Automatic Exposure Control (AEC)

Choose a study mode

Play Quiz
Study Flashcards
Spaced Repetition
Chat to Lesson

Podcast

Play an AI-generated podcast conversation about this lesson
Download our mobile app to listen on the go
Get App

Questions and Answers

What is the primary function of Automatic Exposure Control (AEC) in radiography?

  • Determining the exposure time and amount of radiation to produce a quality image. (correct)
  • Selecting the appropriate focal spot size for the examination.
  • Adjusting kVp to optimize image contrast.
  • Controlling the mA to reduce patient dose.

In an AEC system, what happens once a predetermined amount of radiation has been detected?

  • The mA is decreased to reduce the heat load on the tube.
  • The kVp is automatically increased to improve penetration.
  • The X-ray exposure is terminated. (correct)
  • The technologist is alerted to manually adjust the exposure settings.

What is the primary difference between phototimers and ionization chambers in AEC systems?

  • Phototimers use high kVp, while ionization chambers use low kVp.
  • Phototimers measure radiation before it reaches the patient, while ionization chambers measure it after it exits the patient.
  • Phototimers are entrance-type devices, while ionization chambers are exit-type devices.
  • Phototimers are exit-type devices, while ionization chambers are entrance-type devices. (correct)

Why is it crucial for radiographers to be aware of the mAs readout when using AEC?

<p>To have information to switch to manual techniques as needed. (D)</p> Signup and view all the answers

How does kVp selection affect the exposure time when using AEC?

<p>Higher kVp leads to shorter exposure times. (B)</p> Signup and view all the answers

What is the significance of the minimum response time in AEC systems?

<p>It refers to the shortest exposure time the system can produce, regardless of radiation detected. (C)</p> Signup and view all the answers

Why is it important to set a backup time when using AEC?

<p>As a safety mechanism to protect the patient and tube. (A)</p> Signup and view all the answers

What does the "+1" setting typically represent on the exposure adjustment control of an AEC system?

<p>An increase in the predetermined exposure level needed to terminate the timer. (A)</p> Signup and view all the answers

Why is accurate centering of the anatomical region of interest over the selected AEC detector(s) important?

<p>To prevent underexposure or overexposure of the image receptor. (B)</p> Signup and view all the answers

What is a potential consequence of using a detector size that is larger than the body part being imaged when using AEC?

<p>Underexposure of the image due to premature termination of the exposure. (B)</p> Signup and view all the answers

How does the presence of additive pathology typically affect exposure when using AEC?

<p>It typically requires an increase in exposure to penetrate the denser tissue. (B)</p> Signup and view all the answers

What is the primary purpose of an anatomically programmed technique (APT) system?

<p>To provide preprogrammed exposure factors based on the anatomical area being imaged. (D)</p> Signup and view all the answers

What is the function of exposure technique charts in radiography?

<p>To provide guidelines for selecting exposure factors. (C)</p> Signup and view all the answers

In a variable kVp/fixed mAs technique chart, how is the kVp adjusted for changes in part thickness?

<p>kVp is increased as part thickness increases. (B)</p> Signup and view all the answers

What is a key advantage of using a fixed kVp/variable mAs technique chart?

<p>It provides more consistent image quality and contrast. (D)</p> Signup and view all the answers

What tools are essential for the development of an exposure technique chart?

<p>Calipers, phantoms, and calculators. (C)</p> Signup and view all the answers

Why might AEC be less effective for pediatric patients, especially infants?

<p>The minimum response time may result in over exposure. (B)</p> Signup and view all the answers

How should exposure factors be adjusted when imaging a patient with a plaster cast?

<p>Increase mAs and/or kVp. (A)</p> Signup and view all the answers

What is the general recommendation for adjusting kVp when imaging a part affected by an additive pathology?

<p>Increase kVp by at least 15%. (D)</p> Signup and view all the answers

How do positive contrast agents like barium and iodine affect x-ray attenuation, and what adjustment is typically required?

<p>Increase attenuation, so increase exposure factors. (A)</p> Signup and view all the answers

Flashcards

Automatic Exposure Control (AEC)

A tool that assists in determining the amount of radiation exposure to produce a quality image.

