Podcast
Questions and Answers
What is the primary function of Automatic Exposure Control (AEC) in radiography?
What is the primary function of Automatic Exposure Control (AEC) in radiography?
- Determining the exposure time and amount of radiation to produce a quality image. (correct)
- Selecting the appropriate focal spot size for the examination.
- Adjusting kVp to optimize image contrast.
- Controlling the mA to reduce patient dose.
In an AEC system, what happens once a predetermined amount of radiation has been detected?
In an AEC system, what happens once a predetermined amount of radiation has been detected?
- The mA is decreased to reduce the heat load on the tube.
- The kVp is automatically increased to improve penetration.
- The X-ray exposure is terminated. (correct)
- The technologist is alerted to manually adjust the exposure settings.
What is the primary difference between phototimers and ionization chambers in AEC systems?
What is the primary difference between phototimers and ionization chambers in AEC systems?
- Phototimers use high kVp, while ionization chambers use low kVp.
- Phototimers measure radiation before it reaches the patient, while ionization chambers measure it after it exits the patient.
- Phototimers are entrance-type devices, while ionization chambers are exit-type devices.
- Phototimers are exit-type devices, while ionization chambers are entrance-type devices. (correct)
Why is it crucial for radiographers to be aware of the mAs readout when using AEC?
Why is it crucial for radiographers to be aware of the mAs readout when using AEC?
How does kVp selection affect the exposure time when using AEC?
How does kVp selection affect the exposure time when using AEC?
What is the significance of the minimum response time in AEC systems?
What is the significance of the minimum response time in AEC systems?
Why is it important to set a backup time when using AEC?
Why is it important to set a backup time when using AEC?
What does the "+1" setting typically represent on the exposure adjustment control of an AEC system?
What does the "+1" setting typically represent on the exposure adjustment control of an AEC system?
Why is accurate centering of the anatomical region of interest over the selected AEC detector(s) important?
Why is accurate centering of the anatomical region of interest over the selected AEC detector(s) important?
What is a potential consequence of using a detector size that is larger than the body part being imaged when using AEC?
What is a potential consequence of using a detector size that is larger than the body part being imaged when using AEC?
How does the presence of additive pathology typically affect exposure when using AEC?
How does the presence of additive pathology typically affect exposure when using AEC?
What is the primary purpose of an anatomically programmed technique (APT) system?
What is the primary purpose of an anatomically programmed technique (APT) system?
What is the function of exposure technique charts in radiography?
What is the function of exposure technique charts in radiography?
In a variable kVp/fixed mAs technique chart, how is the kVp adjusted for changes in part thickness?
In a variable kVp/fixed mAs technique chart, how is the kVp adjusted for changes in part thickness?
What is a key advantage of using a fixed kVp/variable mAs technique chart?
What is a key advantage of using a fixed kVp/variable mAs technique chart?
What tools are essential for the development of an exposure technique chart?
What tools are essential for the development of an exposure technique chart?
Why might AEC be less effective for pediatric patients, especially infants?
Why might AEC be less effective for pediatric patients, especially infants?
How should exposure factors be adjusted when imaging a patient with a plaster cast?
How should exposure factors be adjusted when imaging a patient with a plaster cast?
What is the general recommendation for adjusting kVp when imaging a part affected by an additive pathology?
What is the general recommendation for adjusting kVp when imaging a part affected by an additive pathology?
How do positive contrast agents like barium and iodine affect x-ray attenuation, and what adjustment is typically required?
How do positive contrast agents like barium and iodine affect x-ray attenuation, and what adjustment is typically required?
Flashcards
Automatic Exposure Control (AEC)
Automatic Exposure Control (AEC)
A tool that assists in determining the amount of radiation exposure to produce a quality image.
AEC Detecting Devices
AEC Detecting Devices
These detecting devices are referred to as sensors, chambers, cells, or detectors. They measure radiation exposure reaching the image receptor.
Phototiming
Phototiming
An AEC device that uses photomultiplier tubes or photodiodes.
