What must be assessed to ensure proper viewing from either side of the examination table?

  • Rotation of the video display monitor(s)
  • Ease of movement of the video display monitor(s)
  • Range of motion of the video display monitor(s) (correct)
  • Sagging of the video display monitor(s)

What should be within the active range of motion of the fluoroscopists?

  • X-ray generator
  • Protective radiation shields (correct)
  • Radiological exposure parameters
  • Solid-state detector systems

Which system provides detailed operation instructions for solid-state detector systems?

  • X-ray generator parameter settings
  • Fluoroscopists' active range of motion
  • Automatic dose rate control (ADRC) system
  • Manufacturers of solid-state detector systems (correct)

What should remain in a fixed position without movement or sagging when released?

<p>Protective radiation shields (C)</p> Signup and view all the answers

Which equipment must be evaluated to ensure easy movement and stationary positioning?

<p>Ancillary equipment installed on overhead arm(s) (A)</p> Signup and view all the answers

For many IXE systems, which system controls X-ray generator parameter settings?

<p>Automatic dose rate control (ADRC) system (A)</p> Signup and view all the answers

What should be used to cover the image intensifier (II) or flat panel image receptor (FPIR) during generator parameter tests in service mode?

<p>A piece of lead at least 1/8 inch thick (D)</p> Signup and view all the answers

What should the SSDS be positioned as during the experimental setup for acceptance testing?

<p>PERP (C)</p> Signup and view all the answers

What is the purpose of the specially designed protective piece that may be available from the manufacturer?

<p>To protect the II or FPIR from excessive radiation (C)</p> Signup and view all the answers

Who should be consulted if the physicist is not familiar with service mode during generator parameter tests?

<p>The service engineer from the specific vendor (D)</p> Signup and view all the answers

What term has been renamed by the IEC as "PERP" in relation to interventional reference points?

<p>&quot;PERP&quot; (C)</p> Signup and view all the answers

Where should the SID be set during testing for the imaging system?

<p>Maximum available distance (C)</p> Signup and view all the answers

What should be tested during both single-plane and biplane operation in a ceiling-suspended single-plane system?

<p>Collision avoidance safety switches (B)</p> Signup and view all the answers

In the L-arm of the fluoroscopy system, what capabilities are associated with the rotation about the Y-axis?

<p>Angulation capabilities (D)</p> Signup and view all the answers

What should be carefully recorded if the L-arm of a system can be flipped?

<p>Coordinates and signs displayed for peak PSD calculations (C)</p> Signup and view all the answers

Which component's entire movable range should be tested on both sides of the examination table in a ceiling-suspended single-plane system?

<p>C-arm (D)</p> Signup and view all the answers

What should be tested for proper function in a fluoroscopy system with a tabletop offering various motions?

<p>All available table motions and locks (C)</p> Signup and view all the answers

Which feature should be assessed to ensure safe operation during single-plane and biplane use of the fluoroscopy system?

<p>Collision avoidance safety switches (A)</p> Signup and view all the answers