Podcast
Questions and Answers
Which pacemaker mode is MOST suitable for a patient undergoing electrocautery?
Which pacemaker mode is MOST suitable for a patient undergoing electrocautery?
- Demand mode
- Biventricular mode
- Synchronous mode
- Asynchronous/fixed mode (correct)
What does the 'I' in a pacemaker code (e.g., VVI) typically represent?
What does the 'I' in a pacemaker code (e.g., VVI) typically represent?
- Inhibition (correct)
- Intrinsic rate
- Impedance
- Initiation
A patient with a pacemaker experiences a 'failure to capture'. What does this term indicate?
A patient with a pacemaker experiences a 'failure to capture'. What does this term indicate?
- The pacemaker is delivering a stimulus, but it does not result in myocardial depolarization. (correct)
- The pacemaker is firing at an abnormally high rate.
- The pacemaker is not sensing the heart's intrinsic activity.
- The pacemaker is being inhibited by extraneous signals.
What is the primary reason for using bipolar electrocautery in a patient with a pacemaker? select 2
What is the primary reason for using bipolar electrocautery in a patient with a pacemaker? select 2
In the context of pacemakers, what does 'sensitivity' refer to?
In the context of pacemakers, what does 'sensitivity' refer to?
Why might lidocaine be avoided during the initial placement of an ICD?
Why might lidocaine be avoided during the initial placement of an ICD?
What is the MOST likely outcome if a pacemaker is programmed in DOO or VOO mode and the patient's intrinsic heart rate is higher than the programmed pacing rate?
What is the MOST likely outcome if a pacemaker is programmed in DOO or VOO mode and the patient's intrinsic heart rate is higher than the programmed pacing rate?
Which of the following is a common indication for the use of a temporary pacemaker? select 3
Which of the following is a common indication for the use of a temporary pacemaker? select 3
A patient with a pacemaker is experiencing 'undersensing'. What is the MOST likely consequence of this malfunction?
A patient with a pacemaker is experiencing 'undersensing'. What is the MOST likely consequence of this malfunction?
What does triggered response mean in the context of pacemaker sensing?
What does triggered response mean in the context of pacemaker sensing?
Which of the following best describes the purpose of effective refractory periods in the context of pacemaker function?
Which of the following best describes the purpose of effective refractory periods in the context of pacemaker function?
What does the VA interval represent in the context of dual-chamber pacing?
What does the VA interval represent in the context of dual-chamber pacing?
What is the purpose of tiered therapy in implantable cardioverter-defibrillators (ICDs)?
What is the purpose of tiered therapy in implantable cardioverter-defibrillators (ICDs)?
During the implantation of an insertable cardiac loop recorder (ICLR), what action triggers the device to record a cardiac event?
During the implantation of an insertable cardiac loop recorder (ICLR), what action triggers the device to record a cardiac event?
Why are flecainide, procainamide, and quinidine contraindicated for perioperative use in patients with pacemakers?
Why are flecainide, procainamide, and quinidine contraindicated for perioperative use in patients with pacemakers?
What drug class do flecainide, procainamide, and quinidine belong to?
What drug class do flecainide, procainamide, and quinidine belong to?
What are the characteristics of a pacemaker in asynchronous mode? (Select 3)
What are the characteristics of a pacemaker in asynchronous mode? (Select 3)
What are characteristics of a pacemaker in synchronous/demand mode? (Select 3)
What are characteristics of a pacemaker in synchronous/demand mode? (Select 3)
What is the threshold for pacing the heart?
What is the threshold for pacing the heart?
What is the AV interval in cardiac pacing?
What is the AV interval in cardiac pacing?
What are indications for permanent pacing? (Select all that apply)
What are indications for permanent pacing? (Select all that apply)
How does a permanent pacemaker aid in the management of atrial fibrillation?
How does a permanent pacemaker aid in the management of atrial fibrillation?
What does "inhibited" mean regarding pacemakers? select 2
What does "inhibited" mean regarding pacemakers? select 2
What is the purpose of inhibition of pacing?
What is the purpose of inhibition of pacing?
What does 'D' mean regarding sensing in a pacemaker? select 2
What does 'D' mean regarding sensing in a pacemaker? select 2
What does VVI mean in terms of pacemaker function?
