Assessment BeeAT Q1 2025

Choose a study mode

Play Quiz
Study Flashcards
Spaced Repetition
Chat to Lesson

Podcast

Play an AI-generated podcast conversation about this lesson

Questions and Answers

What is the total BeeAT model at JLLM? (Including new model)

  • 3
  • 5
  • 7 (correct)
  • 6

Which statement is true about BeeAT?

  • Use to correct abnormal heart rhythms (arrhythmia) in patients. (correct)
  • To monitor the temperature of the esophageal during catheter ablation for atrial fibrillation
  • Implanted into the bile duct, using endoscope to drain out accumulated bile by securing a passage through the obstructed bile duct and improve fatal symptoms.
  • A left hand model which dial can be manipulated using the left hand.

What is the approximate effective length of the 20884-80S BeeAT model?

  • 799 ± 5 mm
  • 800 ± 15 mm
  • 810 ± 5 mm
  • 799 ± 10 mm (correct)

For the S208282-SAOCR model, what is the approximate electrode gap between Ring #19 and Ring #20?

<p>45 ± 1 mm (D)</p> Signup and view all the answers

What are the additional item need to be check at Dimension Inspection for S208282-SAOCR model?

<p>Distance between laser marking and ring (A)</p> Signup and view all the answers

According to the electrode location diagram, which BeeAT model has electrodes #8-#9 located at 50±1?

<p>S20884-80S (A)</p> Signup and view all the answers

Which of the following BeeAT models has electrode group gaps between #8-#9 at 90±1?

<p>S20884-80L (A)</p> Signup and view all the answers

Which BeeAT model includes a two-step laser marking process as part of its manufacturing?

<p>S208282-SAOCR (B)</p> Signup and view all the answers

What is the return product BeeAT for last year?

<p>Handle Gap (C)</p> Signup and view all the answers

Which statement refer to Bending for S208282-SAOCR BeeAT model?

<p>For maximum curve, make sure ring #9 must inside the shaded box (D)</p> Signup and view all the answers

Flashcards

Objective of BeeAT model transfer?

Transferring the BeeAT model to JLLM and expanding the production base.

BeeAT S20884-80S Electrode Location

The ring has electrodes numbered from #8 to #9.

BeeAT S208282 Electrode Location

The ring has electrodes numbered from #8 to #17.18.

Purpose of S208282-SAOCR model

To prevent other companies from using it with their products, protecting intracardiac defibrillation catheter business.

Signup and view all the flashcards

Documents to revise for SAOCR Model

Documents such as Inspection Batch Record, Inspection Record, QC Flow Chart, Ring Configuration, Ring Gap & Shaft Spec

Signup and view all the flashcards

Distance Checks for laser markings

First ensure #20-width lasers markings are 135±3 mm , second ensure to perform 234±10mm between #20 and the tip.

Signup and view all the flashcards

Study Notes

BeeAT New Models

  • New BeeAT models added include: S20884-80S, S20884-80L, S208282-SAOC2, and S208282-SAOCR

Objective and Target Models

  • The objective is to transfer the BeeAT model to JLLM and expand the production base to respond to production volume changes
  • Due to expected decrease in BeeAT production volume next fiscal year, target models will be produced by JLLM to compensate

BeeAt Ring Configuration

  • The electrode configuration of the BeeAT ring varies across different models.
  • S20884-80S electrode is #8-#9 : 50±1.
  • S20884-80L is #8-#9 : -90±1.
  • S208282-SAOC2 is #8-#17.18 : 10±1.

Model Comparison

  • Models compared are : 20884-80M, 20884-80S, 20884-80L, 208282-SAOC and 208282-SAOC2.
  • All have an effective length of 799±10mm
  • All have a sheath 80M

Model S20884-80S

  • Electrode gap between #8~#9 is 70±1

Model S20884-80L

  • Electrode gap between #8~#9 is 90±1

Model 208282-SAOC

  • The Electrode gap between #817.18: 20±1, and #8#9: 70±1

Model 208282-SAOC2

  • The Electrode gap between #8~17.18: 10±1
  • Electrode gap between #8~#9: 70±1, and #16~#19 is : 15±1

MP Trial BeeAT

  • MP Trial BeeAT models, lot numbers, quantities, plan dates, and ship dates are available, including S20884-80S, S20884-80L, and S208282-SAOC2

Background and Configuration of S208282-SAOCR

  • The purpose of S208282-SAOCR is to prevent transfer to other companies' products by using it with a mapping device
  • Basic configuration of S208282-SAOC shown

Ring Configuration S208282-SAOCR

  • The electrode configuration utilizes electrodes #1 to #8, #9 to #16, #17 and #18, and #19 to #20

Ring Configuration S208282-SAOCR

  • The model S208282-SAOCR has an effective length of 799±10mm
  • The electrode gap between #817.18 : 20±1 and the electrode gap between #8#9 : 70±1. and #16~#19 : 20±1
Model 208282-SAOCR electrode spacing (mm):
  • #1~#8: 2±0.5
  • #817.18 : 2±1 -#9#16 : 2±
  • #19~#20: 45±

Shared Dimension Item S208282-SAOCR

  • S208282-SAOCR has specific requirements for distance between laser markings and #20, distance between #20 and tip, outer diameter, and PVC tube length

Documents to Revise Relating To the SAOCR Model

  • F-8260-BAT-02 Rev4 BAT Final Inspection Batch Record
  • F-8260-BAT-01 Rev4 BAT Final Inspection Record
  • QF-8260-BAT-001 Rev7 BeeAT Final Inspection QC Flow Chart
  • Ap-QF-8260-BAT-001.02 Rev1 BeeAT Ring Configuration
  • Ap-QF-8260-BAT-001.04 Rev1 BeeAT Ring Gap & Shaft Spec