Podcast
Questions and Answers
What happens when the oxygen pressure drops below 20-25 psi?
What happens when the oxygen pressure drops below 20-25 psi?
What is a potential consequence of using the oxygen flush valve incorrectly?
What is a potential consequence of using the oxygen flush valve incorrectly?
How does the oxygen proportioning device ensure patient safety?
How does the oxygen proportioning device ensure patient safety?
What is the function of the low pressure system in gas delivery?
What is the function of the low pressure system in gas delivery?
Signup and view all the answers
What component is designed to prevent the inadvertent activation of the oxygen flush valve?
What component is designed to prevent the inadvertent activation of the oxygen flush valve?
Signup and view all the answers
What is the function of having a minimum oxygen flow rate of 50-150 ml/min even when the flowmeter is in the full-off position?
What is the function of having a minimum oxygen flow rate of 50-150 ml/min even when the flowmeter is in the full-off position?
Signup and view all the answers
Which component of the anesthesia machine helps to prevent accidental disconnection from the breathing circuit?
Which component of the anesthesia machine helps to prevent accidental disconnection from the breathing circuit?
Signup and view all the answers
During manual or assisted bag ventilation, what must be done to the APL valve to ensure adequate gas delivery?
During manual or assisted bag ventilation, what must be done to the APL valve to ensure adequate gas delivery?
Signup and view all the answers
What happens if the APL valve is closed too tightly during ventilation?
What happens if the APL valve is closed too tightly during ventilation?
Signup and view all the answers
What type of flowmeter is specifically designed to be gas-specific and uses a free-floating indicator?
What type of flowmeter is specifically designed to be gas-specific and uses a free-floating indicator?
Signup and view all the answers
Study Notes
Oxygen Failure Safety Valve
- Shuts off flow of all other gases when oxygen pressure drops below a set limit (20-25 psi)
- Activates if there is decreased flow of oxygen
- Some machines have audible and visual alarms, even with low oxygen pressure
- Does not prevent delivery of a hypoxic mixture
- Datex-Ohmeda and Drager have proprietary versions
Oxygen Pressure-Sensor Shutoff
- Datex-Ohmeda proprietary system
O2 Failure Protection Device
- Drager proprietary system
Oxygen Flush Valve
- Provides oxygen at 35-55 L/min directly to the common gas outlet, bypassing flowmeters and vaporizers
- Oxygen may reach a connected patient at a pressure of 45-50 psi
- Potential for barotrauma
- Used to rapidly refill or flush the breathing circuit
- Must have a built-in rim to prevent inadvertent activation and a spring to prevent prolonged activation
Low Pressure System
- Starts at the second stage regulator where all gas pressure is dropped from the intermediate pressure system to about 16 psi
- Includes flowmeters, vaporizers, and the common gas outlet
Flow Valves
- Touch and color-coded to prevent use of the wrong gas
- Oxygen knob is fluted, larger than the others, and protrudes further
- Oxygen is located furthest to the right
Oxygen Proportioning Device
- Ensures that nitrous oxide can only be delivered in the presence of oxygen
- Prevents delivery of low oxygen content nitrous oxide to the patient
- Adds a minimum of 21-25% oxygen to the mixture when nitrous oxide is used
- Oxygen is furthest on the right, so that in the event of a leak, the chances of a hypoxic mixture are lessened
- To ensure oxygen is always delivered, there's a minimum setting of 50-150ml/min constant flow even with the oxygen flowmeter in the full-off position
Flowmeters (aka Thorpe tubes)
- Constant pressure, variable orifices with a low pressure of 16 psi
- Tube is tapered, with the largest diameter on top
- Hand calibrated, no two tubes are alike
- Gas specific
- The indicator floats freely at a point where the downward force (gravity) equals the upward force caused by gas molecules hitting the bottom of the float
- Read the ball at the middle, and the arrow float at the top
Flowmeter Leaks
- Can occur, leading to hypoxic mixture being delivered
Power Failure
- Battery supply for 30 minutes
- Added backup systems
- Pneumatic versus electrical systems
Alternative Oxygen
- Separate axillary flowmeter
- Used to control oxygen in power failure
Common Gas Outlet
- Also known as the fresh gas outlet
- Only one outlet on the anesthesia machine, compared to multiple gas inlets
- Must have an anti-disconnect device to prevent accidental disconnection from the breathing circuit
- Oxygen from the flush valve goes directly from the intermediate pressure system to this outlet, directly to the patient
APL Valve (AKA Pop-off valve)
- Usually fully open during spontaneous ventilation, but must be partially closed during manual or assisted bag ventilation
- If open too wide, too much volume escapes and not enough is delivered to the patient
- If closed too tightly, not enough volume escapes, creating larger volumes and possibly high airway pressures
- Requires very fine tuning
Unidirectional Valves
- Maintain the flow of anesthetic gas in the intended direction
- Maintains a minimum O2/N20 ratio of 1:3
- O2/NO2 Proportioning Device automatically mixes O2 and NO2 to the setting selected on the dial
Proportioning Device
- Datex-Ohmeda Link-25 proprietary system
- Drager’s ORMC/S-ORC proprietary system
Flowmeters
- Constant-pressure variable-orifice flowmeters use an indicator ball or bobbin which displays the flow of gas through a Thorpe tube
- Utilize a ball, bobbin or float as a marker
- The tube is narrow at the bottom and low flow raises the marker; as the flows increase and the marker rises in the widened tube, greater flows pass around it and reach the common gas outlet
- May be single or double tube design
Flowmeters
- Specifically designed and calibrated for each gas
- The flow rate of a gas through a constriction depends on the gas’s viscosity at low laminar flows, and its density at high turbulent flow (Hagen-Poiseuille)
- Floats should rotate freely; inside of tubes is coated to prevent static electricity
- May malfunction due to dirt inside, misalignment, sticking markers
- Leaks may occur and could cause hypoxic gas mixture to be delivered
Flowmeters
- Flow moves from left to right, exiting the common manifold to the right of the oxygen flowmeter
- A leak in the first two flowmeters means oxygen continues uninterrupted out to the patient.
- A leak in the oxygen flowmeter, or somewhere beyond it, can cause delivery of a hypoxic mixture.
Studying That Suits You
Use AI to generate personalized quizzes and flashcards to suit your learning preferences.
Related Documents
Description
Test your knowledge on oxygen failure safety valves and pressure-sensor shutoff technologies. This quiz covers various proprietary systems like Datex-Ohmeda and Drager, and details about the oxygen flush valve and low-pressure systems. Enhance your understanding of medical gas delivery safety mechanisms.