CHEM 2225 LAB: Calibration Methods Part I

SensationalKhaki avatar
SensationalKhaki
·
·
Download

Start Quiz

Study Flashcards

30 Questions

What is the primary reason why instrumental analysis methods are preferred over classical methods?

Instrumental methods provide greater sensitivity and wider concentration ranges

What statistical question is raised when establishing a calibration graph?

All of the above

What is the purpose of a calibration graph in instrumental analysis?

To establish the relationship between instrument response and analyte concentration

What statistical measure is used to quantify the strength of the linear relationship between the instrument response and analyte concentration?

The product-moment correlation coefficient

Which of the following is NOT a key advantage of instrumental analysis methods over classical methods?

Reduced cost per sample

What is the purpose of the 'line of regression of $y$ on $x$' in the analysis of calibration graphs?

To determine the concentration of an unknown sample from its instrument response

If the calibration plot is linear, how are the errors and confidence limits for the slope and intercept determined?

They are calculated from the residuals of the linear regression line.

When using the calibration plot to determine the concentration of a test sample, what is used to calculate the errors and confidence limits?

The errors and confidence limits of the slope and intercept.

What does the limit of detection represent in the context of a calibration plot?

The lowest concentration of the analyte that can be reliably measured.

What is an important consideration when selecting the range of concentrations for calibration standards?

The range should cover the entire possible range of concentrations for the method.

Why is it important to include the value for a blank in the calibration curve?

To account for any instrument signal that may be present even in the absence of the analyte.

Why is it considered wrong to subtract the blank value from the other standard values before plotting the calibration graph?

It introduces additional errors into the calibration plot.

Why is the calibration curve plotted with instrument signals on the y-axis and standard concentrations on the x-axis?

The procedure assumes that errors are only present in the y-values (instrument signals)

Which statement about the assumptions made when plotting calibration graphs is correct?

It is assumed that if multiple measurements are made on a standard, the resulting y-values will have a normal (Gaussian) error distribution

Which assumption about the errors in the y-values requires further discussion?

The assumption that the errors are independent of the analyte concentration

What is the typical coefficient of variation for routine instrumental analyses mentioned in the text?

2-3% or worse

What has put the assumption of negligible errors in the x-values (standard concentrations) into question?

The advent of high-precision automatic methods with low coefficients of variation

What is the typical error in preparing standard solutions?

0.1% or better

What is the primary approach suggested to improve (narrow) the confidence limits in a calibration experiment?

Both (a) and (b)

In the example provided, what term was dominant in the calculation of confidence limits?

Unity (1)

If the unknown concentration (y$_0$) of 13.5 was calculated as the mean of four determinations, what would be the approximate value of the 95% confidence limits?

6.21 ± 0.36

What does the term 's$_{x_0}$' represent in the context of calculating confidence limits?

The standard deviation of the unknown concentration

What result would show improved precision when measuring multiple times and using the mean?

Both (a) and (c)

What is the purpose of calculating confidence limits in the context of this text?

To estimate the uncertainty associated with the concentration estimate

What is the primary purpose of making too many replicate measurements of y0?

To generate more work for only a small additional benefit

How is the limit of detection of an analyte defined?

The concentration that gives an instrument signal significantly different from the blank or background signal

What is the recommended formula for the limit of quantification (LOQ) according to the text?

yB + 10sB

What is the relationship between the limit of detection (LOD) and the limit of quantification (LOQ) according to the text?

Statutory bodies recommend the LOD, which is a lower limit for precise quantitative measurements

How can the standard deviation of the blank (sB) be estimated when using regression line calibration?

sB can be estimated as the standard error of the regression line (sy/x)

What is the main reason that the definition of the limit of detection must be provided whenever it is cited?

To provide context for interpreting the significance of the reported limit of detection

Study Notes

Calibration Methods: Regression and Correlation

  • Instrumental analysis methods offer extreme sensitivity, wide concentration ranges, multi-analyte capability, automation, miniaturized systems, and computer interfacing.

Calibration Graphs

  • A calibration graph is established, and unknown concentrations can be determined via interpolation.
  • The graph is subject to errors, and the best straight line (or curve) through these points must be determined.

Problems with Calibration

  • Is the calibration graph linear or curved?
  • What are the errors and confidence limits for the slope and intercept of the line?
  • What are the errors and confidence limits for the determined concentration?
  • What is the limit of detection of the method?

Aspects to Consider when Plotting Calibration Graphs

  • Calibration standards should cover the whole range of concentrations required.
  • Include a 'blank' value in the calibration curve to account for instrumental signal.
  • Do not subtract the blank value from standard values before plotting the graph, as this gives incorrect information on errors.

Calibration Curve Assumptions

  • Errors are in the y-values (instrument signals), and standard concentrations (x-values) are assumed error-free.
  • The magnitude of errors in y-values is independent of the analyte concentration.

Calculation of Concentration and Random Error

  • The concentration and its random error can be calculated using equations.
  • Increasing the number of calibration points or making multiple measurements can improve precision.

Limits of Detection

  • The limit of detection is the concentration that gives an instrument signal significantly different from the 'blank' or 'background' signal.
  • The limit of quantification (LOQ) is the lower limit for precise quantitative measurements.
  • LOQ can be estimated using the formula yB + 10sB, where yB is the blank signal and sB is the standard deviation of the blank.

This quiz covers topics related to calibration methods in Analytical Chemistry III Laboratory, including regression, correlation, calibration graphs, and error calculations. Learn about instrumental analysis and the limits of detection in this quiz.