AEC Detecting Devices

These detecting devices are referred to as sensors, chambers, cells, or detectors. They measure radiation exposure reaching the image receptor.

Phototiming

An AEC device that uses photomultiplier tubes or photodiodes.

Photomultiplier Tube (PMT)

A device used to detect and amplify low levels of light.

Signup and view all the flashcards

Solid-State Photodiode

A solid-state device that converts light into an electrical signal.

Signup and view all the flashcards

Phototimer AEC

Considered exit-type devices in AEC because detectors are positioned behind the Image Receptor.

Signup and view all the flashcards

Ionization Chambers

Replaced phototimers, remnant radiation from the patient is received into a sensing area that creates ionizations or electron pairs.

Signup and view all the flashcards

Ionization Chamber AEC

AEC devices considered entrance type because the detectors are positioned in front of the IR.

Signup and view all the flashcards

mAs Readout

The total amount of radiation (mAs) required to produce the appropriate exposure to the IR is determined by the system.

Signup and view all the flashcards

AEC

Controls only the quantity of radiation reaching the IR and has no effect on image characteristics, such as contrast.

Signup and view all the flashcards

Minimum Response Time

Refers to the shortest exposure time that the system can produce.

Signup and view all the flashcards

Backup Time

The maximum length of time for which the x-ray exposure will continue when using an AEC system.

Signup and view all the flashcards

Exposure Adjustment

Allows the technologist to adjust the amount of preset radiation detection values (density controls).

Signup and view all the flashcards

Detector Selection

AEC systems allows the tech to select any combination of 1, 2, or 3 detectors. Select detectors that would be superimposed by the anatomical region of interest.

Signup and view all the flashcards

Collimation

Excessive or insufficient collimation may affect the amount of exposure to the IR.

Signup and view all the flashcards

Anatomically Programmed Technique

Allows the technologist to select the anatomic area in the console panel. It will have a preprogrammed set of exposure factors.

Signup and view all the flashcards

Exposure Technique Charts

Pre-established guidelines used by technologists to select exposure factors for x-ray exams.

Signup and view all the flashcards

Technique Charts

Charts are instrumental in the production of consistent quality images, reduction in repeats, and reduction in patient exposure.

Signup and view all the flashcards

Plaster cast

Requires an increase in exposure factors. The technologist may measure the part with the plaster cast on and adjust the technique based on the technical factors used for the part without a cast.

Signup and view all the flashcards

Additive Diseases

Pathologies that increase the absorption characteristics of the part, which makes the part more difficult to penetrate.

Signup and view all the flashcards

Study Notes

Automatic Exposure Control (AEC)

  • AEC is a tool used for determining radiation exposure to produce quality images
  • In operation, the x-ray exposure terminates once a radiation amount transmits through a patient
  • AEC helps determine exposure time and radiation exposure delivered to the image receptor (IR)
  • AEC radiation detectors convert radiation to electrical signals
  • Exposure time terminates when the electrical signal indicates a predetermined radiation amount is reached
  • Most AEC systems use three radiation detectors set in a specific arrangement
  • The radiographer selects the detector configuration to measure radiation exposure reaching the IR
  • Detecting devices: sensors, chambers, cells, or detectors

Types of AEC Devices

  • Phototimers and ionization chambers are the two main types of AEC devices
  • Phototimers were in early AEC systems, hence the term "phototiming"
  • Modern systems no longer use phototimers making "phototiming" an outdated term
  • Phototiming uses an AEC device with photomultiplier tubes or photodiodes
  • Photomultiplier tubes (PMT) detect and amplify low light levels
  • Solid-state photodiodes are photodetectors converting light to electrical current through the photovoltaic effect
  • Solid-state photodiodes operate using solid semiconductors

Phototimers

  • Phototimers use a fluorescent light-producing screen with a device that transforms light into electricity
  • A photodiode is a solid-state device converting light into an electrical signal
  • Phototimer AECs were exit-type devices, placed behind the IR to measure the radiation after exiting the IR
  • Light paddles coated with fluorescent material served as detectors
  • Once radiation interacted, light was converted to electricity by photomultiplier tubes/photodiodes
  • When enough electric charge was reached, the exposure timer stopped the exposure