Photomultiplier Tube (PMT)
Photomultiplier Tube (PMT)
Signup and view all the flashcards
Solid-State Photodiode
Solid-State Photodiode
Signup and view all the flashcards
Phototimer AEC
Phototimer AEC
Signup and view all the flashcards
Ionization Chambers
Ionization Chambers
Signup and view all the flashcards
Ionization Chamber AEC
Ionization Chamber AEC
Signup and view all the flashcards
mAs Readout
mAs Readout
Signup and view all the flashcards
AEC
AEC
Signup and view all the flashcards
Minimum Response Time
Minimum Response Time
Signup and view all the flashcards
Backup Time
Backup Time
Signup and view all the flashcards
Exposure Adjustment
Exposure Adjustment
Signup and view all the flashcards
Detector Selection
Detector Selection
Signup and view all the flashcards
Collimation
Collimation
Signup and view all the flashcards
Anatomically Programmed Technique
Anatomically Programmed Technique
Signup and view all the flashcards
Exposure Technique Charts
Exposure Technique Charts
Signup and view all the flashcards
Technique Charts
Technique Charts
Signup and view all the flashcards
Plaster cast
Plaster cast
Signup and view all the flashcards
Additive Diseases
Additive Diseases
Signup and view all the flashcards
Study Notes
Automatic Exposure Control (AEC)
- AEC is a tool used for determining radiation exposure to produce quality images
- In operation, the x-ray exposure terminates once a radiation amount transmits through a patient
- AEC helps determine exposure time and radiation exposure delivered to the image receptor (IR)
- AEC radiation detectors convert radiation to electrical signals
- Exposure time terminates when the electrical signal indicates a predetermined radiation amount is reached
- Most AEC systems use three radiation detectors set in a specific arrangement
- The radiographer selects the detector configuration to measure radiation exposure reaching the IR
- Detecting devices: sensors, chambers, cells, or detectors
Types of AEC Devices
- Phototimers and ionization chambers are the two main types of AEC devices
- Phototimers were in early AEC systems, hence the term "phototiming"
- Modern systems no longer use phototimers making "phototiming" an outdated term
- Phototiming uses an AEC device with photomultiplier tubes or photodiodes
- Photomultiplier tubes (PMT) detect and amplify low light levels
- Solid-state photodiodes are photodetectors converting light to electrical current through the photovoltaic effect
- Solid-state photodiodes operate using solid semiconductors
Phototimers
- Phototimers use a fluorescent light-producing screen with a device that transforms light into electricity
- A photodiode is a solid-state device converting light into an electrical signal
- Phototimer AECs were exit-type devices, placed behind the IR to measure the radiation after exiting the IR
- Light paddles coated with fluorescent material served as detectors
- Once radiation interacted, light was converted to electricity by photomultiplier tubes/photodiodes
- When enough electric charge was reached, the exposure timer stopped the exposure
Ionization Chambers
- Ionization chambers or solid-state systems are used more than phototimers today
- Remnant radiation from the patient creates ionizations or electron pairs in a sensing area
- The ionizations are proportional to the x-ray exposure
- Ion pairs create an electrical current, generating an amplified signal
- The electrical signal tells the x-ray generator to end the exposure
- An ionization chamber is a hollow cell containing air connected to the timer circuit by electrical wire
- Ionization chamber AEC devices are entrance type devices, positioned in front of the IR
- Radiation interacts with the sensing detectors before reaching the IR
- When radiation hits the ionization chamber, the air ionizes and creates an electrical charge
- The charge flows to the timer circuit, and when sufficient exposure is achieved, the timer is tripped
- The electrical charge is proportional to the radiation the ionization chamber is exposed to
Milliampere-Seconds (mAs) Readout
- AEC systems determine the total radiation (mAs) needed for the proper IR exposure automatically
- Many units display the mAs immediately after exposure
- Radiographers should note mAs readouts to help switch to manual techniques, when needed
Kilovoltage Peak (kVp) and mAs Readout
- AEC solely manages the quantity of radiation reaching the IR, not influencing image characteristics like contrast
- kVp selection should be independent of AEC use and based on required penetration and subject contrast
- The selected kVp affects exposure time, as higher kVp needs less mAs for an earlier termination, reducing patient dose
- Lower kVp values need larger mAs values, extending exposure time and increasing patient dose
- In systems allowing independent mA and time settings, select mA irrespective of AEC usage
- Chosen mA has an inverse relationship with exposure time: low mA means longer exposure and vice versa for AEC
Minimum Response Time
- Refers to the shortest possible exposure duration that a system can produce
- Generally, minimum response time is longer in systems with AEC than other radiographic timers
- Modern AEC systems feature minimum response times of 1ms
- If the required time to produce quality exposure falls under the AEC's minimum response time, overexposure can occur
- Example: Pediatric X-rays use short times to decrease movement but if settings are less than 1ms, increased radiation happens
Backup Time
- It defines the maximum length an X-ray exposure runs when utilizing an AEC system
- Technologists are responsible to set the backup time
- Backup time serves as a safety feature during AEC failure or incorrect equipment use
- It safeguards the patient from excessive exposure and shields the tube from exceeding heat load capacity
- If automatically set, the limit stops at 600 mAs when the equipment runs over 50 kVp
- If the technologist chooses the backup time, setting it at 150% to 200% of the anticipated exposure duration is advised
Exposure Adjustment
- Permits technologists to modify preset radiation detection thresholds using density settings
- Adjustment options on the control panel are labeled from -2 to +3
- Each selection alters exposure time by a pre-set increment