What does VVI mean in terms of pacemaker function?
When would 'O' sensing be appropriate for a pacemaker?
When would 'O' sensing be appropriate for a pacemaker?
When would VOO pacing be appropriate?
When would VOO pacing be appropriate?
Match the pacing mode with the correct description:
Match the pacing mode with the correct description:
What are key features of a single chamber pacemaker? select 2
What are key features of a single chamber pacemaker? select 2
What modes are associated with single chamber pacemakers?
What modes are associated with single chamber pacemakers?
Where are the leads located in dual chamber pacemakers?
Where are the leads located in dual chamber pacemakers?
What are leadless permanent pacemakers indicated for?
What are leadless permanent pacemakers indicated for?
Where are leadless permanent pacemakers typically placed?
Where are leadless permanent pacemakers typically placed?
How do leadless permanent pacemakers work?
How do leadless permanent pacemakers work?
Where are the leads located in transvenous pacing?
Where are the leads located in transvenous pacing?
What are features of epicardial pacing? (Select 2)
What are features of epicardial pacing? (Select 2)
Which of the following are invasive temporary pacemakers? (Select all that apply)
Which of the following are invasive temporary pacemakers? (Select all that apply)
Which of the following is a non-invasive temporary pacemaker?
Which of the following is a non-invasive temporary pacemaker?
What is the polarity of the leads of a pacemaker?
What is the polarity of the leads of a pacemaker?
Why do we check a pacemaker in demand/synchronous mode only?
Why do we check a pacemaker in demand/synchronous mode only?
What is the R-on-T phenomenon?
What is the R-on-T phenomenon?
What are the causes of the R-on-T phenomenon? (Select 2)
What are the causes of the R-on-T phenomenon? (Select 2)
Which pacing mode is notable for sensing the heart's intrinsic rhythm and avoiding R on T? select 2
Which pacing mode is notable for sensing the heart's intrinsic rhythm and avoiding R on T? select 2
What is the result of the intrinsic heart rate being lower than the paced rate in DOO?
What is the result of the intrinsic heart rate being lower than the paced rate in DOO?
What is the consequence of the intrinsic heart rate being faster than the paced rate in DOO mode?
What is the consequence of the intrinsic heart rate being faster than the paced rate in DOO mode?
What are indications for magnet placement intraoperatively?
What are indications for magnet placement intraoperatively?
What is the result of undersensing/failure to sense in a pacemaker?
What is the result of undersensing/failure to sense in a pacemaker?
What is the result of oversensing in a pacemaker?
What is the result of oversensing in a pacemaker?
Which of the following drugs are associated with pacemaker malfunction? (Select 3)
Which of the following drugs are associated with pacemaker malfunction? (Select 3)
How do sodium channel blockers cause pacemaker malfunction? select 2
How do sodium channel blockers cause pacemaker malfunction? select 2
What does failure to pace mean?
What does failure to pace mean?
What is the result of a magnet being placed on a patient with a pacemaker?
What is the result of a magnet being placed on a patient with a pacemaker?
What should be part of your setup for pacemaker insertion? (Select 3)
What should be part of your setup for pacemaker insertion? (Select 3)
What are the electrocautery considerations for permanent pacemaker placement? (Select 3)
What are the electrocautery considerations for permanent pacemaker placement? (Select 3)
What effect does lithotripsy have on pacemakers?
What effect does lithotripsy have on pacemakers?
What effect does TURP or a hysteroscopy have on pacemaker function?
What effect does TURP or a hysteroscopy have on pacemaker function?
What mode should the pacemaker be placed in during ECT, monitoring SSEP, or MEP?
What mode should the pacemaker be placed in during ECT, monitoring SSEP, or MEP?
What are the indications for using an ICLR (Insertable Cardiac Loop Recorder)?
What are the indications for using an ICLR (Insertable Cardiac Loop Recorder)?
What can occur when using electrocautery on a patient with an ICLR?
What can occur when using electrocautery on a patient with an ICLR?
What complication can occur from lithotripsy in a patient with an ICLR?
What complication can occur from lithotripsy in a patient with an ICLR?