Ionization Chambers

  • Ionization chambers or solid-state systems are used more than phototimers today
  • Remnant radiation from the patient creates ionizations or electron pairs in a sensing area
  • The ionizations are proportional to the x-ray exposure
  • Ion pairs create an electrical current, generating an amplified signal
  • The electrical signal tells the x-ray generator to end the exposure
  • An ionization chamber is a hollow cell containing air connected to the timer circuit by electrical wire
  • Ionization chamber AEC devices are entrance type devices, positioned in front of the IR
  • Radiation interacts with the sensing detectors before reaching the IR
  • When radiation hits the ionization chamber, the air ionizes and creates an electrical charge
  • The charge flows to the timer circuit, and when sufficient exposure is achieved, the timer is tripped
  • The electrical charge is proportional to the radiation the ionization chamber is exposed to

Milliampere-Seconds (mAs) Readout

  • AEC systems determine the total radiation (mAs) needed for the proper IR exposure automatically
  • Many units display the mAs immediately after exposure
  • Radiographers should note mAs readouts to help switch to manual techniques, when needed

Kilovoltage Peak (kVp) and mAs Readout

  • AEC solely manages the quantity of radiation reaching the IR, not influencing image characteristics like contrast
  • kVp selection should be independent of AEC use and based on required penetration and subject contrast
  • The selected kVp affects exposure time, as higher kVp needs less mAs for an earlier termination, reducing patient dose
  • Lower kVp values need larger mAs values, extending exposure time and increasing patient dose
  • In systems allowing independent mA and time settings, select mA irrespective of AEC usage
  • Chosen mA has an inverse relationship with exposure time: low mA means longer exposure and vice versa for AEC

Minimum Response Time

  • Refers to the shortest possible exposure duration that a system can produce
  • Generally, minimum response time is longer in systems with AEC than other radiographic timers
  • Modern AEC systems feature minimum response times of 1ms
  • If the required time to produce quality exposure falls under the AEC's minimum response time, overexposure can occur
  • Example: Pediatric X-rays use short times to decrease movement but if settings are less than 1ms, increased radiation happens

Backup Time

  • It defines the maximum length an X-ray exposure runs when utilizing an AEC system
  • Technologists are responsible to set the backup time
  • Backup time serves as a safety feature during AEC failure or incorrect equipment use
  • It safeguards the patient from excessive exposure and shields the tube from exceeding heat load capacity
  • If automatically set, the limit stops at 600 mAs when the equipment runs over 50 kVp
  • If the technologist chooses the backup time, setting it at 150% to 200% of the anticipated exposure duration is advised

Exposure Adjustment

  • Permits technologists to modify preset radiation detection thresholds using density settings
  • Adjustment options on the control panel are labeled from -2 to +3
  • Each selection alters exposure time by a pre-set increment
  • A 25% increment means the timer may be either raised or lowered from standard in this measure
  • For example, +2 raises the timer by 50%

Detector Selection

  • AEC Systems allow various detector combinations or offer five detectors for flexibility
  • Selected detectors actively assess radiation when exposing the image and the electric signals are averaged
  • Detectors measuring the most exposure influence the total exposure greatly
  • Radiographers should position the detectors in line or overlapping with anatomical regions, based on interest
  • Incorrect detector selection leads to either over or underexposure of images
  • Technologist need to check image quality, mAs and exposure while using the AEC System
  • If the region of interest is not precisely aligned with operational sensors
  • It can cause exposure issues, resulting in a computer adjusting image flaws and risks image/exposure issues
  • Underexposure can cause quantum noise
  • Overexposure results in higher radiation hurting the image displayed

Detector Size

  • Detector sizes and shapes are fixed in AEC systems and cannot be changed
  • If the target body part is smaller than chosen detectors, the exposure will end quickly
  • The primary beam exposes some detectors prematurely, causing underexposure
  • For smaller parts, the manual technique should be used

Compensating Issues: Patient Considerations

  • Increased patient thickness requires longer exposure with AEC
  • Conditions like gas in the bowels lead to quicker termination and underexposure
  • Additive pathology, contrast media, metal or prosthetic devices may lengthen exposure

Compensating Issues: Collimation

  • Inadequate collimation can affect radiation exposure levels
  • Insufficient collimation causes increase scatter at the detectors, causing quick termination
  • Excessive collimation increases exposure time, as detectors need sufficient exposure initially

