions should be considered when measuring actual patient entrance air kerma (AKR)?
When conducting acceptance testing for free-in-air AKR measurements, what attenuation phantom is used?
When conducting acceptance testing for free-in-air AKR measurements, what attenuation phantom is used?
Why might a radiotransparent ionization chamber be employed in place of the SSDS for fluoroscopic automatic exposure control?
Why might a radiotransparent ionization chamber be employed in place of the SSDS for fluoroscopic automatic exposure control?
What should be done to minimize the influence of the lead backing of SSDS during fluoroscopic automatic control?
What should be done to minimize the influence of the lead backing of SSDS during fluoroscopic automatic control?
Why is testing dose rates for recorded/acquisition imaging modes problematic?
Why is testing dose rates for recorded/acquisition imaging modes problematic?
What precaution should be taken during measurements to avoid excess X-ray tube heat loading and possible sensor 'burn-in'?
What precaution should be taken during measurements to avoid excess X-ray tube heat loading and possible sensor 'burn-in'?