- A 25% increment means the timer may be either raised or lowered from standard in this measure
- For example, +2 raises the timer by 50%
Detector Selection
- AEC Systems allow various detector combinations or offer five detectors for flexibility
- Selected detectors actively assess radiation when exposing the image and the electric signals are averaged
- Detectors measuring the most exposure influence the total exposure greatly
- Radiographers should position the detectors in line or overlapping with anatomical regions, based on interest
- Incorrect detector selection leads to either over or underexposure of images
- Technologist need to check image quality, mAs and exposure while using the AEC System
- If the region of interest is not precisely aligned with operational sensors
- It can cause exposure issues, resulting in a computer adjusting image flaws and risks image/exposure issues
- Underexposure can cause quantum noise
- Overexposure results in higher radiation hurting the image displayed
Detector Size
- Detector sizes and shapes are fixed in AEC systems and cannot be changed
- If the target body part is smaller than chosen detectors, the exposure will end quickly
- The primary beam exposes some detectors prematurely, causing underexposure
- For smaller parts, the manual technique should be used
Compensating Issues: Patient Considerations
- Increased patient thickness requires longer exposure with AEC
- Conditions like gas in the bowels lead to quicker termination and underexposure
- Additive pathology, contrast media, metal or prosthetic devices may lengthen exposure
Compensating Issues: Collimation
- Inadequate collimation can affect radiation exposure levels
- Insufficient collimation causes increase scatter at the detectors, causing quick termination
- Excessive collimation increases exposure time, as detectors need sufficient exposure initially
Compensating Issues: Image Receptor Variations
- Exchanging IR types can cause problems when devices are preset
- Since calibration sets the AEC device to specific IR/radiation amounts
- AEC devices cannot sense IR types and will produce inaccurate exposure settings
Anatomically Programmed Technique
- This system enables the technologist to select an anatomical area on the console panel
- After selecting the anatomical part (e.g., chest) and projection (e.g., PA or Lat), the system displays pre-programmed exposure settings
Exposure Technique Charts
- Exposure technique charts are pre-established guidelines that assist technologists in selecting the appropriate exposure factors for x-ray examinations
- It includes kVp, mAs, IR type, grid, and SID
- Technique charts promote consistent image quality, decrease the amount of retakes needed, and lessens patient exposure
- The charts are designed for average patients and overlook unusual conditions
- Part measurement accuracy is vital
- Calipers measure part thickness at the location of the central ray midpoint or the thickest portion of the area being radiographed
- Wide exposure latitude equals image quality
Types of Technique Charts
- Primary types are variable kVp/fixed mAs and fixed kVp/variable mAs
Variable kVp/Fixed mAs Technique Chart
- Easy to formulate, decreasing subject contrast and may not ensure penetration
- With this technique chart part measurement must be accurate to ensure that the 2 kVp adjustment is accurately applied
- kVp increases as part thickness increases
- For every 1cm (0.4 inch) increase in part thickness, kVp increases by 2 and mAs stays constant
- Baseline kVp then adjusts for part thickness
- For example if settings are 70 kVp and 10 mAs when imaging a knee but part size goes from 10 cm and up, settings change to 72 kVp/10mAs
- Less accurate for extreme ranges in part size
Fixed kVp/Variable mAs Technique Chart
- The Fixed chart selects the ideal kilovoltage and sets adjustment of mass
- MAs should be chosen depending on thickness
- For every few centimeters, the mast increase or decrease
- Example, if the Kilovoltage is at 70 and mass is at 10 for a patient
- One might have to select twenty mass and keep the Kilovoltage to use adequate settings
Exposure Technique Chart Development
- Calipers, phantoms, and calculations are used for tool chart development
- After test images, extrapolate exposure techniques
- Comparative anatomy says parts of the same side can be read with similar exposure if kVp is well set
Special Considerations: Pediatric Patients
- Pediatric patients require lower kVp and mAs values
- Shortened exposure times are required because of children's inability to suspend breathing
- AEC use may be limited by minimum response time
- Smaller size may not cover the detector, which can lead to inaccurate technique
Special Considerations: Geriatric Patients
- Aging patients may have physical changes, such as sensory and thin broken skin
- Exposure techniques may be decreased
Special Considerations: Bariatric Patients
- An obese patient is someone with a BMI higher than 30
- Imaging a bariatric patient means considering table weight limits and equipment size
- Higher kVp and mAs is needed
- Grids are useful to reduce scatter
Special Considerations: Projections and Positions
- Exposure factors are altered per specific projection
- Lateral knees warrant more technique
Special Considerations: Casts and Splints
- Casts are made from attenuating materials
- Fiberglass casts do not need exposure adjustments
- Plaster casts require exposure increase
- Splints require no adjustments
Special Considerations: Pathologic Conditions
- Additive diseases increases an anatomical part's absorption characteristics, such as ascites, tumors etc
- Destructive decreases absorption, such as atrophy, emphysema ect
- Kilovoltage should increase if adding, and decrease if decreasing
- Change should be at least 15%
Special Considerations: Soft Tissue
- In digital imaging, brightness and contrast can be used instead
- A decrease in mass can display the larynx
Special Considerations: Contrast Media
- Positive contrast calls for higher exposure and are radiopaque
- Negative is air and are normally less exposed
Mastering exposure variable effects is essential for technologists
Studying That Suits You
Use AI to generate personalized quizzes and flashcards to suit your learning preferences.