Compensating Issues: Image Receptor Variations

  • Exchanging IR types can cause problems when devices are preset
  • Since calibration sets the AEC device to specific IR/radiation amounts
  • AEC devices cannot sense IR types and will produce inaccurate exposure settings

Anatomically Programmed Technique

  • This system enables the technologist to select an anatomical area on the console panel
  • After selecting the anatomical part (e.g., chest) and projection (e.g., PA or Lat), the system displays pre-programmed exposure settings

Exposure Technique Charts

  • Exposure technique charts are pre-established guidelines that assist technologists in selecting the appropriate exposure factors for x-ray examinations
  • It includes kVp, mAs, IR type, grid, and SID
  • Technique charts promote consistent image quality, decrease the amount of retakes needed, and lessens patient exposure
  • The charts are designed for average patients and overlook unusual conditions
  • Part measurement accuracy is vital
  • Calipers measure part thickness at the location of the central ray midpoint or the thickest portion of the area being radiographed
  • Wide exposure latitude equals image quality

Types of Technique Charts

  • Primary types are variable kVp/fixed mAs and fixed kVp/variable mAs

Variable kVp/Fixed mAs Technique Chart

  • Easy to formulate, decreasing subject contrast and may not ensure penetration
  • With this technique chart part measurement must be accurate to ensure that the 2 kVp adjustment is accurately applied
  • kVp increases as part thickness increases
  • For every 1cm (0.4 inch) increase in part thickness, kVp increases by 2 and mAs stays constant
  • Baseline kVp then adjusts for part thickness
  • For example if settings are 70 kVp and 10 mAs when imaging a knee but part size goes from 10 cm and up, settings change to 72 kVp/10mAs
  • Less accurate for extreme ranges in part size

Fixed kVp/Variable mAs Technique Chart

  • The Fixed chart selects the ideal kilovoltage and sets adjustment of mass
  • MAs should be chosen depending on thickness
  • For every few centimeters, the mast increase or decrease
  • Example, if the Kilovoltage is at 70 and mass is at 10 for a patient
  • One might have to select twenty mass and keep the Kilovoltage to use adequate settings

Exposure Technique Chart Development

  • Calipers, phantoms, and calculations are used for tool chart development
  • After test images, extrapolate exposure techniques
  • Comparative anatomy says parts of the same side can be read with similar exposure if kVp is well set

Special Considerations: Pediatric Patients

  • Pediatric patients require lower kVp and mAs values
  • Shortened exposure times are required because of children's inability to suspend breathing
  • AEC use may be limited by minimum response time
  • Smaller size may not cover the detector, which can lead to inaccurate technique

Special Considerations: Geriatric Patients

  • Aging patients may have physical changes, such as sensory and thin broken skin
  • Exposure techniques may be decreased

Special Considerations: Bariatric Patients

  • An obese patient is someone with a BMI higher than 30
  • Imaging a bariatric patient means considering table weight limits and equipment size
  • Higher kVp and mAs is needed
  • Grids are useful to reduce scatter

Special Considerations: Projections and Positions

  • Exposure factors are altered per specific projection
  • Lateral knees warrant more technique

Special Considerations: Casts and Splints

  • Casts are made from attenuating materials
  • Fiberglass casts do not need exposure adjustments
  • Plaster casts require exposure increase
  • Splints require no adjustments

Special Considerations: Pathologic Conditions

  • Additive diseases increases an anatomical part's absorption characteristics, such as ascites, tumors etc
  • Destructive decreases absorption, such as atrophy, emphysema ect
  • Kilovoltage should increase if adding, and decrease if decreasing
  • Change should be at least 15%

Special Considerations: Soft Tissue

  • In digital imaging, brightness and contrast can be used instead
  • A decrease in mass can display the larynx

Special Considerations: Contrast Media

  • Positive contrast calls for higher exposure and are radiopaque
  • Negative is air and are normally less exposed

Mastering exposure variable effects is essential for technologists

Studying That Suits You

Use AI to generate personalized quizzes and flashcards to suit your learning preferences.

Quiz Team

Related Documents

More Like This

Use Quizgecko on...
Browser
Browser