What is a consideration for a patient with an Implantable Cardioverter-Defibrillator (ICD) undergoing MRI?
What is a consideration for a patient with an Implantable Cardioverter-Defibrillator (ICD) undergoing MRI?
What are the functions of an ICD? (Select all that apply)
What are the functions of an ICD? (Select all that apply)
What should occur prior to surgery with a patient who has an ICD? (Select 2)
What should occur prior to surgery with a patient who has an ICD? (Select 2)
Which of the following intraoperative conditions can cause pacemakers to malfunction? (Select all that apply)
Which of the following intraoperative conditions can cause pacemakers to malfunction? (Select all that apply)
What are complications of permanent pacemaker placement? (Select all that apply)
What are complications of permanent pacemaker placement? (Select all that apply)
What surgical technique can interfere with pacemaker function?
What surgical technique can interfere with pacemaker function?
Why should procainamide be avoided in a patient with an ICD?
Why should procainamide be avoided in a patient with an ICD?
What anesthetic technique is typically used to place an ICD?
What anesthetic technique is typically used to place an ICD?
Flashcards
Synchronous Mode
Synchronous Mode
Delivers voltage to generate depolarization when the heart's intrinsic rate is less than the set rate.
Threshold (Pacemaker)
Threshold (Pacemaker)
The MINIMUM current needed to stimulate the heart.
Effective Refractory Period
Effective Refractory Period
Time when a cardiac cell cannot respond to a new stimulus, protecting against arrhythmias.
Atrial Fibrillation Pacing
Atrial Fibrillation Pacing
Signup and view all the flashcards
Triggered (T)
Triggered (T)
Signup and view all the flashcards
Inhibited (I)
Inhibited (I)
Signup and view all the flashcards
DDD Pacing
DDD Pacing
Signup and view all the flashcards
VVI Pacing
VVI Pacing
Signup and view all the flashcards
RV Lead Placement
RV Lead Placement
Signup and view all the flashcards
R on T Phenomenon
R on T Phenomenon
Signup and view all the flashcards
Pacemaker Undersensing
Pacemaker Undersensing
Signup and view all the flashcards
Pacemaker Oversensing
Pacemaker Oversensing
Signup and view all the flashcards
Drugs Affecting Pacemakers
Drugs Affecting Pacemakers
Signup and view all the flashcards
ICD - Medications to Avoid
ICD - Medications to Avoid
Signup and view all the flashcards
Biventricular Pacemaker
Biventricular Pacemaker
Signup and view all the flashcards
Electrocautery
Electrocautery
Signup and view all the flashcards
Electrosurgery
Electrosurgery
Signup and view all the flashcards
Lead in RV
Lead in RV
Signup and view all the flashcards
Study Notes
Pacemaker Types
- Pacemakers can be single chamber, dual chamber, or biventricular
- They can be temporary (external) or permanent (internal)
Pacemaker Modes
- Pacemakers operate in two modes: asynchronous or synchronous.
