During the initial setup of the Puritan Bennett 840 ventilator for an SST, if the error message 'short self test not allowed after patient connection' appears, what is the most likely cause?
During the initial setup of the Puritan Bennett 840 ventilator for an SST, if the error message 'short self test not allowed after patient connection' appears, what is the most likely cause?
- The ventilator's internal battery is low and needs to be charged.
- The Y connector is attached to the ventilator circuit. (correct)
- The ventilator is not properly connected to an oxygen source.
- The humidifier is already connected to the circuit.
After initiating the short self test (SST) on the Puritan Bennett 840 ventilator, what is the correct sequence of actions when prompted to 'Block the Y'?
After initiating the short self test (SST) on the Puritan Bennett 840 ventilator, what is the correct sequence of actions when prompted to 'Block the Y'?
- Disconnect the Y connector from the circuit, then press accept.
- Attach the Y connector back to the circuit, then press accept. (correct)
- Connect the humidifier, then press accept.
- Leave the Y connector as it is, the ventilator will automatically continue the test.
Which of the following components is tested during the early stages of the Puritan Bennett 840 ventilator's short self test (SST)?
Which of the following components is tested during the early stages of the Puritan Bennett 840 ventilator's short self test (SST)?
- The flow sensor functionality (correct)
- The internal battery's charge capacity
- The calibration of the touch screen display
- The external oxygen tank pressure
During the Puritan Bennett 840 ventilator's SST, you are prompted to disconnect a tube. Which tube should be disconnected to proceed with the test?
During the Puritan Bennett 840 ventilator's SST, you are prompted to disconnect a tube. Which tube should be disconnected to proceed with the test?
If an incorrect tube is disconnected during the Puritan Bennett 840 ventilator's short self test (SST) when prompted, what is the likely outcome?
If an incorrect tube is disconnected during the Puritan Bennett 840 ventilator's short self test (SST) when prompted, what is the likely outcome?
What actions are performed involving the Y connector during the final stages of the circuit resistance testing of the Puritan Bennett 840 ventilator's SST?
What actions are performed involving the Y connector during the final stages of the circuit resistance testing of the Puritan Bennett 840 ventilator's SST?
After the Puritan Bennett 840 ventilator displays the 'SST testing is completed' message, what is the next appropriate action?
After the Puritan Bennett 840 ventilator displays the 'SST testing is completed' message, what is the next appropriate action?
What is the estimated total duration of the Short Self Test (SST) on the Puritan Bennett 840 ventilator?
What is the estimated total duration of the Short Self Test (SST) on the Puritan Bennett 840 ventilator?
What parameters or conditions are assessed by the Puritan Bennett 840 ventilator during the circuit leak test stage of the SST?
What parameters or conditions are assessed by the Puritan Bennett 840 ventilator during the circuit leak test stage of the SST?
What initial default patient setting does the Puritan Bennett 840 ventilator assume when starting the Short Self Test (SST)?
What initial default patient setting does the Puritan Bennett 840 ventilator assume when starting the Short Self Test (SST)?
Flashcards
Short Self Test (SST)
Short Self Test (SST)
A brief diagnostic procedure to ensure the ventilator is functioning correctly before patient use.
"Short self test not allowed" error
"Short self test not allowed" error
Remove the Y-connector from the ventilator tubing to clear this initial error.
Test Button
Test Button
The button located on the side of the ventilator used to initiate the Short Self Test sequence.
Circuit Connection
Circuit Connection
Signup and view all the flashcards
Blocking the Y-Connector
Blocking the Y-Connector
Signup and view all the flashcards
Humidifier Connection
Humidifier Connection
Signup and view all the flashcards
Tube Disconnection/Reconnection
Tube Disconnection/Reconnection
Signup and view all the flashcards
SST Completion Message
SST Completion Message
Signup and view all the flashcards
Study Notes
- This tutorial details performing a Short Self Test (SST) on the Puritan Bennett 840 ventilator.
- Before starting, ensure the ventilator is plugged into the wall and connected to oxygen.
- The ventilator should have the tubing with the Y connector attached.
Initial Setup and Error Handling
- The ventilator may initially display a "short self test not allowed after patient connection" error.
- This alarm can be silenced by pressing the silence button.
- The error typically occurs when the Y connector is attached.
- Remove the Y connector prior to beginning to resolve the error, then select the short self test option.
Starting the Short Self Test
- Initiate the SST by pressing the test button on the side of the ventilator.
- The default patient setting is adult; press accept to continue and start the test sequence.
Step-by-Step Testing Procedure
- The ventilator provides instructions in the lower right corner of the screen.
- Prompts will guide you to connect the circuit, block the Y, and connect the humidifier.
- Connect the circuit and press accept to proceed.
- Block the Y connector by reattaching it, and then press accept to start the test.
- The total test time is approximately 5 minutes.
Testing Stages
- The ventilator conducts tests on the flow sensor, pressure, and circuit leak.
- The flow sensor test is performed first.
- The user is then prompted to connect the humidifier and press accept.
- After verifying the humidifier, the ventilator tests for circuit leaks, monitoring for any leaks.
Disconnecting and Reconnecting Tubes
- During the test, the ventilator will prompt disconnection from the patient port.
- Disconnect the indicated blue tube located by the expiratory filter.
- Incorrectly disconnecting a tube will cause the test to fail, and it will need to be repeated.
- After disconnecting the correct tube, press accept, then reconnect it and press accept again.
Final Steps and Completion
- The ventilator tests circuit resistance at this stage.
- Instructions will include unblocking the Y, then blocking it again, and finally unblocking it.
- The ventilator displays "SST testing is completed" when finished and indicates if all tests passed.
- Press accept to exit the SST.
- The ventilator is safe to connect to a patient after successfully completing the SST.
Studying That Suits You
Use AI to generate personalized quizzes and flashcards to suit your learning preferences.
Description
Learn how to perform a Short Self Test (SST) on the Puritan Bennett 840 ventilator. This tutorial covers initial setup, error handling, and step-by-step instructions for running the test. Ensure proper connections and settings before initiating the SST.