Radiation Measurement Acceptance Testing Report
12 Questions
0 Views

Choose a study mode

Play Quiz
Study Flashcards
Spaced Repetition
Chat to Lesson

Podcast

Play an AI-generated podcast conversation about this lesson

Questions and Answers

What is the purpose of consulting the X-ray imaging manufacturer regarding the system’s default configuration for calibration?

  • To streamline the testing process and ensure correct data scaling. (correct)
  • To determine the pulse rate for high-output mode.
  • To select the FOV for calibration.
  • To understand how to program every procedure into the system.

How does selecting high-output mode impact the pulse rate in the IXE system?

  • It decreases the pulse rate compared to low or normal mode.
  • It maintains a constant pulse rate across all modes.
  • It eliminates the need for pulse rate adjustment.
  • It increases the pulse rate compared to low or normal mode. (correct)

What is recommended in selecting procedures for acceptance testing of an IXE system?

  • Avoid selecting any procedures as it can be time-consuming.
  • Select procedures based on the system's primary function. (correct)
  • Choose every available procedure for thorough evaluation.
  • Opt for procedures related to neurological angiography only.

Which factor influences the default FOV for calibration in an IXE system?

<p>The size of the installed FPIR (B)</p> Signup and view all the answers

Why is it essential to understand the numerous control parameters that affect radiation output?

<p>To streamline the testing process and achieve accurate data scaling. (B)</p> Signup and view all the answers

What is depicted on the right side of Figure 8 in the described geometry?

<p>A homemade supporting table with an opening (C)</p> Signup and view all the answers

What conditions should be considered when measuring actual patient entrance air kerma (AKR)?

<p>Automatic dose control logic enabled, normal positioning of the examination table, use of a table pad, presence of back scatter from any phantom used (B)</p> Signup and view all the answers

When conducting acceptance testing for free-in-air AKR measurements, what attenuation phantom is used?

<p>PMMA slabs (C)</p> Signup and view all the answers

Why might a radiotransparent ionization chamber be employed in place of the SSDS for fluoroscopic automatic exposure control?

<p>To maintain fidelity with clinical patient radiation exposure (A)</p> Signup and view all the answers

What should be done to minimize the influence of the lead backing of SSDS during fluoroscopic automatic control?

<p>Place the radiation detector/probe outside the sensing area(s) for the fluoroscopic automatic control (B)</p> Signup and view all the answers

Why is testing dose rates for recorded/acquisition imaging modes problematic?

<p>Exceedingly high dose rates and absence of regulatory limits on these rates (C)</p> Signup and view all the answers

What precaution should be taken during measurements to avoid excess X-ray tube heat loading and possible sensor 'burn-in'?

<p>Carefully monitor X-ray tube heat loading and sensor 'burn-in' (A)</p> Signup and view all the answers