Asynchronous Mode
- Ignores the heart's intrinsic activity
- Does not sense intrinsic activity
- Paces at a fixed rate set
Synchronous Mode
- Senses the intrinsic heart rate and delivers voltage to cause depolarization
- Paces only when the intrinsic heart rate is less than the set rate
- WILL NOT fire if the heart rate is above the set rate
Pacemaker Function
- Paces the heart, measured in milliamperes (mA)
- Senses the heart, measured in millivolts (mV)
Wave Detection and Sensitivity
- R wave detection relies on a voltage level above the pacemaker's set level
- Sensitivity relates to the voltage level needing to be exceeded to detect an R or P wave
Threshold and Intervals
- Threshold denotes the minimum current needed to pace the heart
- AV interval is the duration between paced or sensed atrial depolarization and the subsequent ventricular stimulus
- VA interval is the duration between sensed or paced ventricular depolarization and the next atrial stimulus
Effective Refractory Periods
- The amount of time in which a cell is unable to respond to a new stimulus
- Protects against arrhythmias
Temporary Pacemaker Indications
- Used for unstable bradyarrhythmias or tachyarrhythmias
- Employed in third-degree heart block
- Provides support post open-heart surgery
- Use ends upon resolution of the problem or permanent pacemaker implantation
Permanent Pacing Indications
- Sinus node dysfunction
- Symptomatic impulse formation diseases
- Sinus bradycardia, sinus pause, SA block
- AV node disease
- Atrial fibrillation, using dual-site pacing as a means to decrease intra-atrial conduction time
Pacemaker Code
- Indicates the chamber paced, chamber sensed, and response to sensing
Sensing Response
- Triggered (T) means that a sensed intrinsic depolarization results in pacemaker firing
- Heart will respond to a sensed intrinsic activity by delivering a pacing stimulus
- Inhibited (I) means the device DOES NOT PACE
- Pacemaker will withhold a pacing stimulus on sensing the heart’s electrical activity
- Prevents competition with the heart’s intrinsic rhythm and reduces the risk of inducing arrhythmias
- The Pacemaker will fire if the heart fails to beat on its own
- Dual (D) defines the dual inhibition of atrial and ventricular pacing in response to intrinsic ventricular depolarization
- The pacemaker inhibits both atrial and ventricular pacing if the heart is functioning normally
- Pacemaker will fire in the needed chamber (atria or ventricle) if intrinsic heart activity is missing or delayed
- None (O) Utilized for asynchronous pacing, regardless of intrinsic activity
- Useful when electrocautery, which interferes with pacemakers, is being used
- Placing a magnet on the patient’s chest puts the pacemaker in asynchronous pacing mode, such as D00 at 99 HR
- Example DDD signifies both atrium and ventricle paced, both atrium and ventricle depolarizations sensed, inhibition in response to depolarization, and can trigger a response to depolarization
- Example: VVI indicates ventricle paced, ventricle sensed, and that a ventricle event outside of the refractory period will be inhibited
Chamber Pacemakers
- Lead placement in the right ventricle is most common
- Paces or senses in one chamber only, for example, AAI, VVI
Dual Chamber Pacemakers
- Leads placed in the right atrium (RA) and right ventricle (RV)
- DDD can pace and sense in both the atrium and ventricle
- Features 2 timing intervals, the AV interval and VA interval
Leadless Permanent Pacemakers
- Indicated for AV block
- Implanted under fluoroscopy into the RV septum
- Contains 4 tines instead of leads
- Utilizes internal atrial sensing algorithms to adjust pacing which coordinates with the atria
Temporary Pacemakers
- Transvenous pacing is invasive; wire is placed from the internal jugular (IJ) to the RV
- Epicardial pacing is invasive, with most surgeons placing 2 atrial and 2 ventricular wires, with some surgeons placing two ventricular wires only
- Transcutaneous pacing is non-invasive
- Esophageal pacing is invasive
Pacing Wires and Cable Connectors
- Polarity of wires, black to negative, red to positive.
- Leads connect from the wire insertion to the controller
Checking Pacemaker Function
- Confirming all connections are secure
- Settings must be confirmed before starting (mode, rate, output, sensitivity, AV interval)
- Check in Demand mode only, which prevents firing and causing R on T
R on T Phenomenon
- Ventricular extrasystole caused by ventricular depolarization superimposing on a previous beat’s repolarization
- Causes include undersensing of the pacemaker
- Pacemaker starting in asynchronous mode, allowing an erroneous beat to fall on the T wave
- Avoided if the pacemaker is in DDD or VVI mode, because can sense the heart’s intrinsic rhythm
- If the pacemaker is in DOO or VOO mode, and the intrinsic heart rate is lower than the paced rate, It will speed up the HR
- Alternatively, if the intrinsic heart rate is higher than the paced rate, the pacer will pace on a T wave, and VT or VF may occur
Pacemaker Components
- Leads
- Box
- Battery (with a life of 7-10 years)
- Electrodes
- Barbs/screws
Magnets with Pacemakers
- Usage can cause a device-specific rate change or a percentage decrease
- Useful with DDD-VVI
- A magnet would be used on a pacemaker-dependent patient during electrocautery
- Magnets are placed over the pacemaker any time EMI is expected
- Magnet pacing rates may differ from the set fixed pacemaker rate
Pacemaker Malfunctions
- Failure to pace
- Failure to capture
- Pacing at abnormal rates
- Failure to sense (undersensing) leads to inappropriate pacing
- Undersensing causes cardiac activity to go undetected
- The device mistakenly believes the heart is not beating, so an unnecessary unneeded impulse is fired
- Main clinical risk from this includes competitive pacing, leading to arrhythmias like R on T
- Oversensing causes inhibition of pacing
- The pacemaker detects extraneous signals as intrinsic activity
- The device is inhibited from pacing
- Clinical risk from this includes inappropriate pauses or bradycardia
- Oversensing and malfunction is a unique issue for dual-chamber devices
- Malfunction can be caused by drugs and certain conditions
- Drugs such as procainamide may cause failure to capture
- Quinidine use may increase the threshold.
