Cupping Artefact in Radiography

ComfortingCaesura6782 avatar
ComfortingCaesura6782
·
·
Download

Start Quiz

Study Flashcards

24 Questions

If there are objects lying outside the field of view, especially high attenuation objects such as the arms, this will create _____________ artefacts within the imaged area.

streak

The metal produces a _____________ and photon starvation artefact.

beam-hardening

Motion artefact can be caused by ____________________.

patient

To reduce motion artefact, the patient should be in a ____________________ position.

comfortable

Spiral scanning is a scan parameter solution to reduce ____________________ artefact.

motion

Incomplete projections are caused by objects lying ____________________ the field of view.

outside

A faulty detector can cause a ____________________ artefact.

ring

Cone beam artefact is a particular artefact caused by ____________________ scanners.

multislice

This beam hardening artefact also produces another type of artefact called the ______ artefact.

cupping

The ______ of an object is usually the thickest and, therefore, the beam will become harder in the centre than at the periphery.

centre

This can be corrected with a '______ correction' algorithm.

beam hardening

A ______ filter pre-harden the x-ray beam.

bow-tie

In detector stream 1 the object is only partially imaged and so the attenuation is an average between the dense object and the less dense ______.

background

Partial volume artefact is the result of averaging the linear attenuation coefficient in a voxel that is not uniform in ______.

composition

An object may protrude into the slice in one projection but not in the opposing projection, especially at the ______ of the image where the beam is more divergent.

periphery

These streak artefacts can be caused, for example, when a patient's arms are by their side and are imaged in some projections but not ______.

others

Multislice helical CT results in the acquisition of multiple slices in the same time previously required for a single ______.

slice

The slice acquisition rate (SAR) is one measure of the efficiency of the multislice helical CT ______ system.

imaging

A volume of tissue is being imaged, and this volume is represented by ______ coverage.

z-axis

In computed tomography (CT), the term artefact is applied to any systematic discrepancy between the CT numbers in the reconstructed image and the true ______ coefficients of the object.

attenuation

Causes of image artefacts can be grouped into a few categories, including ______ based artefacts.

physics

Beam hardening is a type of ______ based artefact.

physics

The dense petrous bones in the posterior fossa on a CT head scan can cause a ______ hardening artefact.

beam

Beam hardening artefact results in a higher mean ______ energy.

beam

Study Notes

Cupping Artefact

  • Cupping artefact occurs when the beam becomes harder in the centre of an object, resulting in lower Hounsfield units.
  • This artefact can be corrected with a 'beam hardening correction' algorithm.

Solutions to Beam Hardening

  • Pre-patient filter: absorbs soft x-rays and minimizes beam hardening artefact.
  • Bow-tie filter: pre-hardens the x-ray beam.

Partial Volume Artefact

  • Partial volume artefact occurs when a dense object is partially imaged in a voxel, resulting in an average attenuation between the dense object and the less dense background.
  • Partial volume artefact can only reduce the apparent attenuation of an object, never increase it.

Incomplete Projection

  • Incomplete projection occurs when an object protrudes into the slice in one projection but not in the opposing projection.
  • This can cause streak artefacts due to inconsistencies produced during imaging.
  • Solutions: smaller slice thickness.

Patient Properties

  • Metallic artefacts: caused by high attenuation materials such as metal, which can produce beam-hardening and photon starvation artefacts.
  • Patient motion: can be caused by patient swallowing, breathing, pulsatility of heart and vessels, and patient movement.
  • Solutions: scan parameters (shorten scan time, spiral scanning), patient parameters (breath hold, comfortable patient position, clear instructions).

Incomplete Projections (Out of Field)

  • If objects lie outside the field of view, especially high attenuation objects, it can create streak artefacts within the imaged area.
  • This is due to the object being detected in some projections and not others, leading to inconsistencies in the data.

Scanner Based Artefacts

  • Ring artefact: caused by a faulty detector with a different gain relative to other detectors.
  • Cone beam artefact: caused by multislice scanners, where the x-ray beam becomes cone-shaped instead of fan-shaped, and the area imaged by each detector is a volume instead of a flat plane.
  • Solution: reconstruction algorithm minimizes cone beam artefacts.

Multislice Helical CT

  • Principal advantage: imaging speed, with 80 ms imaging possible.
  • Data Acquisition Rate: multislice helical CT results in acquisition of multiple slices in the same time previously required for a single slice.
  • The principal advantage of multislice helical CT is that a larger volume of tissue can be imaged.
  • It is possible to image the entire body—from head to toe—in a single breath-hold.

Types of Artefacts

  • Artefacts can be classified into four types based on appearance: streaking, shading, rings, and distortion.
  • Causes of image artefacts can be grouped into four categories: physics-based, patient properties, scanner-based, and helical and multislice artefacts.

Physics Based Artefacts

  • Beam hardening: an x-ray beam has photons of different energies that vary around a mean 'beam energy'.
  • As the beam passes through a dense area, the lower energy photons are more likely to be absorbed, and the higher energy photons are more likely to remain, resulting in a higher mean beam energy.
  • This is particularly common in the posterior fossa on a CT head scan due to the dense petrous bones.

This quiz covers the concept of cupping artefact, a type of beam hardening artefact, and its correction in radiography. Learn how the centre of an object is affected by beam hardening and how it can be corrected.

Make Your Own Quizzes and Flashcards

Convert your notes into interactive study material.

Get started for free

More Quizzes Like This

X-Ray Tube Quiz
3 questions

X-Ray Tube Quiz

JawDroppingWalrus1974 avatar
JawDroppingWalrus1974
The X-Ray Intensity Quiz
30 questions
Use Quizgecko on...
Browser
Browser