- Flecainide (Na+ channel blocker) raises the threshold
- Electrical interference, hypokalemia, hyperkalemia, myocardial ischemia, and issues with pacemaker placement, leads, or the generator are potential conditions
Anesthetic Considerations
- Requires knowledge of external or existing pacemaker
- Determine if patient is pacemaker dependent
- Permanent pacemaker insertion may involve MAC (monitored anesthesia care) in an EP (electrophysiology) lab or cath lab
- NC (nasal cannula) is preferred for oxygenation to avoid potential fire hazards associated with masks
- Location of the pacemaker
- Verification of pacemaker function, typically by a device representative
- Magnet availability
Electrocautery and Electrosurgery
- Electrocautery passes through alternating current through a resistant metal wire electrode for coagulation only
- Electrosurgery involves direct heating of tissue for cutting and coagulation
- Monopolar electrosurgery directs current from the electrode, through the body, to a grounding pad, and back to the generator
- Bipolar electrosurgery restricts the current to a small tissue volume for better control
- Advantages of bipolar include reduced damage to sensitive tissues near the instrument
- Reduced chance of current arcs directly to the surrounding surface
Electrocautery Considerations for Permanent Pacemakers
- Bovie grounding pad placement should be away from the generator
- Return electron should be opposite the surgical site
- Bipolar, short bursts, and the lowest effective current should be used by the surgeon
Complications of Permanent Pacemaker Placement
- Lead dislodgement
- Pneumothorax
- Infection
- Bleeding
Pacemaker Adjustments for Surgical Procedures
- Lithotripsy is safe if the generator is protected, and the beam is not directed towards the device
- TURP/hysteroscopy: the pacemaker may need to be asynchronous because the surgeon will use monopolar cautery
- ECT (electroconvulsive therapy) can be performed with the pacemaker in asynchronous mode
- MRI considerations
- TENS/SSEP/MEP are generally safe
Insertable Cardiac Loop Recorders (ICLR)
- Implanted in the left pectoral region within subcutaneous fat
- Alternative to the Holter cardiac monitor
- No absolute contraindications for implantation
- Absence of wiring between the device and the heart
- Device activation is triggered by a patient using a magnet during a presyncopal or syncope episode
- Used for diagnosis and management of various cardiac arrhythmias and symptoms
ICLRs Anesthetic Considerations
- Need a cardiologist's progress notes
- Interrogate the device before surgery
- Assess recordings of recent life-threatening arrhythmias
- Made with a titanium shell patients may feel a pulling during MRI
- Electrocautery is safe but data loss can occur if the device is reset
- Focal point of the beam should not be set on the device as it can cause damage during lithotripsy
ICDs
- The size of a pacemaker
- Two types: transvenous or subcutaneous
- Transvenous leads sense, pace, and deliver biphasic shocks
- Can be single, dual, or biventricular (cardiac resynchronization therapy)
- Pulse generator and leads come standard on all types
ICD Functions
- Able to sense ventricular depolarization
- Provide VF detection
- Provide tachycardia detection and discrimination
- Tiered therapy options
- Biphasic shocks use lower energy requirements
ICDs Anesthetic Considerations
- GA vs MAC (most devices are placed under MAC)
- Avoid Lidocaine, as it obscures checking; also avoid procainamide (both affect defibrillation energy requirements)
- Maintenance: technique and caution with EF
Studying That Suits You
Use AI to generate personalized quizzes and flashcards to suit your learning preferences.